IMACC Script Setting
Modbus Control Version
E-12
Omni FTIR Multi Gas CEMS Instruction Manual
"alarm". When "alarm" equals 0, the measurement data would be valid
defined by variable "Data_Valid". Non valid data is defined by either
"Alarm=1" or "Alarm=2". The corresponding settings are defined by three
setting blocks starting with "if expression", as indicated.
Script setting for system command and status
This setting is included in the second container, as shown.
There are three script setting blocks and one calculation block. The first
script setting is used to assign the digital input channels to five system
commands stating from 2 up to 6. The subsequent calculation block is used
to send the value of "system_cmd" to the variable of "system_status".
The second script setting is used to assign the digital I/O output channels
for each system mode using the "system_status" variable.
Thermo Fisher Scientific