Summary of Contents for Thermo Scientific Omni FTIR Multi Gas CEMS

  • Page 1 Omni FTIR Multi Gas CEMS Instruction Manual Part Number 108004-00 11Apr2013...
  • Page 2 © 2013 Thermo Fisher Scientific Inc. All rights reserved. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Fisher Scientific Air Quality Instruments 27 Forge Parkway Franklin, MA 02038 1-508-520-0430 www.thermoscientific.com/aqi...
  • Page 3 WEEE Compliance This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the following symbol: Thermo Fisher Scientific has contracted with one or more recycling/disposal companies in each EU Member State, and this product should be disposed of or recycled through them.
  • Page 4 Thermo Fisher Scientific WEEE Compliance...
  • Page 5 About This Manual This manual provides information about installing, maintaining, and servicing the Omni FTIR Multi Gas CEMS. It also contains important alerts to ensure safe operation and prevent equipment damage. The manual is organized into the following chapters and appendixes to provide direct access to specific operation and service information.
  • Page 6 Do not allow cleaning solutions or liquids to run into the analyzer, its power ▲ supply, or any powered components. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 7 The source element becomes extremely hot during normal analyzer operation. Always turn off the analyzer and allow the elements to cool for at least 15 minutes Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 8 Marking of electrical and electronic equipment which applies to electrical and electronic equipment falling under the Directive 2002/96/EC (WEEE) and the ▲ equipment that has been put on the market after 13 August 2005. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 9 Available 24-hours a day and seven-days a week, the online library provides quick access to information regardless of time zone or office hours. To register for an account or log in, please visit www.thermoscientific.com/aqilibrary. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 10 About This Manual Where to Get Help Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 11: Table Of Contents

    Connect the Power to the FTIR............2-6 Connecting the Computer Cable to the FTIR........2-7 Connect the Pressure Transducer Cable to the FTIR ....... 2-7 Connect the Cell Heater Cable to the FTIR........2-7 Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 12 Server ..................... 3-10 Script Engine ................. 3-10 OMNIC ................... 3-11 Result....................3-13 NI-DAQ ................... 3-13 CALGrafix ..................3-13 CDM-USB Driver ................3-13 Data Server Software ................. 3-13 ImaccCurrentLoop................. 3-13 Configure..................3-14 viii Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 13 Replacing the Gas Cell ..............5-42 Gas Cell Alignment................5-44 Replacing the Gas Cell Heater............5-51 Replacing the Heated Lines............... 5-54 Replacing the PC ................5-55 Replacing the PC Console..............5-57 Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 14 System Component Description ..............7-1 Sample Probe ..................7-2 Temperature Controller ..............7-3 Heated Sample Line ................7-3 Heated Pump/Valve Assembly ............7-4 Hydrator ..................... 7-4 Pneumatic Panel ................. 7-4 Sample mode ................... 7-5 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 15 Function Code .................B-3 Data ....................B-3 Error Check ..................B-3 Function Codes...................B-3 (0x03) Read Holding Registers............B-3 Modbus Parameters Supported ............B-5 Appendix C Interfacing a DCS/PLC with the Omni FTIR Multi Gas CEMS....C-1 Introduction..................C-1 Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 16 Contents Connecting a DCS/PLC to the Omni FTIR Multi Gas CEMS ..C-2 Establishing Communications............C-3 Analog Outputs................C-3 Digital I/O..................C-4 Solid-State Relays (SSR)..............C-6 SSR Rail..................C-6 Appendix D Calogix Setup....................D-1 Hardware Description................ D-1 Finding the Correct COM Port ............D-2 Setting the Virtual COM Port ............
  • Page 17 Figures Figure 1–1. Omni FTIR Multi Gas CEMS ............1-2 Figure 1–2. Omni FTIR Multi Gas CEMS System Overview....... 1-3 Figure 2–1. Removing the shipping Screws............2-3 Figure 2–2. Analyzer Installation................ 2-4 Figure 2–3. Purge and Power Connections ............2-5 Figure 2–4.
  • Page 18 Figure 5–59. Sample System Heater Assembly Component Replacement and Heated Block Removal..................5-80 Figure 5–60. Check Valve Replacement ............5-84 Figure 5–61. Solenoid Valve Replacement............5-87 Figure 5–62. Pneumatic Panel Components Replacement.......5-90 Figure 5–63. Solenoid Valve Replacement............5-93 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 19 Figure C–3. Digital I/O - Rear Panel Terminals..........C-5 Figure C–4. SSR Rail................... C-6 Figure C–5. SSR Indicator Light Configurations..........C-6 Figure D–1. Base Unit and Four Controllers............D-1 Figure D–2. Temperature Controller Connection ..........D-2 Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 20 Figures Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 21 Table B–3. System Command Values ..............B-6 Table B–4. System Flag Bits................B-7 Table C–1. Analog Output Connections ............. C-3 Table C–2. Digital I/O Connections..............C-5 Table D–1. Controller Values................D-10 Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual xvii...
  • Page 22 Tables xviii Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 23: Chapter 1 Introduction

    The entire sample stream, from the probe stinger to the analysis bench, is maintained at a temperature of 190 °C to prevent sample loss or condensation. The Omni FTIR Multi Gas CEMS has the unique ability to switch to a Flowback mode that purges the sample stream all the way from the analyzer through the sample line to the tip of the probe.
  • Page 24: Figure 1-1. Omni Ftir Multi Gas Cems

    ● “Typical System Flow” on page ● “System Power Distribution” on page ● “Specifications” on page ● “Other Manuals” on page 1-10 ● Figure 1–1. Omni FTIR Multi Gas CEMS Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 25: System Overview

    Note This manual is for use with the Omni FTIR Multi Gas CEMS integrated by Thermo Fisher Scientific. Systems integrated by others will vary from the information provided in this document.
  • Page 26: Sample Probe

    Instrument air filter assembly ● FTIR The FTIR analyzer is a Thermo Scientific Antaris IGS instrument which is based on a Michelson interferometer which measures the interaction of infrared (IR) radiation with the sample. The FTIR analyzer includes a 5.2-...
  • Page 27: Data Logger

    Capture measurement data in real-time ● Allow the operator to trigger a calibration check sequence ● Manually control calibration gases ● Mark data taken during any calibration purge or activity ● Generate reports ● Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 28: Data Communications

    ● Temperature controller ● Specifications This section provides technical data on: Compounds ● Performance ● Electrical, data, and pneumatic connections ● Power and instrument air ● Dimensions and weight ● Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 29: Measuring Components With Minimum Measuring Range And Required Mdl

    <3 minutes for all non-reactive gases <5 minutes for NH , HCl and HF spiking <10 minutes for NH , HCl and HF dry Availability >98% +60 to +95 °F/+15 to +35 °C Ambient temperature Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 30: Connections

    Total Consumption 9080 W FTIR electronics 700 W PC with console 1200 W Air Purifier 100 W 24 VDC Supply 250 W Temp Controllers <6000 W Sample Probe 630 W Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 31: Pneumatics

    787.4 x 1778 x 914.4 mm (built system) Weight ~440 lbs/200 kg (without doors) Color Thermo Fisher Scientific Blue PMS 646 +60 to +95 °F/+15 to +35 °C (without doors) Ambient temperature Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 32: Other Manuals

    Result Software User’s Guide ● M&C Operating Instructions—Gas Sample Probe Series SP® ● Version SP2000, SP2000-H, SP2300-H, SP2400-H Parker Balson® FT-IR Purge Gas Generator ● Models 75-45, 75-52, and 75-62 1-10 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 33: Chapter 2 System Installation, Start Up And Shut Down

    Shut Down This chapter provides the recommended procedures and settings for installing the Omni FTIR Multi Gas CEMS. Installation includes lifting the FTIR instrument, unpacking and inspection, installing and connecting the FTIR in the rack and connecting sample and exhaust lines to the complete system.
  • Page 34: Site Requirements

    Custom flange and barrel configuration may be ordered. Remove Shipping To prevent damage during shipment, the analyzer is shipped with a restraint in the Michelson interferometer. You will need to open the Restraint Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 35: Installing The Ftir Analyzer Into The Rack

    CAUTION Lift from each side of the analyzer, not from the front and back. Damage to the unit may occur. ▲ Position the analyzer on the rack slide per Figure 2–2. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 36: Figure 2-2. Analyzer Installation

    Use 2 people to install FTIR to shelf. Carefully place FTIR on top of shelf while lining up shock mount feet holes to the 8 studs on the FTIR shelf. Secure with 8 nuts. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 37: Figure 2-3. Purge And Power Connections

    Connect the FTIR The following connections need to be made to the FTIR unit: Purge, Power, Computer, Pressure, Cell heater and Sample lines (Figure 2–3). Unit Figure 2–3. Purge and Power Connections Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 38: Connect The Purge Lines To The Ftir

    Connect the power cord from the power supply to the circular socket on to the FTIR the back chassis of the FTIR. Tighten the two thumb screws attached to the connector. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 39: Connecting The Computer Cable To The Ftir

    The flowback hydrator is located on the right side of the rack. At this time we do not recommend filling the hydrator with water for normal operation of the system, but it may be used in the future. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 40: External System Connections

    2 = Probe Temperature Alarm In 3 = Probe Blowback Valve + 4 = Probe Blowback Valve – Terminal Block X4 5 = Ground (Green/Yellow) 6 = Neutral (Blue) 7 = Hot (Brown) Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 41: Instrument Air

    Consult Thermo Fisher Scientific for longer lengths. Calibration Gases The Omni FTIR Multi Gas CEMS has the capability of simultaneous measurement of ten or more gases. The actual gases required will depend on the application. Two ports are provided for the connection of separately controlled calibration gases, marked ‘Spike’...
  • Page 42: Analog

    LEDs indicate instrument status. The ‘Scan’ light should blink intermittently in normal operation once the PC has booted and the IMACC FSS software is running; the other lights should be steady. 2-10 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 43: Adjust/Confirm Flow And Heat Settings

    MODBUS or manually controlled with buttons located on the bottom of the corresponding solenoid valves located on the inside of the pneumatic panel (Figure 2–6). Figure 2–6. Pneumatic Panel Rear View Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 2-11...
  • Page 44: Heat

    C:\Program files\Imacc\IMACC FTIR Software Suite\Imacc System Process Monitor.exe C:\Program files\Imacc\IMACC FTIR Software Suite\Imacc Modbus Server.exe C:\Program files\Imacc\IMACC FTIR Software Suite\Imacc Modbus TCP Server.exe C:\Program files\Imacc\IMACC FTIR Software Suite\Imacc CurrentLoop Server.exe 2-12 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 45: Maintenance Switch

    System Installation, Start Up and Shut Down System Startup Drag the shortcuts into the Startup folder. Close the folder. Maintenance Turn the switch to the “Measurement” position to complete the Start-up procedure. Switch Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 2-13...
  • Page 46: System Shut Down

    Note Once the initial startup procedure has been done, remember to turn the Maintenance switch to the “Measurement” position each time the system is powered up. ▲ 2-14 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 47: Operation

    IFSS – Imacc FTIR Software Suite: ● Monitor – User Interface ● Server – Data Collection & Analysis ● Script Engine ● Data Server Software ● ImaccCurrentloop ● ImaccModbus ● ImaccModbusTCP ● Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 48: Imacc Ftir Software Suite (Ifss)

    This .layout will be designed in advance for each installation by Thermo Fisher Scientific and will call a pre-configured script file. Both the layout and the script are also user-configurable. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 49 The lower half of the screen consists of four tabbed displays: Spectrum Plot that displays spectral plots of single beam intensities and absorbance, and three Concentration Plot tabs that display plots of concentrations versus time. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 50: Concentrations Pane

    Optical path – The optical path for which this control displays results. Display style – There are three styles: list, grid, and dialog. Compound name – Determined by the spectral processing method used in the script. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 51: Processstatus

    FTIR status, percentage complete, numbers of scans completed, and pressure. Settings: Select Parameters – Check the parameters to be displayed. Order Parameters – Use arrows to reorder the selected parameters. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 52: Concentrationplot/Spectrum Plot

    Show as percentage – Shows the data scaled to 100%. Load Configuration – Loads a saved configuration of the data limits and colors (*.desc file). Save Configuration – Saves a configuration of data limits and colors. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 53: Timespan Tab

    Leaving the Timespan box blank will display all of the data. Autoscroll Timespan – This is a shorter time period (subset) of the Timespan over which data is displayed on the top graph in the concentration plot control. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 54: Datasource Tab

    The computer must be connected to a running FTIR. Directory – Pointing the control to a directory containing *.con files. File – Pointing the control to a single *.con file. Data files See the following: Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 55 There is the option to collect a time and date stamped spectrum file for each minute’s data point in C:\Spec. This allows for post-processing of the data at any time in the future. There is also the capability to not save these files. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 56: Remote Control

    The script editor can be launched by double clicking on the ImaccScriptEeditor icon (See diagram). A script file can be loaded in the editor or by using the file menu. 3-10 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 57: Omnic

    It can also be used for diagnostic purposes and when re-aligning the spectrometer after changing the beamsplitter, IR source, laser, or detector on the optical bench (See Chapter 5, “Preventive Maintenance and Servicing”). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 3-11...
  • Page 58 Selecting the ‘Diagnostic’ tab brings up a menu with the ‘Align’ button and a readout of the detector maximum and minimum voltage. The align button performs an automatic optimization of the mirror position. See Chapter 6, Troubleshooting > “Concentration Errors”. 3-12 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 59: Result

    The Valid Data Settings section transmits a TTL signal that describes the status of the instrument on any of the digital output lines. The Configure section allows Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 3-13...
  • Page 60: Configure

    The DAQ Board specifies the National Instruments device number assigned to the current loop board. Value 1 must be specified. The AutoStart feature automatically begins data transmission when the current loop program is opened. 3-14 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 61: Analog Output Board

    0–20 mA. Table 3–1. Analog Current Channel Assignments Compound Channel Sulphur dioxide Nitric oxide Nitrogen dioxide Nitrous oxide Ammonia Hydrogen chloride Hydrogen fluoride Carbon monoxide Carbon dioxide Water Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 3-15...
  • Page 62 MODBUS master to ensure that it matches the register number on the instrument. ▲ Note For additional information on how to read registers and interpret the data, refer to Appendix B, “Modbus Protocol”. ▲ 3-16 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 63 Modbus registers from the concentration file data. The nominal value is 10000 ms. The Settings button can be used to configure the COM port associated with the Modbus network for serial Modbus. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 3-17...
  • Page 64 Windows. If IMACC Modbus is added to the Windows Startup folder and the AutoStart capability is enabled, the program will automatically open and start upon bootup. 3-18 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 65: Chapter 4 Calibration

    ~10 minutes, a new background spectrum is collected and compared with the previous one. A good newly acquired background spectrum is saved for subsequent sample spectrum processing. If the new zero background Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 66: Periodic System Zero Check

    PLC to perform system daily zero check. To ensure a sufficient zero air flow into the sample probe, it would be necessary to adjust the rotameter, shown in Figure 4–1. Figure 4–1. Adjusting the Rotameter on the Pneumatic Panel Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 67: Periodic System Span Check

    The only difference is that the Gas Procedure system operator needs to swap the span gas cylinder with the audit gas source (dry gas cylinder or wet/hot HovaCal output). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 68: Calibration Gases

    0.005 mol/l In addition to a HovaCAL, a high concentration gas cylinder for HCl (~1000 ppm) is recommended. A gas calibrator such as Thermo Scientific Model 146i should be used to dilute HCl to the desired value for both span check and spike check.
  • Page 69: Manual Background Spectrum Collection Procedure

    When running IMACC Monitor, press Stop Script and press OK go to tab OMNIC. Spectrum Collection Procedure In OMNIC, select Exp Set, tab Collect, and set the following. No. of scans: 30, Resolution: 0.5, Final Format: Interferogram, Correction: None. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 70 400, Gain: 1.0, Velocity: 0.3165, Aperture: 15. Press OK. Then go to Col Smp (Collect Sample). Set the name of the new background in the following format: zero air background and date (year-mm-dd) Press OK. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 71 Make sure the system is flushed with zero air (wait 5 min) then press OK again. The system will now collect 30 scans when prompted with “Add to IMACC capture?” press Yes. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 72 Select tab Process and select Reprocess…. Verify or set the following. Resolution: 0.5, Apodization: Happ-Genzel, Phase correction: Mertz, Zero filling: None, Final format: SingleBeam, Correction: None, Saved spectral range: 4500 – 400. Press OK. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 73 Select tab File and select Save As. Go to directory BKGS and press Set Filename to Title. Go to tab My Documents (If not opened yet then open it with Explorer). Move previous background file to monthly archive and delete BKG. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 74: Manual Background Spectrum Collection Using Thermo Modbus Controller

    “save background” mode. The system starts to collect the spectrum and saves it as “c:\backgrounds\bkg.spa. After completion, switch the system command to “sample”. 4-10 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 75: System Calibration Procedure Using Imacc Collect

    Scan time: 0:56, Low Cutoff: 400, High Cutoff: 4500. Use Method Shift Settings: Not enabled; Use Dynamic References: Enabled; Use Bias Correction: Not enabled; Use Dynamic Shifting: Enabled; Forget Analysis State: Enabled Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 4-11...
  • Page 76 Bench Parameters: Enabled Pathlength: 5.209, Source: IR, Aperture: 15.0, Velocity, 0.2659, Gain: 4.0, Resolution: 0.5, Correlation: Simple, Default Temperature: 185, Default Pressure: 750 Press OK to close Collect EZ Setup box. 4-12 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 77 Optical Path: FrontMain, Type of data: IEEE float, Code: CellPressure, Register:0x0a2a, COM Port: 1, Parity: None, Baud: 9600, Stopbits: 1, Databits: 8, Flow Control: None Press OK to close ModbusMultiSensor Settings box. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 4-13...
  • Page 78 Press map to check that all compounds are listed. If not, enter the required compound and click the add button. Check that the corresponding compounds in the method are selected for the highlighted compound. 4-14 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 79 Scan Time (MM:SS): 01:00; Averaging Count: 3 Press GO. Upon completion of the process, a new calibration factor will be calculated. Then press “Save Factor to Method File” button to save the calibration factor. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 4-15...
  • Page 80 (without saving changes again). The reading should equal the value of the calibration gas (spike value). When you return to ImaccMonitor, the reading should equal the value of the calibration gas. 4-16 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 81 Calibration System Calibration Procedure Using IMACC Collect When pressing Start Script, the new zero air background will now be used. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 4-17...
  • Page 82: System Calibration Procedure Using Calibration Factor Table In Method

    Click the “linearity” tab and you will see the tab button “Calibration Factors”. Click the button and open the table as shown below. 4-18 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 83 O value should be calibrated with dry certified gases. For those with an H O content range, such as HCl_wet, the calibration must be done in that wet condition. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 4-19...
  • Page 84: Table 4-1. Calibration Factors

    >0.5% NH3_wet 0 to 106 ppm >0.5% NH3_dry 0 to 106 ppm <0.5% NO_dry_Hi 60 to 1253 ppm <1% NO2_dry 1 to 50 ppm <1% 0 to 6 ppm 4-20 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 85: Chapter 5 Preventive Maintenance And Servicing

    “Replacing the Gas Cell Heater” on page 5-51 ● “Replacing the Heated Lines” on page 5-54 ● “Replacing the PC” on page 5-55 ● “Replacing the PC Console” on page 5-57 ● Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 86 ● “Electrical Panel Components” on page 5-71 ● “Sample System Heater Assembly Components” on page 5-76 ● “Pneumatic Panel Components” on page 5-90 ● “Service Locations” on page 5-97 ● Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 87: Replacement Parts List

    FTIR Purge Gas Generator Maintenance Kit (kit contains 113309-00 1 year one BX instrument air filter) DX Instrument Air Filter (box of 10) 9810 9-12 months BX Instrument Air Filter (box of 10) 9811 9-12 months Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 88 Assembly, Replacement, Heated Line (FTIR to Sample 113390-00 System) USB Cable, 10' (FTIR to PC) 112796-00 Temperature Controller Part Number Replacement Period Temperature Controller Assembly 112701-00 Relay, SSR, Crydom 2425 103344-00 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 89: Table 1-4. Power

    Solid State Relay, DIN Rail, 5VDC 112224-00 Digital I/O 100 Pin to 32 Terminal Breakout PCBA 112165-00 Analog I/O 68 Terminal Breakout PCBA 109665-00 Cable, Temperature Alarm, 10' (Temp Controller to 110252-00 Electrical Panel) Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 90: Safety Precautions

    Important Do not attempt to clean or even touch the mirror surfaces inside the analyzer. These front-surfaced mirrors are easily scratched. Dust will not harm the mid-infrared signal, but fingerprints and scratches can degrade spectral performance or permanently damage the mirrors. ▲ Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 91: Maintenance For The Sample Probe

    Graphite filter seals (as required) CAUTION High surface temperatures present. Allow sufficient time to cool. ▲ 1. Flush the probe with inert gas or air to clear any toxic gases. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 92: Figure 5-1. Probe Assembly

    Remove the filter element, the 3 graphite gasket filter element seals, the retaining washer, and the graphite gasket lid seal from the assembly. Properly dispose of the used filter element and the graphite gasket lid seal (Figure 5–2). Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 93: Figure 5-2. Filter Housing Lid Assembly

    Turn the “T” handle approximately 1 turn clockwise or until snug. 11. Re-install the probe protection shield onto the probe. Close the two pressure clamps to secure the shield to the probe assembly. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 94: Replacing The Pre-Filter

    The pre-filter could be cleaned with water in an ultrasonic bath and then re-used. If doing so, perform several blow backs on the system after installation is complete to dry the pre-filter. 5-10 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 95: Maintenance And Servicing Of The Ftir Analyzer Components

    WARNING The desiccant is harmful if swallowed. Follow the manufacturer’s instructions to dispose of desiccant packets properly. ▲ CAUTION Avoid burn and fire hazards. Do not attempt to regenerate spent desiccant packets. ▲ Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-11...
  • Page 96: Figure 5-3. Locating The Humidity Indicator

    Be careful not to unplug or damage any cables connected to the panel (Figure 5–4). 5-12 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 97: Replacing The Humidity Indicator

    1. Using the Phillips screwdriver, remove the FTIR cover panel in the front of the system rack. Slide the FTIR drawer out of the rack to gain access to the analyzer. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-13...
  • Page 98: Cleaning Or Replacing The Fan Filter

    Slide the FTIR drawer out of the rack to gain access to the analyzer. 2. Using the flat blade screwdriver, carefully pry the fan cover grill off of the fan filter housing. Remove the dirty filter (Figure 5–5). 5-14 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 99: Figure 5-5. Removing The Fan Filter

    Whenever these procedures are performed, there is a potential Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-15...
  • Page 100: Replacing The Source

    1. Using the Phillips screwdriver, remove the FTIR cover panel in the front of the system rack. Slide the FTIR drawer out of the rack to gain access to the analyzer. 5-16 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 101: Figure 5-6. Removing The Right Side Panel

    5. Loosen the thumbscrew that secures the source assembly to the analyzer. Rotate the clamp bar that holds the source in place 90 degrees counterclockwise (Figure 5–7). Figure 5–7. Removing the Source Assembly Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-17...
  • Page 102: Figure 5-8. Source Element Is Inside Can Cut-Out

    13. Wait 15 minutes for the new components to stabilize thermally and then align the analyzer by following the instructions in Chapter 3. (See “OMNIC” software on page 3-11.) 5-18 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 103: Replacing The Laser

    If more room is needed, use the 9/16-inch wrench to loosen the nuts to disconnect the heated lines from the sample system heater assembly and/or the FTIR Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-19...
  • Page 104: Figure 5-9. Locating The Laser Assembly

    2 springs and an alignment pin. Release the spring clamps by pushing down on the clamp release hooks near the bottom of the analyzer compartment (Figure 5–10). 5-20 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 105: Figure 5-10. Unlocking The Laser Spring Clamps

    If the laser is misaligned, the analyzer will not scan. ▲ 9. Place the springs over the laser tube and re-attach the clamp release hooks. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-21...
  • Page 106: Replacing The Detector

    Important Note A system operator can replace the laser, however it is recommended that only our certified service engineers or on-site maintenance personnel who have been trained by us replace the laser. 5-22 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 107 Lift the detector straight up to disengage it from the connector and to clear the 2 alignment pins. Remove it from the analyzer (Figure 5–12). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-23...
  • Page 108: Figure 5-12. Locating The Detector Assembly

    6. Install the new detector assembly in the analyzer over the 2 alignment pins, taking care to engage the connector. Tighten the thumbscrews on the base plate until snug. 5-24 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 109: Replacing The Beamsplitter

    Whenever the analyzer compartment is open, there is a potential for damage to sensitive components, which will affect the performance of the system. ▲ Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-25...
  • Page 110: Figure 5-14. Ftir Wiring And Tubing Connections

    3. Using the small Phillips screwdriver, loosen the captive screw that holds the pressure sensor cable in place, then pull the cable assembly off of the sensor (Figure 5–14). Figure 5–14. FTIR Wiring and Tubing Connections 5-26 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 111: Figure 5-15. Corner Screws On Coupling Optics Plate

    9. Locate the rectangular top access panel toward the rear of the analyzer. Using the 5/32-inch hex wrench, remove the 4 corner mounting screws and remove the top access panel (Figure 5–16). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-27...
  • Page 112: Figure 5-16. Removing The Beamsplitter Access Panel

    12. Place the rectangular top access panel back into position. Using the 5/32-inch hex wrench, re-install the 4 corner mounting screws and tighten until snug. 13. Follow steps 2–8 in reverse to complete the installation procedure. 5-28 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 113: Replacing The Air Filters

    2. Grasp the bottom of the filter assembly, then with a slight upward pressure, turn it to the left until it stops (the word “Lock” should be directly facing you) and pull it straight down (Figure 5–19). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-29...
  • Page 114: Figure 5-19. Removing Filter Cap

    4. Take the replacement filter and screw it onto the threaded rod. Once some resistance is felt, turn it approximately another 1/2 turn until it is snug (Figure 5–21). 5-30 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 115: Removing The Purge Gas Generator

    Large Phillips screwdriver 1. Note the pressure on the house air regulator, then pull up on the knob to unlock it and turn the pressure down to zero (Figure 5–22). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-31...
  • Page 116: Figure 5-22. Removing The Purge Gas Generator

    5. Using the large Phillips screwdriver, loosen but do not remove the 4 mounting screws (Figure 5–22). 6. Lift the purge gas generator up and off of the 4 mounting screws (Figure 5–22). 5-32 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 117: Removing The Left Side Panel

    Simply grasp the hand grips on the panel and carefully lift it about an inch, then pull the panel away from the rack to remove it (Figure 5–23 Figure 5–24). Figure 5–23. Lock and Hand Grips Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-33...
  • Page 118: Removing The Ftir

    Panel” for better accessibility (optional step). 3. Using the Phillips screwdriver, remove the FTIR cover panel in the front of the system and the blank panel above it (Figure 5–25). 5-34 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 119: Figure 5-25. Remove Screws And Two Lowest Panels

    Figure 5–25. Remove Screws and Two Lowest Panels 4. Slide the FTIR drawer out of the system to access the rear of the unit (Figure 5–26). Figure 5–26. FTIR Shelf Out Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-35...
  • Page 120 12. Using the 9/16-inch wrench, disconnect the two 1-meter heated lines from the elbow fittings at the rear of the unit (Figure 5–27). Note or mark their position to facilitate re-connection. 5-36 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 121: Figure 5-27. Ftir Wiring And Tubing Connections

    14. Using 2 people, carefully lift the FTIR unit off of its base. For details, see “Lifting” on page 2-1. 15. Re-install the FTIR by following the previous steps in reverse order. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-37...
  • Page 122: Replacing The Ftir Power Supply

    2. Using the Phillips screwdriver, remove the 8 mounting screws and lift the panel off of the base (Figure 5–29). Figure 5–29. FTIR Power Supply Replacement 3. Disconnect the power cord from the back of the power supply. 5-38 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 123: Replacing The Temperature Sensor

    4. Slide the FTIR drawer out of the system to access the rear of the unit. 5. At the rear of the FTIR cell, disconnect the temperature sensor connector from the sensor by pulling it off, exposing the 3 metal prongs (Figure 5–30). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-39...
  • Page 124: Figure 5-30. Ftir Cell Rtd Temperature Sensor Replacement

    10. Re-install the sensor connector onto the 3 metal prongs of the sensor, with the red dots lining up with each other. 11. Follow steps 2–4 in reverse order to complete the procedure. 5-40 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 125: Replacing The Pressure Sensor

    4. Using the small Phillips screwdriver, loosen the captive screw that holds the pressure sensor cable in place, then pull the cable assembly off of the sensor (Figure 5–31). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-41...
  • Page 126: Replacing The Gas Cell

    Gas Cell preferably by a trained service technician. Equipment Required: Wrench, 9/16-inch, 7/16-inch, and 3/4-inch Hex wrench, 5/32-inch and 7/64-inch Phillips screwdriver Gas cell 5-42 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 127 27, but with the tubing removed with the sensor.) 9. Using the 5/32-inch hex wrench, remove the 4 cap screws that mount the gas cell to the mounting plate (Figure 5–32). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-43...
  • Page 128: Gas Cell Alignment

    Wrench, 9/16-inch and 11/32-inch Hex wrench, 5/32-inch and 9/64-inch Phillips screwdriver Visible laser module for gas cell alignment (field service only) Refer to “Replacing the Gas Cell” for additional tools needed 5-44 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 129: Figure 5-33. Gas Cell Mirror Configuration

    4. Using the 5/32-inch hex wrench, remove the 4 screws that mount the nickel plated aluminum cell to the plate. Carefully remove the gas cell casing, leaving the mirrors in place (Figure 5–33). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-45...
  • Page 130: Figure 5-34. Optical Layout In The Relay Mirror Compartment

    “Open area 1” where the IR light comes out of the bench illustrated in Figure 5–36. Figure 5–35. Visible Laser Module for Gas Cell Alignment 5-46 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 131: Figure 5-36. Visible Laser Used To Align The Gas Cell

    Mirror #1 is used to adjust the beam image on mirror #2, mirror #2 is used to adjust the beam image on mirror #3, and mirror #3 is used to position the beam image on the objective Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-47...
  • Page 132: Figure 5-38. Typical Image Pattern On The Field Mirror

    Figure 5–38. Typical Image Pattern on the Field Mirror Figure 5–39. Location of Components on Back Plate of Gas Cell 5-48 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 133 For the 5.2 meter gas cell, the maximum number of passes should be 20, with the number of 5 images on the bottom row as shown in Figure 5–40. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-49...
  • Page 134: Figure 5-40. Real Red Laser Image Pattern

    4, 5 and 6 shown in Figure 5–34. Manipulate these mirrors to maximize the IR signal (Max) on the interferometer display as shown in Figure 5–41. Figure 5–41. Typical Interferometer of Omni FTIR CEMS 5-50 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 135: Replacing The Gas Cell Heater

    Use the following procedure to replace the FTIR gas cell heater. Gas Cell Heater Equipment Required: Wrench, 9/16-inch, 7/16-inch, and 3/4-inch Hex wrench, 5/32-inch and 7/64-inch Phillips screwdriver Razor blade/utility knife Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-51...
  • Page 136: Figure 5-42. Cutting Insulation Paper At The Seam

    (Figure 5–42). Figure 5–42. Cutting Insulation Paper at the Seam 4. Remove the insulation from the gas cell. Wearing gloves is suggested when handling the insulation (Figure 5–43). 5-52 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 137: Figure 5-43. Insulation Removal

    8. Place the insulation back on the gas cell and attach it with a couple of pieces of high temperature tape, making sure that it fits snugly. 9. Refer to the procedure for “Replacing the Gas Cell” to re-install the gas cell. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-53...
  • Page 138: Replacing The Heated Lines

    7. Using the 9/16-inch wrench, loosen the nut and disconnect the heated line(s) to be replaced from the elbow fittings at the rear of the FTIR (Figure 5–45). Note their position to facilitate re-connection. 5-54 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 139: Replacing The Pc

    9. Install the new heated line(s) by following the previous steps in reverse order. Attach insulation and tie wraps where appropriate. Replacing the PC Use the following procedure to replace the PC. Equipment Required: Phillips screwdriver Flat blade screwdriver Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-55...
  • Page 140: Figure 5-46. Pc Removal

    Figure 5–46. PC Removal 5. Slide the PC console back into the rack. Depress the tab at the rear of each slide rail of the PC console to unlock it (Figure 5–46). 5-56 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 141: Replacing The Pc Console

    (Figure 5–46). 3. At the rear of the system, unplug the PC console power connector, as well as the connectors for the console, mouse, and keyboard (Figure 5– 47). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-57...
  • Page 142: Figure 5-47. Pc Console Removal

    8. Push the PC console in just enough to get the rear rails on each side started in the rack sliders. Once both sides are in place, push the console in a couple of inches (Figure 5–48). 5-58 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 143: Replacing The National Instruments Boards

    A properly grounded antistatic wrist strap must be worn while handling any internal component. If an antistatic wrist strap is not available, be sure to touch the instrument chassis before Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-59...
  • Page 144: Figure 5-49. Removing The Pc Cover

    (Figure 5–50). Note that 4 of the screws are on the back of the PC and 1 is inside. Figure 5–50. Removing the Card Cage Mounting Screws 5-60 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 145: Figure 5-51. Positioning The Card Cage

    (See Figure 5–52 Figure 5–53 that show the Digital Board.) Figure 5–52. Removing NI Board Mounting Screw Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-61...
  • Page 146: Temperature Control Box Components

    Box Components Replacing External Use the following procedure to replace the 2 external fuses on the FTIR temperature control box. Fuses Equipment Required: Phillips screwdriver Flat blade screwdriver Replacement Fuses 5-62 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 147: Figure 5-54. External Fuse Replacement

    5. The external fuse holders are located under the above cable connectors. To remove the fuses, push the holder in while turning it approximately 1/4 turn counterclockwise and then pull it out (Figure 5–55). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-63...
  • Page 148: Replacing Internal Components

    3. To remove the cover, use the small Phillips screwdriver to remove the 8 screws that mount the cover to the temperature control box and lift the cover off. 5-64 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 149: Replacing The Internal Fuses

    2. There are 16 fuses located on the PC Board inside the temperature control box (plus 4 spare fuses.) To remove any one of them, simply grasp the fuse with your fingers and pull it up and out of the connector. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-65...
  • Page 150: Replacing The Temperature Control Board

    Use the following procedure to replace the temperature control board Temperature Control inside of the temperature control box. Board Equipment Required: Phillips screwdriver Small Phillips screwdriver Temperature Control Board 5-66 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 151 7. Re-connect all of the cables to their proper locations on the temperature control board. 8. Re-install the temperature control box cover per the above procedure “Removing and Installing the Cover”. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-67...
  • Page 152 6. Pop the end covers up on the new relay and place it in the temperature control box over the mounting studs in the proper orientation. Thread on the 2 mounting nuts, and then use the 5/16-inch nut driver to tighten them until snug. 5-68 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 153: Replacing The Temperature Control Modules

    5–55). Using the 3/8-inch nut driver, remove the 2 nuts that mount the control module and base unit assembly to the temperature control box. Note the orientation of the assembly. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-69...
  • Page 154: Replacing The Temperature Control Base Unit

    A properly grounded antistatic wrist strap must be worn while handling any internal components. ▲ 1. Remove the temperature control box cover per the previously mentioned procedure “Removing and Installing the Cover”. 5-70 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 155: Electrical Panel Components

    11. Re-install the temperature control box cover per the above procedure, “Removing and Installing the Cover”. Electrical Panel Use the following procedures to replace the electrical panel components (Figure 5–56 Figure 5–57). Components Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-71...
  • Page 156: Figure 5-56. Electrical Panel Component Location (Front Side)

    Preventive Maintenance and Servicing Electrical Panel Components Figure 5–56. Electrical Panel Component Location (Front Side) Figure 5–57. Electrical Panel Component Location (Rear Side) 5-72 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 157: Replacing The Ethernet Switch

    Use the following procedure to replace the solid state relays on the electrical State Relays panel in the rear of the system. There are 3 24VDC relays and 15 5VDC relays (Figure 5–57). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-73...
  • Page 158 Be sure that the voltage of the new relay matches the one being replaced. 8. Re-install the electrical panel by following steps 1–4 in reverse order. 5-74 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 159: Replacing The 24V Power Supply

    7. To install the new power supply, hook the upper tabs of the rear of the power supply onto the din rail, and then firmly push the bottom of the supply toward the panel until it locks into place. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-75...
  • Page 160: Sample System Heater Assembly Components

    They are the instrument zero air, the flow back air, the heated sample line, and the auxiliary sample line (if equipped). Note or mark the location of each line to facilitate reconnection. 5-76 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 161: Figure 5-58. Sample System Heater Assembly And Cover Removal

    “Pump”. Turn the connector nut counterclockwise and pull it out of its socket. Remove the cable assembly tie-wraps using the wire cutters. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-77...
  • Page 162: Replacing The Thermostat

    “Removing the Sample System Heater Assembly Cover”, skipping steps 5, 6, and 7 (sample system heater assembly doesn’t need to be removed from the rack for this procedure). 5-78 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 163: Replacing The Eductor Pump

    “Removing the Sample System Heater Assembly Cover”, skipping steps 6 and 7 (sample system heater assembly doesn’t need to be removed from the rack for this procedure). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-79...
  • Page 164: Figure 5-59. Sample System Heater Assembly Component Replacement And

    Ejector Air In and the FTIR Out (Figure 5– 59). 6. Pull the entire eductor pump plumbing assembly out of the sample system heater assembly box. 5-80 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 165 13. Follow steps 1–4 in reverse order to complete the procedure. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-81...
  • Page 166: Removing The Heated Block Assembly

    RTD is properly seated. 7. Using the 3/16-inch hex wrench, install and tighten the 4 mounting screws in the corners of the heated block assembly until snug. 5-82 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 167: Replacing The Check Valve

    2. Remove the heated block assembly from the heater box per the previously mentioned procedure, “Removing the Heated Block Assembly”. 3. Turn the heated block assembly around to access the back of the assembly. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-83...
  • Page 168: Figure 5-60. Check Valve Replacement

    8. Using the wrenches, connect the other end of the check valve to the brass nut with copper tubing and tighten until snug. Some movement of the copper tubing will be required to get it into the check valve. 5-84 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 169 13. Follow the procedures referenced in step 2, “Removing the Heated Block Assembly”, and in step 1, “Removing the Sample System Heater Assembly and Cover”, in reverse order to complete the procedure. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-85...
  • Page 170: Replacing The Sample Solenoid Valve

    4. Using the 9/64-inch hex wrench, loosen slightly but do not remove the 5 screws that mount the double coil clamp plate to the heated block. This will allow some freedom of movement of the copper tubing (Figure 5–60). 5-86 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 171: Figure 5-61. Solenoid Valve Replacement

    10. Lift the sample solenoid valve and plumbing assembly out of the heated block (Figure 5–61). Note the orientation of the tubing and fittings to facilitate reconnection. Figure 5–61. Solenoid Valve Replacement Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-87...
  • Page 172 16. Attach the other piece of copper tubing with brass nut to the union fitting on the adjacent side, but do not tighten it all the way (Figure 5– 61). 5-88 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 173 22. Follow the procedures referenced in step 2, “Removing the Heated Block Assembly”, and in step 1, “Removing the Sample System Heater Assembly and Cover”, in reverse order to complete the procedure. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-89...
  • Page 174: Pneumatic Panel Components

    “Removing the Left Side Panel” for better accessibility. 2. Note the pressure on the gauge above the pressure reducer to be replaced, so that it can be reset later. 5-90 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 175: Replacing The Pressure Gauges

    The procedure is the same for either pressure gauge. Pressure Gauges Equipment Required: Wrench, 9/16-inch Pressure gauge 1. Remove the left side panel per procedure “Removing the Left Side Panel” for better accessibility. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-91...
  • Page 176: Replacing The Solenoid Valves

    (from the rear of the panel, it is the furthest to the right.) (See (Figure 5–62).)Be sure to order the correct valve. Equipment Required: Phillips screwdriver Small Phillips screwdriver Wrench, 9/16-inch and 3/4-inch 5-92 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 177: Figure 5-63. Solenoid Valve Replacement

    4. Using the 11/32-inch nut driver, loosen and remove the 2 nuts that mount the valve bank assembly to the flow meter panel. Pull the valve bank assembly away from the flow meter panel to access the back side (Figure 5–62). Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-93...
  • Page 178: Replacing The Flow Meters

    The 4 center flow meters have the same part number, but the 2 end flow meters have a different part number. Be sure to order the correct flow meter. Equipment Required: Phillips screwdriver 5-94 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 179 7. Insert the new flow meter into position on the flow meter panel, making sure the orientation is correct. Insert the 2 mounting screws and tighten using the Phillips screwdriver until snug. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-95...
  • Page 180 12. Re-install the left side panel to complete the procedure. 5-96 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 181: Service Locations

    Contact one of the phone numbers Locations below for product support and technical information or visit us on the web at www. thermoscientific.com/aqi. 1-866-282-0430 Toll Free 1-508-520-0430 International Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 5-97...
  • Page 183 Chapter 6 Troubleshooting This chapter provides the following troubleshooting and service support information for the Omni FTIR Multi Gas CEMS: “Concentration Errors” on page ● “Boot Up Error/IMACC Software Error” on page ● “Temperature Errors” on page ● “Cell Pressure Errors/Sampling System Errors”...
  • Page 184: Troubleshooting

    Cell Pressure = <40 mmHg less than ambient pressure during ● sample mode The following outline addresses possible instrument and alarm problems and proposes checks and possible troubleshooting solutions. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 185: Concentration Errors

    Go to ‘C:\Backgrounds\’ in Windows Explorer and verify that ● file ‘C:\Backgrounds\BKG.spa’ has been updated in the last 24 hours. Refer to Appendix E, “IMACC Script Setting” for proper script configuration. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 186: Boot Up Error/Imacc Software Error

    ● temperatures are OK. Check thermostat inside pump box (See Chapter 5, “Preventive ● Maintenance and Servicing”. ERROR_QUANT_TEMPRESSCON in ImaccMonitor ● Check temp control breaker and AC wiring connections. ● Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 187: Cell Pressure Errors/Sampling System Errors

    Check for blocked exhaust system; make sure exhaust lines are ● less than 10 feet long. Check all temperature readings; make sure they are within 10 ● degrees of setpoints. Check probe breaker. ● Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 188: Communication Errors

    Check ‘Hardware’ setup in Windows system menu. ● Internal communications errors No temperature readings or heating of system components ● See ‘Temperature errors’ in this chapter. ● No communications with FTIR ● Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 189: I/O Errors

    Gas mode output or FTIR data valid digital output is not reporting ● the expected mode Click the “View Script” button in ImaccMonitor and verify ● that IMACC Script Editor shows that the script is running. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 190: Service Locations

    Contact one of the phone numbers Locations below for product support and technical information or visit us on the web at www. thermoscientific.com/aqi. 1-866-282-0430 Toll Free 1-508-520-0430 International Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 191 “Industrial PC” on page 7-13 ● “Console” on page 7-14 ● “Datalogger” on page 7-14 ● “Electrical Panel Assembly” on page 7-14 ● “AC Power Input” on page 7-15 ● Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 192: Chapter 7 System Component Description

    The standard probe has a 4-inch ANSI flange and a 1.0 meter probe barrel length. Other flange sizes and barrel lengths may be purchased depending on requirements. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 193: Temperature Controller

    15 Amps at 220-240VAC 50/60 Hz. The sample line temperature is monitored using a 2-wire 100 ohm RTD. Additional electrical lines are included to hook up the probe temperature alarm and blowback valve. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 194: Heated Pump/Valve Assembly

    FTIR Zero Air Flow (4.0 L/min) ● Sample Line Flowback Flow (4.0 L/min) ● Probe Zero Flow (6.0 L/min) ● Probe Span Flow (6.0 L/min) ● Probe Spike Flow (0.4 L/min) ● Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 195: Sample Mode

    This mode should be executed once every day for collecting instrument background spectrum. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 196: Check Zero

    6 LPM span gas into the stinger via the cal/zero port. This mode is executed for the daily system span check. Figure 7–4. Sample and Calibration Gas Flow Diagram Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 197: Ftir

    When the beams recombine, constructive or destructive interference occurs, depending on the relationship between the position of the moving mirror relative to the fixed mirror and the wavelength of the light. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 198: Figure 7-6. Ftir Flow Schematic

    Fourier Transform or FT. The function calculates the amplitude of each of the component signals. The amplitude gives the intensity at the corresponding wavelength of light. A Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 199: Optical Components

    Figure 7–7. Spectrometer Compartment Source The FTIR analyzer uses Thermo Fishser Scientific’s EverGlo source. The source unit can be replaced as needed. It is pre-aligned and needs no adjustment when it is installed. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 200: Interferometer

    The DTGS detector can accommodate most of the energy from the source, which means the software selects a large aperture size. 7-10 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 201: Detector

    Jacket accuracy and prevent condensation of sample gases. The heating jacket is designed to heat the gas cell to a temperature of 185 °C. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 7-11...
  • Page 202: Temperature Sensor

    1 ppm. The generator uses a combination of coalescing filtration, regenerative pressure swing adsorption, and high efficiency particulate filtration to produce laboratory quality, dry, CO -free air from a standard compressed air supply. 7-12 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 203: Industrial Pc

    Front-panel DVD+R drive, 3.5” floppy drive, and USB connectors are provided as additional means to move data off the system as well as install system updates. Figure 7–10. FTIR PC Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 7-13...
  • Page 204: I/O Boards

    Digital solid state relay outputs (Temperature Alarm, Pressure Alarm, ● Maintenance, FID Alarm, FTIR Data Valid, System Gas Mode) Digital Inputs (System Gas Mode) ● Probe Connections (AC Power, Temperature Alarm, and Valve ● Control) 7-14 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 205: Ac Power Input

    The system requires 60 Amp 220-240 VAC connections. This box includes screw terminals for these connections. Figure 7–12. AC Mains Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual 7-15...
  • Page 207: Appendix A Warranty

    (i) normal wear and tear, (ii) accident, disaster or event of force majeure, (iii) misuse, fault or negligence of or by Buyer, (iv) use of the Products in a manner for which Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 208 OR WRITTEN, WITH RESPECT TO THE PRODUCTS, INCLUDING WITHOUT LIMITATION ALL IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE. SELLER DOES NOT WARRANT THAT THE PRODUCTS ARE ERROR-FREE OR WILL ACCOMPLISH ANY PARTICULAR RESULT. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 209: Appendix B Modbus Protocol

    ImaccModbus uses two Modbus registers to transmit concentration and measurement data in floating point format. The TCP version also has the ability to accept an F06: Write Single Register command. For details of the Omni FTIR Multi Gas CEMS Modbus Protocol specification, see the following topics: “Serial Communication Parameters”...
  • Page 210: Application Data Unit Definition

    1 and 255 decimal (i.e. 0x01 hex to 0xFF hex). The default value is 70. MBAP Header In Modbus over TCP/IP, a Modbus Application Protocol Header (MBAP) is used to identify the message. This header consists of the following components: Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 211: Function Code

    For each register, the first byte contains the high order bits and the second contains the low order bits. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 212 0x64 The contents of register 9 are shown as the two byte values of 0x02 0x2B. Then contents of registers 10–12 are 0x00 0x00, 0x00 0x64 and 0x00 0x64, respectively. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 213: Modbus Parameters Supported

    Table B–1 through Table B–4 lists the Modbus addresses supported for the Omni FTIR Multi Gas CEMS. Parameters Supported IMPORTANT NOTE The addresses in the following table are Protocol Data Unit (PDU) addresses. For Application Data Unit (ADU) addresses, add 40000 to these numbers for holding registers.
  • Page 214: Table B-2. Modbus Write Registers

    This mode is used to check the system span CHECK SPAN2 MODE This mode is used to perform an extra span check, in (N/A) addition to the target gas. CHECK SPAN3 MODE (N/A) Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 215: Table B-4. System Flag Bits

    Modbus Protocol Modbus Parameters Supported Table B–4. System Flag Bits Value Description 0x00000000 Data Valid 0x00000001 Data Invalid Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 217: Appendix C Interfacing A Dcs/Plc With The Omni Ftir Multi Gas Cems

    Checking data values Reading system and alarm status Triggering events (such as calibration checks and probe blowbacks) The Omni FTIR Multi Gas CEMS provides several methods for establishing data communication with a DCS/PLC including: Analog output connections (4-20mA) Digital I/O connections (5V TTL logic)
  • Page 218: Connecting A Dcs/Plc To The Omni Ftir Multi Gas Cems

    Interfacing a DCS/PLC with the Omni FTIR Multi Gas CEMS Connecting a DCS/PLC to the Omni FTIR Multi Gas CEMS Connecting a You can connect to the FTIR via Serial or Ethernet communications. Refer Figure C–1 and the “Installation” chapter. For Modbus over Ethernet, DCS/PLC to the use TCP port 502 by default (user selectable).
  • Page 219: Establishing Communications

    Interfacing a DCS/PLC with the Omni FTIR Multi Gas CEMS Establishing Communications Establishing This section provides information on the methods available for retrieving system data, accessing system status conditions, initiating system events, Communications and checking system calibration. Analog Outputs The 4-20mA analog output signals are labeled “ANALOG OUTPUT X1”...
  • Page 220: Digital I/O

    Interfacing a DCS/PLC with the Omni FTIR Multi Gas CEMS Establishing Communications Figure C–2. Analog Outputs Digital I/O The FTIR allows the user to access system status conditions via the digital I/O connections. The digital I/O signals are labelled “DIGITAL OUTPUT X2”...
  • Page 221: Figure C-3. Digital I/O - Rear Panel Terminals

    Interfacing a DCS/PLC with the Omni FTIR Multi Gas CEMS Establishing Communications Table C–2. Digital I/O Connections Digital No. Block Terminal Description Digital Input Ground Gas Mode Input: Check Span Gas Mode Input: Check Spike Gas Mode Input: Check Zero Gas Mode Input: Bkg &...
  • Page 222: Solid-State Relays (Ssr

    Interfacing a DCS/PLC with the Omni FTIR Multi Gas CEMS Establishing Communications Solid-State Relays TTL outputs from the National Instruments 6509 board are used to drive solid state relays (SSR) for valve control and digital status outputs. The (SSR) SSRs are mounted on a DIN rail on the interior of the electrical panel (Figure C–4).
  • Page 223: Appendix D Calogix Setup

    The USB cable should always be connected to the computer at the lower-left USB socket (Figure D–2). The hardware is also connected to a 220 VAC supply from circuit breaker panel. Figure D–1. Base Unit and Four Controllers Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 224: Finding The Correct Com Port

    Use the following procedure to determine the COM port number. 1. Turn the power ON to the Temperature control box. 2. Select “Start” menu and open ‘Control Panel’. Run ‘System’. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 225 Calogix Setup Finding the Correct COM Port 3. Select the Hardware tab in System Properties. 4. Select ‘Device Manager’. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 226: Setting The Virtual Com Port

    Use the following procedure to set the virtual COM port. Virtual COM Port 1. Open CALGrafix software with the icon on the desktop. 2. Click the second icon from the left in the icon bar. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 227 Calogix Setup Setting the Virtual COM Port 3. Click ‘Browse’ in the ‘Select Devices’ popup menu. 4. Select the correct number for the COM port, Select ‘OK’. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 228: Programming The Controllers And Setting Parameters

    ‘On’ before the controllers are set to run. If the controllers are set to run with the ramp program not enabled, heated line and sample cell temperatures could rise too quickly and damage those components. ▲ Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 229: Creating Ramp Programs

    None. Enter Segment 2: Segment type SOAK, Soak time Continuous, EOP Status None. 3. Select the ‘Module 2’ tab. Enter Segment 1: Segment type RAMP, Ramp rate 90, Target Setpoint 185, Holdback band 0, EOP Status Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 230 Ramp rate 90, Target Setpoint 185, Holdback band 0, EOP Status None. Enter Segment 2: Segment type SOAK, Soak time Continuous, EOP Status None. 5. Close the program menu. The changes will be automatically written to the controllers. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 231: Inputting Controller Values

    2. Once a module is selected, enter the correct values according to Table D–1. Values are grouped according to subheadings. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 232: Table D-1. Controller Values

    Output1 Inversion FALSE FALSE FALSE FALSE Output2 Inversion FALSE FALSE FALSE FALSE Output3 Inversion FALSE FALSE FALSE FALSE Output1 Inhibit None None None None Channel 1 Control Setpoint Mode PARK D-10 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 233 PARK Subsidiary Mode None None None None Setpoint Band Cycle Time Maximum Power Minimum Power Output Power Manual Power Display Alarm Reset Latched Alarm Programmer Program number Run mode Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual D-11...
  • Page 235: Appendix Eimacc Script Setting

    (“Sample Mode”) and one main setting block on the right (“System Running Status”). The vertically aligned sub setting blocks contained in the main script setting body are constructed in a sequential loop. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 236 Status” setting. This setting is used to log six monitored parameters into the “cons” files. They are gas cell temperature, sample temperature, short inlet sample line temperature, heated box temperature, short outlet sample line Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 237 “Register” context is determined by Calogix setting (0x0a1c, in this case). Go back to the “Modbus Register Setting” panel, click the “serial” button on the top, and view the following setting panel, as shown. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 238 High Cutoff” define the low end and high end of the Mid IR spectrum range. They are not recommended to be adjusted by users. The background spectrum used for the absorption spectrum conversion is saved in the Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 239 0.25 ppm is applied in the script setting for HCl measurement. If the users need to make an adjustment, click the “Calculating Setting” block, right below the “Collect EZ” setting block. Its setting subpanel pops up, as shown. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 240 3:15 am. If a change is needed, get access to the script block, labeled “Background Spectrum Collection”. Click the green shadow block “ifTimeCondition: equal to ==”, and the subpanel “time setup” pops up. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 241 In general, one daily background collection is adequate. Script setting for background spectrum collection ● Double click the script setting block named “Background Spectrum Collection”so its setting panel pops up, as shown. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 242 “C:\Backgrounds\BKG.spa”. Right after the calculation setting is another 5 second time delay setting. At the end of the block is the solenoid valve operation setting for normal sample mode. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 243 00:30:00. Users can adjust the blow back frequency by clicking the “add” and “remove” button. The time window for each blowback action is set as 5 minutes. Right below the schedule setting is the script for the blowback valve operation. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 244: Modbus Control Version

    The Modbus based script named Modbus_control is saved in the script folder of the directory of c:\scripts. Use script edit to open it so that its setting block pops up. E-10 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 245 “alarm” would be flagged. Logically, there would be three alarm output results based on the system temperature status. The validation of CEMS measurement data is determined by the variable Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual E-11...
  • Page 246 “system_cmd” to the variable of “system_status”. The second script setting is used to assign the digital I/O output channels for each system mode using the “system_status” variable. E-12 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 247 “dev2/port1/line0:7”, pattern “0x050” and time delay setting “0”. Script setting for ModbusMultiSensor ● Next to the script setting for system command and status is the container for “ModbusMultiSensor” setting, as shown. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual E-13...
  • Page 248 Under the “setup” configuration, the setting panel is similar to the figure below. Highlight the target parameter, such as SampleT, in the list. Double click it so its subpanel pops up, as shown. E-14 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 249 “ModbusMultiSensor” container. As shown, it is the same as described in the CEMS_Samplemode section. Note that all the settings should be the same as the one included in the background spectrum collection. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual E-15...
  • Page 250 The first setting block is used to define the file directory for all saved data. The second setting block is used to display real time system status on the monitor. E-16 Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 251: Appendix F Concentration Data Transfer Using A Web Server

    Do not uncheck any of the existing components or you will uninstall them. 4. Click Next to begin installation. 5. Allow Windows to reboot after installation is complete. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 252 6. Enter FTIR Data into the Alias window and click Next. 7. Browse for the data transfer folder and click Next. 8. Check Read and Run scripts, and click Next and Finish to complete the setup. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
  • Page 253 4. In the entry box Copy CON file to directory, browse for the data transfer folder created earlier. IFSS will save a copy of the current concentration file to the file current.con in that folder. 5. Click OK and save the script. Thermo Fisher Scientific Omni FTIR Multi Gas CEMS Instruction Manual...
  • Page 254 Access to the data will now be restricted to users who possess valid usernames and passwords for the FTIR computer. In this case, the software written to use IIS access will also have to provide this additional information. Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...