Removing The Ftir; Figure 5-24. Posts And Keyhole Slots - Thermo Scientific Omni FTIR Multi Gas CEMS Instruction Manual

Table of Contents

Advertisement

Preventive Maintenance and Servicing

Removing the FTIR

Removing the
5-34
Omni FTIR Multi Gas CEMS Instruction Manual
Figure 5–24. Posts and Keyhole Slots
2. To re-install, lift the panel and line up the posts to the 6 keyhole slots.
Push the panel against the rack until it is flush using the weight of the
panel to drop it into place.
Use the following procedure to remove the FTIR from the system.
FTIR
Equipment Required:
Wrench, 9/16-inch and 7/16-inch
Phillips screwdriver
Small Phillips screwdriver
CAUTION High surface temperatures present. Allow sufficient time to
cool.
1. Shut down the FTIR and heaters by turning off the FTIR circuit
breaker and the Temperature Control circuit breaker on the breaker
box. Allow it to cool for 1 hour. Turn off all calibration gases and the
air regulator on the right side of the system. Note the pressure so it can
be reset when completed.
2. Remove the left side panel per procedure
Panel" for better accessibility (optional step).
3. Using the Phillips screwdriver, remove the FTIR cover panel in the
front of the system and the blank panel above it
"Removing the Left Side
(Figure
5–25).
Thermo Fisher Scientific