IMACC Script Setting
CEMS_SampleMode
Auto blowback setting
By default, the system blowback is performed every two hours, starting
from 00:30:00 am. The schedule and the valve operation setting are
included in the same green block.
Click the left top corner labeled "ifTimeCondition:equal to=="so the
schedule setting block pops up.
As seen in the pull down manual, there is a list of times starting from
00:30:00. Users can adjust the blow back frequency by clicking the "add"
and "remove" button. The time window for each blowback action is set as
5 minutes.
Right below the schedule setting is the script for the blowback valve
operation.
Thermo Fisher Scientific
Omni FTIR Multi Gas CEMS Instruction Manual
E-9