Thermo Fisher Scientific
Figure 5–2. Filter Housing Lid Assembly
6. Replace the large graphite gasket lid seal with a new one. Line up the
through hole in the lid seal with the hole in the lid.
7. Inspect the 3 graphite gasket filter element seals and replace if
necessary.
8. Install the retaining washer, the 3 graphite filter seals, and the new filter
element onto the assembly. See
installation. Thread on the knurled filter lock screw and tighten until
snug.
9. Before re-installing the filter housing lid assembly, the filter housing
itself can be cleaned using a clean cloth dampened only with clean
water.
10. Re-install the filter housing lid assembly into the filter housing by
pushing it all the way in. Flip the retaining clamp back into position so
that it engages the small bar handle, and then turn the bar handle
clockwise to tighten. Turn the "T" handle approximately 1 turn
clockwise or until snug.
11. Re-install the probe protection shield onto the probe. Close the two
pressure clamps to secure the shield to the probe assembly.
Preventive Maintenance and Servicing
Maintenance for the Sample Probe
Figure 5–2
for the correct order of
Omni FTIR Multi Gas CEMS Instruction Manual
5-9