Thermo Scientific Omni FTIR Multi Gas CEMS Instruction Manual page 247

Table of Contents

Advertisement

IMACC Script Setting
Modbus Control Version
For each system mode, the corresponding system control is configured by
the third script setting. The description for each system mode is
summarized in
Table B–3
in Appendix B,
"Modbus
Protocol".
Each mode involves a couple of valve operations. Follow the similar
procedure as described in the section of CEMS_samplemode to set the
valve control setting. For instance, to configure the blowback valve to open,
just double click its setting block and set the correct DIO channel
"dev2/port1/line0:7", pattern "0x050" and time delay setting "0".
Script setting for ModbusMultiSensor
Next to the script setting for system command and status is the container
for "ModbusMultiSensor" setting, as shown.
Thermo Fisher Scientific
Omni FTIR Multi Gas CEMS Instruction Manual
E-13