IMACC Script Setting
CEMS_SampleMode
E-6
Omni FTIR Multi Gas CEMS Instruction Manual
Users can edit the calculation expression displayed in the expression box. In
this example, 0.3 ppm is the zero offset. Type in the new zero offset value,
and click the "OK" button to save it. The zero offset corrected output
results will be saved as "HCl_offset" in the "con" file.
Schedule setting for auto zero background collection
●
By default, the zero background collection is executed daily, starting at
3:15 am. If a change is needed, get access to the script block, labeled
"Background Spectrum Collection".
Click the green shadow block "ifTimeCondition: equal to ==", and the
subpanel "time setup" pops up.
Thermo Fisher Scientific