WEEE Compliance This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the following symbol: Thermo Fisher Scientific has contracted with one or more recycling/disposal companies in each EU Member State, and this product should be disposed of or recycled through them.
Page 5
About This Manual This manual provides information about installing, maintaining, and servicing the Omni FTIR Multi Gas CEMS. It also contains important alerts to ensure safe operation and prevent equipment damage. The manual is organized into the following chapters and appendixes to provide direct access to specific operation and service information.
Page 6
Do not allow cleaning solutions or liquids to run into the analyzer, its power ▲ supply, or any powered components. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 7
The source element becomes extremely hot during normal analyzer operation. Always turn off the analyzer and allow the elements to cool for at least 15 minutes Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Page 8
Marking of electrical and electronic equipment which applies to electrical and electronic equipment falling under the Directive 2002/96/EC (WEEE) and the ▲ equipment that has been put on the market after 13 August 2005. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 9
Available 24-hours a day and seven-days a week, the online library provides quick access to information regardless of time zone or office hours. To register for an account or log in, please visit www.thermoscientific.com/aqilibrary. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Page 10
About This Manual Where to Get Help Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Connect the Power to the FTIR............2-6 Connecting the Computer Cable to the FTIR........2-7 Connect the Pressure Transducer Cable to the FTIR ....... 2-7 Connect the Cell Heater Cable to the FTIR........2-7 Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Page 12
Server ..................... 3-10 Script Engine ................. 3-10 OMNIC ................... 3-11 Result....................3-13 NI-DAQ ................... 3-13 CALGrafix ..................3-13 CDM-USB Driver ................3-13 Data Server Software ................. 3-13 ImaccCurrentLoop................. 3-13 Configure..................3-14 viii Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 13
Replacing the Gas Cell ..............5-42 Gas Cell Alignment................5-44 Replacing the Gas Cell Heater............5-51 Replacing the Heated Lines............... 5-54 Replacing the PC ................5-55 Replacing the PC Console..............5-57 Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Page 14
System Component Description ..............7-1 Sample Probe ..................7-2 Temperature Controller ..............7-3 Heated Sample Line ................7-3 Heated Pump/Valve Assembly ............7-4 Hydrator ..................... 7-4 Pneumatic Panel ................. 7-4 Sample mode ................... 7-5 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 15
Function Code .................B-3 Data ....................B-3 Error Check ..................B-3 Function Codes...................B-3 (0x03) Read Holding Registers............B-3 Modbus Parameters Supported ............B-5 Appendix C Interfacing a DCS/PLC with the Omni FTIR Multi Gas CEMS....C-1 Introduction..................C-1 Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Page 16
Contents Connecting a DCS/PLC to the Omni FTIR Multi Gas CEMS ..C-2 Establishing Communications............C-3 Analog Outputs................C-3 Digital I/O..................C-4 Solid-State Relays (SSR)..............C-6 SSR Rail..................C-6 Appendix D Calogix Setup....................D-1 Hardware Description................ D-1 Finding the Correct COM Port ............D-2 Setting the Virtual COM Port ............
Page 17
Figures Figure 1–1. Omni FTIR Multi Gas CEMS ............1-2 Figure 1–2. Omni FTIR Multi Gas CEMS System Overview....... 1-3 Figure 2–1. Removing the shipping Screws............2-3 Figure 2–2. Analyzer Installation................ 2-4 Figure 2–3. Purge and Power Connections ............2-5 Figure 2–4.
Page 18
Figure 5–59. Sample System Heater Assembly Component Replacement and Heated Block Removal..................5-80 Figure 5–60. Check Valve Replacement ............5-84 Figure 5–61. Solenoid Valve Replacement............5-87 Figure 5–62. Pneumatic Panel Components Replacement.......5-90 Figure 5–63. Solenoid Valve Replacement............5-93 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 19
Figure C–3. Digital I/O - Rear Panel Terminals..........C-5 Figure C–4. SSR Rail................... C-6 Figure C–5. SSR Indicator Light Configurations..........C-6 Figure D–1. Base Unit and Four Controllers............D-1 Figure D–2. Temperature Controller Connection ..........D-2 Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Page 20
Figures Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 21
Table B–3. System Command Values ..............B-6 Table B–4. System Flag Bits................B-7 Table C–1. Analog Output Connections ............. C-3 Table C–2. Digital I/O Connections..............C-5 Table D–1. Controller Values................D-10 Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual xvii...
Page 22
Tables xviii Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
The entire sample stream, from the probe stinger to the analysis bench, is maintained at a temperature of 190 °C to prevent sample loss or condensation. The Omni FTIR Multi Gas CEMS has the unique ability to switch to a Flowback mode that purges the sample stream all the way from the analyzer through the sample line to the tip of the probe.
● “Typical System Flow” on page ● “System Power Distribution” on page ● “Specifications” on page ● “Other Manuals” on page 1-10 ● Figure 1–1. Omni FTIR Multi Gas CEMS Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Note This manual is for use with the Omni FTIR Multi Gas CEMS integrated by Thermo Fisher Scientific. Systems integrated by others will vary from the information provided in this document.
Instrument air filter assembly ● FTIR The FTIR analyzer is a Thermo Scientific Antaris IGS instrument which is based on a Michelson interferometer which measures the interaction of infrared (IR) radiation with the sample. The FTIR analyzer includes a 5.2-...
Capture measurement data in real-time ● Allow the operator to trigger a calibration check sequence ● Manually control calibration gases ● Mark data taken during any calibration purge or activity ● Generate reports ● Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
● Temperature controller ● Specifications This section provides technical data on: Compounds ● Performance ● Electrical, data, and pneumatic connections ● Power and instrument air ● Dimensions and weight ● Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
<3 minutes for all non-reactive gases <5 minutes for NH , HCl and HF spiking <10 minutes for NH , HCl and HF dry Availability >98% +60 to +95 °F/+15 to +35 °C Ambient temperature Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Total Consumption 9080 W FTIR electronics 700 W PC with console 1200 W Air Purifier 100 W 24 VDC Supply 250 W Temp Controllers <6000 W Sample Probe 630 W Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
787.4 x 1778 x 914.4 mm (built system) Weight ~440 lbs/200 kg (without doors) Color Thermo Fisher Scientific Blue PMS 646 +60 to +95 °F/+15 to +35 °C (without doors) Ambient temperature Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Shut Down This chapter provides the recommended procedures and settings for installing the Omni FTIR Multi Gas CEMS. Installation includes lifting the FTIR instrument, unpacking and inspection, installing and connecting the FTIR in the rack and connecting sample and exhaust lines to the complete system.
Custom flange and barrel configuration may be ordered. Remove Shipping To prevent damage during shipment, the analyzer is shipped with a restraint in the Michelson interferometer. You will need to open the Restraint Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
CAUTION Lift from each side of the analyzer, not from the front and back. Damage to the unit may occur. ▲ Position the analyzer on the rack slide per Figure 2–2. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Use 2 people to install FTIR to shelf. Carefully place FTIR on top of shelf while lining up shock mount feet holes to the 8 studs on the FTIR shelf. Secure with 8 nuts. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Connect the FTIR The following connections need to be made to the FTIR unit: Purge, Power, Computer, Pressure, Cell heater and Sample lines (Figure 2–3). Unit Figure 2–3. Purge and Power Connections Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Connect the power cord from the power supply to the circular socket on to the FTIR the back chassis of the FTIR. Tighten the two thumb screws attached to the connector. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
The flowback hydrator is located on the right side of the rack. At this time we do not recommend filling the hydrator with water for normal operation of the system, but it may be used in the future. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Consult Thermo Fisher Scientific for longer lengths. Calibration Gases The Omni FTIR Multi Gas CEMS has the capability of simultaneous measurement of ten or more gases. The actual gases required will depend on the application. Two ports are provided for the connection of separately controlled calibration gases, marked ‘Spike’...
LEDs indicate instrument status. The ‘Scan’ light should blink intermittently in normal operation once the PC has booted and the IMACC FSS software is running; the other lights should be steady. 2-10 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
MODBUS or manually controlled with buttons located on the bottom of the corresponding solenoid valves located on the inside of the pneumatic panel (Figure 2–6). Figure 2–6. Pneumatic Panel Rear View Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 2-11...
System Installation, Start Up and Shut Down System Startup Drag the shortcuts into the Startup folder. Close the folder. Maintenance Turn the switch to the “Measurement” position to complete the Start-up procedure. Switch Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 2-13...
Note Once the initial startup procedure has been done, remember to turn the Maintenance switch to the “Measurement” position each time the system is powered up. ▲ 2-14 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
This .layout will be designed in advance for each installation by Thermo Fisher Scientific and will call a pre-configured script file. Both the layout and the script are also user-configurable. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 49
The lower half of the screen consists of four tabbed displays: Spectrum Plot that displays spectral plots of single beam intensities and absorbance, and three Concentration Plot tabs that display plots of concentrations versus time. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Optical path – The optical path for which this control displays results. Display style – There are three styles: list, grid, and dialog. Compound name – Determined by the spectral processing method used in the script. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
FTIR status, percentage complete, numbers of scans completed, and pressure. Settings: Select Parameters – Check the parameters to be displayed. Order Parameters – Use arrows to reorder the selected parameters. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Show as percentage – Shows the data scaled to 100%. Load Configuration – Loads a saved configuration of the data limits and colors (*.desc file). Save Configuration – Saves a configuration of data limits and colors. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Leaving the Timespan box blank will display all of the data. Autoscroll Timespan – This is a shorter time period (subset) of the Timespan over which data is displayed on the top graph in the concentration plot control. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
The computer must be connected to a running FTIR. Directory – Pointing the control to a directory containing *.con files. File – Pointing the control to a single *.con file. Data files See the following: Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 55
There is the option to collect a time and date stamped spectrum file for each minute’s data point in C:\Spec. This allows for post-processing of the data at any time in the future. There is also the capability to not save these files. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
The script editor can be launched by double clicking on the ImaccScriptEeditor icon (See diagram). A script file can be loaded in the editor or by using the file menu. 3-10 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
It can also be used for diagnostic purposes and when re-aligning the spectrometer after changing the beamsplitter, IR source, laser, or detector on the optical bench (See Chapter 5, “Preventive Maintenance and Servicing”). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 3-11...
Page 58
Selecting the ‘Diagnostic’ tab brings up a menu with the ‘Align’ button and a readout of the detector maximum and minimum voltage. The align button performs an automatic optimization of the mirror position. See Chapter 6, Troubleshooting > “Concentration Errors”. 3-12 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
The Valid Data Settings section transmits a TTL signal that describes the status of the instrument on any of the digital output lines. The Configure section allows Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 3-13...
The DAQ Board specifies the National Instruments device number assigned to the current loop board. Value 1 must be specified. The AutoStart feature automatically begins data transmission when the current loop program is opened. 3-14 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
0–20 mA. Table 3–1. Analog Current Channel Assignments Compound Channel Sulphur dioxide Nitric oxide Nitrogen dioxide Nitrous oxide Ammonia Hydrogen chloride Hydrogen fluoride Carbon monoxide Carbon dioxide Water Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 3-15...
Page 62
MODBUS master to ensure that it matches the register number on the instrument. ▲ Note For additional information on how to read registers and interpret the data, refer to Appendix B, “Modbus Protocol”. ▲ 3-16 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 63
Modbus registers from the concentration file data. The nominal value is 10000 ms. The Settings button can be used to configure the COM port associated with the Modbus network for serial Modbus. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 3-17...
Page 64
Windows. If IMACC Modbus is added to the Windows Startup folder and the AutoStart capability is enabled, the program will automatically open and start upon bootup. 3-18 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
~10 minutes, a new background spectrum is collected and compared with the previous one. A good newly acquired background spectrum is saved for subsequent sample spectrum processing. If the new zero background Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
PLC to perform system daily zero check. To ensure a sufficient zero air flow into the sample probe, it would be necessary to adjust the rotameter, shown in Figure 4–1. Figure 4–1. Adjusting the Rotameter on the Pneumatic Panel Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
The only difference is that the Gas Procedure system operator needs to swap the span gas cylinder with the audit gas source (dry gas cylinder or wet/hot HovaCal output). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
0.005 mol/l In addition to a HovaCAL, a high concentration gas cylinder for HCl (~1000 ppm) is recommended. A gas calibrator such as Thermo Scientific Model 146i should be used to dilute HCl to the desired value for both span check and spike check.
When running IMACC Monitor, press Stop Script and press OK go to tab OMNIC. Spectrum Collection Procedure In OMNIC, select Exp Set, tab Collect, and set the following. No. of scans: 30, Resolution: 0.5, Final Format: Interferogram, Correction: None. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Page 70
400, Gain: 1.0, Velocity: 0.3165, Aperture: 15. Press OK. Then go to Col Smp (Collect Sample). Set the name of the new background in the following format: zero air background and date (year-mm-dd) Press OK. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 71
Make sure the system is flushed with zero air (wait 5 min) then press OK again. The system will now collect 30 scans when prompted with “Add to IMACC capture?” press Yes. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Page 72
Select tab Process and select Reprocess…. Verify or set the following. Resolution: 0.5, Apodization: Happ-Genzel, Phase correction: Mertz, Zero filling: None, Final format: SingleBeam, Correction: None, Saved spectral range: 4500 – 400. Press OK. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 73
Select tab File and select Save As. Go to directory BKGS and press Set Filename to Title. Go to tab My Documents (If not opened yet then open it with Explorer). Move previous background file to monthly archive and delete BKG. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
“save background” mode. The system starts to collect the spectrum and saves it as “c:\backgrounds\bkg.spa. After completion, switch the system command to “sample”. 4-10 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Scan time: 0:56, Low Cutoff: 400, High Cutoff: 4500. Use Method Shift Settings: Not enabled; Use Dynamic References: Enabled; Use Bias Correction: Not enabled; Use Dynamic Shifting: Enabled; Forget Analysis State: Enabled Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 4-11...
Page 76
Bench Parameters: Enabled Pathlength: 5.209, Source: IR, Aperture: 15.0, Velocity, 0.2659, Gain: 4.0, Resolution: 0.5, Correlation: Simple, Default Temperature: 185, Default Pressure: 750 Press OK to close Collect EZ Setup box. 4-12 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 77
Optical Path: FrontMain, Type of data: IEEE float, Code: CellPressure, Register:0x0a2a, COM Port: 1, Parity: None, Baud: 9600, Stopbits: 1, Databits: 8, Flow Control: None Press OK to close ModbusMultiSensor Settings box. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 4-13...
Page 78
Press map to check that all compounds are listed. If not, enter the required compound and click the add button. Check that the corresponding compounds in the method are selected for the highlighted compound. 4-14 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 79
Scan Time (MM:SS): 01:00; Averaging Count: 3 Press GO. Upon completion of the process, a new calibration factor will be calculated. Then press “Save Factor to Method File” button to save the calibration factor. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 4-15...
Page 80
(without saving changes again). The reading should equal the value of the calibration gas (spike value). When you return to ImaccMonitor, the reading should equal the value of the calibration gas. 4-16 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 81
Calibration System Calibration Procedure Using IMACC Collect When pressing Start Script, the new zero air background will now be used. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 4-17...
Click the “linearity” tab and you will see the tab button “Calibration Factors”. Click the button and open the table as shown below. 4-18 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 83
O value should be calibrated with dry certified gases. For those with an H O content range, such as HCl_wet, the calibration must be done in that wet condition. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 4-19...
>0.5% NH3_wet 0 to 106 ppm >0.5% NH3_dry 0 to 106 ppm <0.5% NO_dry_Hi 60 to 1253 ppm <1% NO2_dry 1 to 50 ppm <1% 0 to 6 ppm 4-20 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
“Replacing the Gas Cell Heater” on page 5-51 ● “Replacing the Heated Lines” on page 5-54 ● “Replacing the PC” on page 5-55 ● “Replacing the PC Console” on page 5-57 ● Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Page 86
● “Electrical Panel Components” on page 5-71 ● “Sample System Heater Assembly Components” on page 5-76 ● “Pneumatic Panel Components” on page 5-90 ● “Service Locations” on page 5-97 ● Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
FTIR Purge Gas Generator Maintenance Kit (kit contains 113309-00 1 year one BX instrument air filter) DX Instrument Air Filter (box of 10) 9810 9-12 months BX Instrument Air Filter (box of 10) 9811 9-12 months Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Page 88
Assembly, Replacement, Heated Line (FTIR to Sample 113390-00 System) USB Cable, 10' (FTIR to PC) 112796-00 Temperature Controller Part Number Replacement Period Temperature Controller Assembly 112701-00 Relay, SSR, Crydom 2425 103344-00 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Solid State Relay, DIN Rail, 5VDC 112224-00 Digital I/O 100 Pin to 32 Terminal Breakout PCBA 112165-00 Analog I/O 68 Terminal Breakout PCBA 109665-00 Cable, Temperature Alarm, 10' (Temp Controller to 110252-00 Electrical Panel) Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Important Do not attempt to clean or even touch the mirror surfaces inside the analyzer. These front-surfaced mirrors are easily scratched. Dust will not harm the mid-infrared signal, but fingerprints and scratches can degrade spectral performance or permanently damage the mirrors. ▲ Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Graphite filter seals (as required) CAUTION High surface temperatures present. Allow sufficient time to cool. ▲ 1. Flush the probe with inert gas or air to clear any toxic gases. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Remove the filter element, the 3 graphite gasket filter element seals, the retaining washer, and the graphite gasket lid seal from the assembly. Properly dispose of the used filter element and the graphite gasket lid seal (Figure 5–2). Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Turn the “T” handle approximately 1 turn clockwise or until snug. 11. Re-install the probe protection shield onto the probe. Close the two pressure clamps to secure the shield to the probe assembly. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
The pre-filter could be cleaned with water in an ultrasonic bath and then re-used. If doing so, perform several blow backs on the system after installation is complete to dry the pre-filter. 5-10 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
WARNING The desiccant is harmful if swallowed. Follow the manufacturer’s instructions to dispose of desiccant packets properly. ▲ CAUTION Avoid burn and fire hazards. Do not attempt to regenerate spent desiccant packets. ▲ Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-11...
Be careful not to unplug or damage any cables connected to the panel (Figure 5–4). 5-12 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
1. Using the Phillips screwdriver, remove the FTIR cover panel in the front of the system rack. Slide the FTIR drawer out of the rack to gain access to the analyzer. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-13...
Slide the FTIR drawer out of the rack to gain access to the analyzer. 2. Using the flat blade screwdriver, carefully pry the fan cover grill off of the fan filter housing. Remove the dirty filter (Figure 5–5). 5-14 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
1. Using the Phillips screwdriver, remove the FTIR cover panel in the front of the system rack. Slide the FTIR drawer out of the rack to gain access to the analyzer. 5-16 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
5. Loosen the thumbscrew that secures the source assembly to the analyzer. Rotate the clamp bar that holds the source in place 90 degrees counterclockwise (Figure 5–7). Figure 5–7. Removing the Source Assembly Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-17...
13. Wait 15 minutes for the new components to stabilize thermally and then align the analyzer by following the instructions in Chapter 3. (See “OMNIC” software on page 3-11.) 5-18 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
If more room is needed, use the 9/16-inch wrench to loosen the nuts to disconnect the heated lines from the sample system heater assembly and/or the FTIR Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-19...
2 springs and an alignment pin. Release the spring clamps by pushing down on the clamp release hooks near the bottom of the analyzer compartment (Figure 5–10). 5-20 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
If the laser is misaligned, the analyzer will not scan. ▲ 9. Place the springs over the laser tube and re-attach the clamp release hooks. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-21...
Important Note A system operator can replace the laser, however it is recommended that only our certified service engineers or on-site maintenance personnel who have been trained by us replace the laser. 5-22 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 107
Lift the detector straight up to disengage it from the connector and to clear the 2 alignment pins. Remove it from the analyzer (Figure 5–12). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-23...
6. Install the new detector assembly in the analyzer over the 2 alignment pins, taking care to engage the connector. Tighten the thumbscrews on the base plate until snug. 5-24 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Whenever the analyzer compartment is open, there is a potential for damage to sensitive components, which will affect the performance of the system. ▲ Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-25...
3. Using the small Phillips screwdriver, loosen the captive screw that holds the pressure sensor cable in place, then pull the cable assembly off of the sensor (Figure 5–14). Figure 5–14. FTIR Wiring and Tubing Connections 5-26 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
9. Locate the rectangular top access panel toward the rear of the analyzer. Using the 5/32-inch hex wrench, remove the 4 corner mounting screws and remove the top access panel (Figure 5–16). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-27...
12. Place the rectangular top access panel back into position. Using the 5/32-inch hex wrench, re-install the 4 corner mounting screws and tighten until snug. 13. Follow steps 2–8 in reverse to complete the installation procedure. 5-28 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
2. Grasp the bottom of the filter assembly, then with a slight upward pressure, turn it to the left until it stops (the word “Lock” should be directly facing you) and pull it straight down (Figure 5–19). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-29...
4. Take the replacement filter and screw it onto the threaded rod. Once some resistance is felt, turn it approximately another 1/2 turn until it is snug (Figure 5–21). 5-30 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Large Phillips screwdriver 1. Note the pressure on the house air regulator, then pull up on the knob to unlock it and turn the pressure down to zero (Figure 5–22). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-31...
5. Using the large Phillips screwdriver, loosen but do not remove the 4 mounting screws (Figure 5–22). 6. Lift the purge gas generator up and off of the 4 mounting screws (Figure 5–22). 5-32 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Simply grasp the hand grips on the panel and carefully lift it about an inch, then pull the panel away from the rack to remove it (Figure 5–23 Figure 5–24). Figure 5–23. Lock and Hand Grips Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-33...
Panel” for better accessibility (optional step). 3. Using the Phillips screwdriver, remove the FTIR cover panel in the front of the system and the blank panel above it (Figure 5–25). 5-34 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Figure 5–25. Remove Screws and Two Lowest Panels 4. Slide the FTIR drawer out of the system to access the rear of the unit (Figure 5–26). Figure 5–26. FTIR Shelf Out Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-35...
Page 120
12. Using the 9/16-inch wrench, disconnect the two 1-meter heated lines from the elbow fittings at the rear of the unit (Figure 5–27). Note or mark their position to facilitate re-connection. 5-36 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
14. Using 2 people, carefully lift the FTIR unit off of its base. For details, see “Lifting” on page 2-1. 15. Re-install the FTIR by following the previous steps in reverse order. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-37...
2. Using the Phillips screwdriver, remove the 8 mounting screws and lift the panel off of the base (Figure 5–29). Figure 5–29. FTIR Power Supply Replacement 3. Disconnect the power cord from the back of the power supply. 5-38 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
4. Slide the FTIR drawer out of the system to access the rear of the unit. 5. At the rear of the FTIR cell, disconnect the temperature sensor connector from the sensor by pulling it off, exposing the 3 metal prongs (Figure 5–30). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-39...
10. Re-install the sensor connector onto the 3 metal prongs of the sensor, with the red dots lining up with each other. 11. Follow steps 2–4 in reverse order to complete the procedure. 5-40 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
4. Using the small Phillips screwdriver, loosen the captive screw that holds the pressure sensor cable in place, then pull the cable assembly off of the sensor (Figure 5–31). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-41...
Gas Cell preferably by a trained service technician. Equipment Required: Wrench, 9/16-inch, 7/16-inch, and 3/4-inch Hex wrench, 5/32-inch and 7/64-inch Phillips screwdriver Gas cell 5-42 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 127
27, but with the tubing removed with the sensor.) 9. Using the 5/32-inch hex wrench, remove the 4 cap screws that mount the gas cell to the mounting plate (Figure 5–32). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-43...
Wrench, 9/16-inch and 11/32-inch Hex wrench, 5/32-inch and 9/64-inch Phillips screwdriver Visible laser module for gas cell alignment (field service only) Refer to “Replacing the Gas Cell” for additional tools needed 5-44 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
4. Using the 5/32-inch hex wrench, remove the 4 screws that mount the nickel plated aluminum cell to the plate. Carefully remove the gas cell casing, leaving the mirrors in place (Figure 5–33). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-45...
“Open area 1” where the IR light comes out of the bench illustrated in Figure 5–36. Figure 5–35. Visible Laser Module for Gas Cell Alignment 5-46 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Mirror #1 is used to adjust the beam image on mirror #2, mirror #2 is used to adjust the beam image on mirror #3, and mirror #3 is used to position the beam image on the objective Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-47...
Figure 5–38. Typical Image Pattern on the Field Mirror Figure 5–39. Location of Components on Back Plate of Gas Cell 5-48 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 133
For the 5.2 meter gas cell, the maximum number of passes should be 20, with the number of 5 images on the bottom row as shown in Figure 5–40. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-49...
4, 5 and 6 shown in Figure 5–34. Manipulate these mirrors to maximize the IR signal (Max) on the interferometer display as shown in Figure 5–41. Figure 5–41. Typical Interferometer of Omni FTIR CEMS 5-50 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Use the following procedure to replace the FTIR gas cell heater. Gas Cell Heater Equipment Required: Wrench, 9/16-inch, 7/16-inch, and 3/4-inch Hex wrench, 5/32-inch and 7/64-inch Phillips screwdriver Razor blade/utility knife Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-51...
(Figure 5–42). Figure 5–42. Cutting Insulation Paper at the Seam 4. Remove the insulation from the gas cell. Wearing gloves is suggested when handling the insulation (Figure 5–43). 5-52 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
8. Place the insulation back on the gas cell and attach it with a couple of pieces of high temperature tape, making sure that it fits snugly. 9. Refer to the procedure for “Replacing the Gas Cell” to re-install the gas cell. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-53...
7. Using the 9/16-inch wrench, loosen the nut and disconnect the heated line(s) to be replaced from the elbow fittings at the rear of the FTIR (Figure 5–45). Note their position to facilitate re-connection. 5-54 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
9. Install the new heated line(s) by following the previous steps in reverse order. Attach insulation and tie wraps where appropriate. Replacing the PC Use the following procedure to replace the PC. Equipment Required: Phillips screwdriver Flat blade screwdriver Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-55...
Figure 5–46. PC Removal 5. Slide the PC console back into the rack. Depress the tab at the rear of each slide rail of the PC console to unlock it (Figure 5–46). 5-56 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
(Figure 5–46). 3. At the rear of the system, unplug the PC console power connector, as well as the connectors for the console, mouse, and keyboard (Figure 5– 47). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-57...
8. Push the PC console in just enough to get the rear rails on each side started in the rack sliders. Once both sides are in place, push the console in a couple of inches (Figure 5–48). 5-58 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
A properly grounded antistatic wrist strap must be worn while handling any internal component. If an antistatic wrist strap is not available, be sure to touch the instrument chassis before Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-59...
(Figure 5–50). Note that 4 of the screws are on the back of the PC and 1 is inside. Figure 5–50. Removing the Card Cage Mounting Screws 5-60 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
(See Figure 5–52 Figure 5–53 that show the Digital Board.) Figure 5–52. Removing NI Board Mounting Screw Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-61...
Box Components Replacing External Use the following procedure to replace the 2 external fuses on the FTIR temperature control box. Fuses Equipment Required: Phillips screwdriver Flat blade screwdriver Replacement Fuses 5-62 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
5. The external fuse holders are located under the above cable connectors. To remove the fuses, push the holder in while turning it approximately 1/4 turn counterclockwise and then pull it out (Figure 5–55). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-63...
3. To remove the cover, use the small Phillips screwdriver to remove the 8 screws that mount the cover to the temperature control box and lift the cover off. 5-64 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
2. There are 16 fuses located on the PC Board inside the temperature control box (plus 4 spare fuses.) To remove any one of them, simply grasp the fuse with your fingers and pull it up and out of the connector. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-65...
Use the following procedure to replace the temperature control board Temperature Control inside of the temperature control box. Board Equipment Required: Phillips screwdriver Small Phillips screwdriver Temperature Control Board 5-66 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 151
7. Re-connect all of the cables to their proper locations on the temperature control board. 8. Re-install the temperature control box cover per the above procedure “Removing and Installing the Cover”. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-67...
Page 152
6. Pop the end covers up on the new relay and place it in the temperature control box over the mounting studs in the proper orientation. Thread on the 2 mounting nuts, and then use the 5/16-inch nut driver to tighten them until snug. 5-68 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
5–55). Using the 3/8-inch nut driver, remove the 2 nuts that mount the control module and base unit assembly to the temperature control box. Note the orientation of the assembly. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-69...
A properly grounded antistatic wrist strap must be worn while handling any internal components. ▲ 1. Remove the temperature control box cover per the previously mentioned procedure “Removing and Installing the Cover”. 5-70 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
11. Re-install the temperature control box cover per the above procedure, “Removing and Installing the Cover”. Electrical Panel Use the following procedures to replace the electrical panel components (Figure 5–56 Figure 5–57). Components Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-71...
Use the following procedure to replace the solid state relays on the electrical State Relays panel in the rear of the system. There are 3 24VDC relays and 15 5VDC relays (Figure 5–57). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-73...
Page 158
Be sure that the voltage of the new relay matches the one being replaced. 8. Re-install the electrical panel by following steps 1–4 in reverse order. 5-74 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
7. To install the new power supply, hook the upper tabs of the rear of the power supply onto the din rail, and then firmly push the bottom of the supply toward the panel until it locks into place. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-75...
They are the instrument zero air, the flow back air, the heated sample line, and the auxiliary sample line (if equipped). Note or mark the location of each line to facilitate reconnection. 5-76 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
“Pump”. Turn the connector nut counterclockwise and pull it out of its socket. Remove the cable assembly tie-wraps using the wire cutters. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-77...
“Removing the Sample System Heater Assembly Cover”, skipping steps 5, 6, and 7 (sample system heater assembly doesn’t need to be removed from the rack for this procedure). 5-78 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
“Removing the Sample System Heater Assembly Cover”, skipping steps 6 and 7 (sample system heater assembly doesn’t need to be removed from the rack for this procedure). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-79...
Ejector Air In and the FTIR Out (Figure 5– 59). 6. Pull the entire eductor pump plumbing assembly out of the sample system heater assembly box. 5-80 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 165
13. Follow steps 1–4 in reverse order to complete the procedure. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-81...
RTD is properly seated. 7. Using the 3/16-inch hex wrench, install and tighten the 4 mounting screws in the corners of the heated block assembly until snug. 5-82 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
2. Remove the heated block assembly from the heater box per the previously mentioned procedure, “Removing the Heated Block Assembly”. 3. Turn the heated block assembly around to access the back of the assembly. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-83...
8. Using the wrenches, connect the other end of the check valve to the brass nut with copper tubing and tighten until snug. Some movement of the copper tubing will be required to get it into the check valve. 5-84 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 169
13. Follow the procedures referenced in step 2, “Removing the Heated Block Assembly”, and in step 1, “Removing the Sample System Heater Assembly and Cover”, in reverse order to complete the procedure. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-85...
4. Using the 9/64-inch hex wrench, loosen slightly but do not remove the 5 screws that mount the double coil clamp plate to the heated block. This will allow some freedom of movement of the copper tubing (Figure 5–60). 5-86 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
10. Lift the sample solenoid valve and plumbing assembly out of the heated block (Figure 5–61). Note the orientation of the tubing and fittings to facilitate reconnection. Figure 5–61. Solenoid Valve Replacement Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-87...
Page 172
16. Attach the other piece of copper tubing with brass nut to the union fitting on the adjacent side, but do not tighten it all the way (Figure 5– 61). 5-88 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 173
22. Follow the procedures referenced in step 2, “Removing the Heated Block Assembly”, and in step 1, “Removing the Sample System Heater Assembly and Cover”, in reverse order to complete the procedure. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-89...
“Removing the Left Side Panel” for better accessibility. 2. Note the pressure on the gauge above the pressure reducer to be replaced, so that it can be reset later. 5-90 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
The procedure is the same for either pressure gauge. Pressure Gauges Equipment Required: Wrench, 9/16-inch Pressure gauge 1. Remove the left side panel per procedure “Removing the Left Side Panel” for better accessibility. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-91...
(from the rear of the panel, it is the furthest to the right.) (See (Figure 5–62).)Be sure to order the correct valve. Equipment Required: Phillips screwdriver Small Phillips screwdriver Wrench, 9/16-inch and 3/4-inch 5-92 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
4. Using the 11/32-inch nut driver, loosen and remove the 2 nuts that mount the valve bank assembly to the flow meter panel. Pull the valve bank assembly away from the flow meter panel to access the back side (Figure 5–62). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-93...
The 4 center flow meters have the same part number, but the 2 end flow meters have a different part number. Be sure to order the correct flow meter. Equipment Required: Phillips screwdriver 5-94 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 179
7. Insert the new flow meter into position on the flow meter panel, making sure the orientation is correct. Insert the 2 mounting screws and tighten using the Phillips screwdriver until snug. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-95...
Page 180
12. Re-install the left side panel to complete the procedure. 5-96 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Contact one of the phone numbers Locations below for product support and technical information or visit us on the web at www. thermoscientific.com/aqi. 1-866-282-0430 Toll Free 1-508-520-0430 International Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-97...
Page 183
Chapter 6 Troubleshooting This chapter provides the following troubleshooting and service support information for the Omni FTIR Multi Gas CEMS: “Concentration Errors” on page ● “Boot Up Error/IMACC Software Error” on page ● “Temperature Errors” on page ● “Cell Pressure Errors/Sampling System Errors”...
Cell Pressure = <40 mmHg less than ambient pressure during ● sample mode The following outline addresses possible instrument and alarm problems and proposes checks and possible troubleshooting solutions. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Go to ‘C:\Backgrounds\’ in Windows Explorer and verify that ● file ‘C:\Backgrounds\BKG.spa’ has been updated in the last 24 hours. Refer to Appendix E, “IMACC Script Setting” for proper script configuration. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
● temperatures are OK. Check thermostat inside pump box (See Chapter 5, “Preventive ● Maintenance and Servicing”. ERROR_QUANT_TEMPRESSCON in ImaccMonitor ● Check temp control breaker and AC wiring connections. ● Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Check for blocked exhaust system; make sure exhaust lines are ● less than 10 feet long. Check all temperature readings; make sure they are within 10 ● degrees of setpoints. Check probe breaker. ● Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Check ‘Hardware’ setup in Windows system menu. ● Internal communications errors No temperature readings or heating of system components ● See ‘Temperature errors’ in this chapter. ● No communications with FTIR ● Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Gas mode output or FTIR data valid digital output is not reporting ● the expected mode Click the “View Script” button in ImaccMonitor and verify ● that IMACC Script Editor shows that the script is running. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Contact one of the phone numbers Locations below for product support and technical information or visit us on the web at www. thermoscientific.com/aqi. 1-866-282-0430 Toll Free 1-508-520-0430 International Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 191
“Industrial PC” on page 7-13 ● “Console” on page 7-14 ● “Datalogger” on page 7-14 ● “Electrical Panel Assembly” on page 7-14 ● “AC Power Input” on page 7-15 ● Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
The standard probe has a 4-inch ANSI flange and a 1.0 meter probe barrel length. Other flange sizes and barrel lengths may be purchased depending on requirements. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
15 Amps at 220-240VAC 50/60 Hz. The sample line temperature is monitored using a 2-wire 100 ohm RTD. Additional electrical lines are included to hook up the probe temperature alarm and blowback valve. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
This mode should be executed once every day for collecting instrument background spectrum. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
6 LPM span gas into the stinger via the cal/zero port. This mode is executed for the daily system span check. Figure 7–4. Sample and Calibration Gas Flow Diagram Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
When the beams recombine, constructive or destructive interference occurs, depending on the relationship between the position of the moving mirror relative to the fixed mirror and the wavelength of the light. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Fourier Transform or FT. The function calculates the amplitude of each of the component signals. The amplitude gives the intensity at the corresponding wavelength of light. A Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Figure 7–7. Spectrometer Compartment Source The FTIR analyzer uses Thermo Fishser Scientific’s EverGlo source. The source unit can be replaced as needed. It is pre-aligned and needs no adjustment when it is installed. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
The DTGS detector can accommodate most of the energy from the source, which means the software selects a large aperture size. 7-10 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Jacket accuracy and prevent condensation of sample gases. The heating jacket is designed to heat the gas cell to a temperature of 185 °C. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 7-11...
1 ppm. The generator uses a combination of coalescing filtration, regenerative pressure swing adsorption, and high efficiency particulate filtration to produce laboratory quality, dry, CO -free air from a standard compressed air supply. 7-12 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Front-panel DVD+R drive, 3.5” floppy drive, and USB connectors are provided as additional means to move data off the system as well as install system updates. Figure 7–10. FTIR PC Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 7-13...
Digital solid state relay outputs (Temperature Alarm, Pressure Alarm, ● Maintenance, FID Alarm, FTIR Data Valid, System Gas Mode) Digital Inputs (System Gas Mode) ● Probe Connections (AC Power, Temperature Alarm, and Valve ● Control) 7-14 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
The system requires 60 Amp 220-240 VAC connections. This box includes screw terminals for these connections. Figure 7–12. AC Mains Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 7-15...
(i) normal wear and tear, (ii) accident, disaster or event of force majeure, (iii) misuse, fault or negligence of or by Buyer, (iv) use of the Products in a manner for which Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Page 208
OR WRITTEN, WITH RESPECT TO THE PRODUCTS, INCLUDING WITHOUT LIMITATION ALL IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE. SELLER DOES NOT WARRANT THAT THE PRODUCTS ARE ERROR-FREE OR WILL ACCOMPLISH ANY PARTICULAR RESULT. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
ImaccModbus uses two Modbus registers to transmit concentration and measurement data in floating point format. The TCP version also has the ability to accept an F06: Write Single Register command. For details of the Omni FTIR Multi Gas CEMS Modbus Protocol specification, see the following topics: “Serial Communication Parameters”...
1 and 255 decimal (i.e. 0x01 hex to 0xFF hex). The default value is 70. MBAP Header In Modbus over TCP/IP, a Modbus Application Protocol Header (MBAP) is used to identify the message. This header consists of the following components: Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
For each register, the first byte contains the high order bits and the second contains the low order bits. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Page 212
0x64 The contents of register 9 are shown as the two byte values of 0x02 0x2B. Then contents of registers 10–12 are 0x00 0x00, 0x00 0x64 and 0x00 0x64, respectively. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Table B–1 through Table B–4 lists the Modbus addresses supported for the Omni FTIR Multi Gas CEMS. Parameters Supported IMPORTANT NOTE The addresses in the following table are Protocol Data Unit (PDU) addresses. For Application Data Unit (ADU) addresses, add 40000 to these numbers for holding registers.
This mode is used to check the system span CHECK SPAN2 MODE This mode is used to perform an extra span check, in (N/A) addition to the target gas. CHECK SPAN3 MODE (N/A) Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Modbus Protocol Modbus Parameters Supported Table B–4. System Flag Bits Value Description 0x00000000 Data Valid 0x00000001 Data Invalid Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Checking data values Reading system and alarm status Triggering events (such as calibration checks and probe blowbacks) The Omni FTIR Multi Gas CEMS provides several methods for establishing data communication with a DCS/PLC including: Analog output connections (4-20mA) Digital I/O connections (5V TTL logic)
Interfacing a DCS/PLC with the Omni FTIR Multi Gas CEMS Connecting a DCS/PLC to the Omni FTIR Multi Gas CEMS Connecting a You can connect to the FTIR via Serial or Ethernet communications. Refer Figure C–1 and the “Installation” chapter. For Modbus over Ethernet, DCS/PLC to the use TCP port 502 by default (user selectable).
Interfacing a DCS/PLC with the Omni FTIR Multi Gas CEMS Establishing Communications Establishing This section provides information on the methods available for retrieving system data, accessing system status conditions, initiating system events, Communications and checking system calibration. Analog Outputs The 4-20mA analog output signals are labeled “ANALOG OUTPUT X1”...
Interfacing a DCS/PLC with the Omni FTIR Multi Gas CEMS Establishing Communications Figure C–2. Analog Outputs Digital I/O The FTIR allows the user to access system status conditions via the digital I/O connections. The digital I/O signals are labelled “DIGITAL OUTPUT X2”...
Interfacing a DCS/PLC with the Omni FTIR Multi Gas CEMS Establishing Communications Table C–2. Digital I/O Connections Digital No. Block Terminal Description Digital Input Ground Gas Mode Input: Check Span Gas Mode Input: Check Spike Gas Mode Input: Check Zero Gas Mode Input: Bkg &...
Interfacing a DCS/PLC with the Omni FTIR Multi Gas CEMS Establishing Communications Solid-State Relays TTL outputs from the National Instruments 6509 board are used to drive solid state relays (SSR) for valve control and digital status outputs. The (SSR) SSRs are mounted on a DIN rail on the interior of the electrical panel (Figure C–4).
The USB cable should always be connected to the computer at the lower-left USB socket (Figure D–2). The hardware is also connected to a 220 VAC supply from circuit breaker panel. Figure D–1. Base Unit and Four Controllers Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Use the following procedure to determine the COM port number. 1. Turn the power ON to the Temperature control box. 2. Select “Start” menu and open ‘Control Panel’. Run ‘System’. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 225
Calogix Setup Finding the Correct COM Port 3. Select the Hardware tab in System Properties. 4. Select ‘Device Manager’. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Use the following procedure to set the virtual COM port. Virtual COM Port 1. Open CALGrafix software with the icon on the desktop. 2. Click the second icon from the left in the icon bar. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 227
Calogix Setup Setting the Virtual COM Port 3. Click ‘Browse’ in the ‘Select Devices’ popup menu. 4. Select the correct number for the COM port, Select ‘OK’. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
‘On’ before the controllers are set to run. If the controllers are set to run with the ramp program not enabled, heated line and sample cell temperatures could rise too quickly and damage those components. ▲ Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
None. Enter Segment 2: Segment type SOAK, Soak time Continuous, EOP Status None. 3. Select the ‘Module 2’ tab. Enter Segment 1: Segment type RAMP, Ramp rate 90, Target Setpoint 185, Holdback band 0, EOP Status Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Page 230
Ramp rate 90, Target Setpoint 185, Holdback band 0, EOP Status None. Enter Segment 2: Segment type SOAK, Soak time Continuous, EOP Status None. 5. Close the program menu. The changes will be automatically written to the controllers. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
2. Once a module is selected, enter the correct values according to Table D–1. Values are grouped according to subheadings. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Output1 Inversion FALSE FALSE FALSE FALSE Output2 Inversion FALSE FALSE FALSE FALSE Output3 Inversion FALSE FALSE FALSE FALSE Output1 Inhibit None None None None Channel 1 Control Setpoint Mode PARK D-10 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 233
PARK Subsidiary Mode None None None None Setpoint Band Cycle Time Maximum Power Minimum Power Output Power Manual Power Display Alarm Reset Latched Alarm Programmer Program number Run mode Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual D-11...
(“Sample Mode”) and one main setting block on the right (“System Running Status”). The vertically aligned sub setting blocks contained in the main script setting body are constructed in a sequential loop. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Page 236
Status” setting. This setting is used to log six monitored parameters into the “cons” files. They are gas cell temperature, sample temperature, short inlet sample line temperature, heated box temperature, short outlet sample line Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 237
“Register” context is determined by Calogix setting (0x0a1c, in this case). Go back to the “Modbus Register Setting” panel, click the “serial” button on the top, and view the following setting panel, as shown. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Page 238
High Cutoff” define the low end and high end of the Mid IR spectrum range. They are not recommended to be adjusted by users. The background spectrum used for the absorption spectrum conversion is saved in the Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 239
0.25 ppm is applied in the script setting for HCl measurement. If the users need to make an adjustment, click the “Calculating Setting” block, right below the “Collect EZ” setting block. Its setting subpanel pops up, as shown. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Page 240
3:15 am. If a change is needed, get access to the script block, labeled “Background Spectrum Collection”. Click the green shadow block “ifTimeCondition: equal to ==”, and the subpanel “time setup” pops up. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 241
In general, one daily background collection is adequate. Script setting for background spectrum collection ● Double click the script setting block named “Background Spectrum Collection”so its setting panel pops up, as shown. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Page 242
“C:\Backgrounds\BKG.spa”. Right after the calculation setting is another 5 second time delay setting. At the end of the block is the solenoid valve operation setting for normal sample mode. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 243
00:30:00. Users can adjust the blow back frequency by clicking the “add” and “remove” button. The time window for each blowback action is set as 5 minutes. Right below the schedule setting is the script for the blowback valve operation. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
The Modbus based script named Modbus_control is saved in the script folder of the directory of c:\scripts. Use script edit to open it so that its setting block pops up. E-10 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 245
“alarm” would be flagged. Logically, there would be three alarm output results based on the system temperature status. The validation of CEMS measurement data is determined by the variable Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual E-11...
Page 246
“system_cmd” to the variable of “system_status”. The second script setting is used to assign the digital I/O output channels for each system mode using the “system_status” variable. E-12 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 247
“dev2/port1/line0:7”, pattern “0x050” and time delay setting “0”. Script setting for ModbusMultiSensor ● Next to the script setting for system command and status is the container for “ModbusMultiSensor” setting, as shown. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual E-13...
Page 248
Under the “setup” configuration, the setting panel is similar to the figure below. Highlight the target parameter, such as SampleT, in the list. Double click it so its subpanel pops up, as shown. E-14 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 249
“ModbusMultiSensor” container. As shown, it is the same as described in the CEMS_Samplemode section. Note that all the settings should be the same as the one included in the background spectrum collection. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual E-15...
Page 250
The first setting block is used to define the file directory for all saved data. The second setting block is used to display real time system status on the monitor. E-16 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Do not uncheck any of the existing components or you will uninstall them. 4. Click Next to begin installation. 5. Allow Windows to reboot after installation is complete. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Page 252
6. Enter FTIR Data into the Alias window and click Next. 7. Browse for the data transfer folder and click Next. 8. Check Read and Run scripts, and click Next and Finish to complete the setup. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 253
4. In the entry box Copy CON file to directory, browse for the data transfer folder created earlier. IFSS will save a copy of the current concentration file to the file current.con in that folder. 5. Click OK and save the script. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
Page 254
Access to the data will now be restricted to users who possess valid usernames and passwords for the FTIR computer. In this case, the software written to use IIS access will also have to provide this additional information. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Need help?
Do you have a question about the Omni FTIR Multi Gas CEMS and is the answer not in the manual?