System Component Description
FTIR

Detector

The FTIR analyzer is configured with a TE cooled DTGS detector with
Potassium bromide (KBr) window, which provides usable data in the range
between 12,500 cm-1 and 350 cm-1.
Note The KBr window material is hygroscopic, or absorbs water from the
air, so the FTIR instrument must be continuously purged to remove
water.

Gas Cell

The FTIR analyzer is using a gas cell with a 5.2 meter optical path length.
The standard gas cell is a general-purpose, low-pressure gas cell, designed to
be used with a variety of sample gases and sampling environments. This gas
cell is suitable for use in both ambient and elevated temperature conditions,
operating reliably at temperatures up to 185 °C. The gas cell is fitted with
zinc selenide (ZnSe) sample windows.
Figure 7–8. Major System Components
Gas Cell Heating
The 5.2-meter gas cell is supplied with a heating jacket. The heating jacket
can maintain the gas cell at a constant temperature to improve quantitative
Jacket
accuracy and prevent condensation of sample gases. The heating jacket is
designed to heat the gas cell to a temperature of 185 °C.