Calibration
System Calibration Procedure Using Calibration Factor Table in Method
4-20
Omni FTIR Multi Gas CEMS Instruction Manual
Table 4–1. Calibration Factors
Gases
CO2
NH3_Hi
NH3_lo
CO2_HI
SO2_LO
SO2_dry
NO2_wet
NO_wet
CO_HI
N2O
SO2_Hi
HCl_Lo
HCl_Hi
H2O_LO
HF_dry
H2O_HI
CH4
CO_LO
NO_dry_LO
HF_wet
HCl_dry
HCl_wet
NH3_wet
NH3_dry
NO_dry_Hi
H2O
NO2_dry
SF6
Range
0 to 30%
16 to 105 ppm
0 to 70 ppm
0 to 30%
0 to 50 ppm
0 to 350 ppm
0 to 60 ppm
10 500 ppm
57 to 766 ppm
0 to 50 ppm
0 to 350 ppm
0 to 13 ppm
15 to 100 ppm
5% to 40%
0 to 14 ppm
3% to 35%
0 to 1000 ppm
10 to 100 ppm
1 to 103 ppm
0 to 15 ppm
0 to 13 ppm
13 to 100 ppm
0 to 106 ppm
0 to 106 ppm
60 to 1253 ppm
1 to 50 ppm
0 to 6 ppm
H20
<13%
1%
1%
<2%
<0.5 %
>2%
<1%
>0.5 %
<0.5%
>0.5%
>0.5%
<0.5%
<1%
<1%