System Installation, Start Up and Shut Down

System Shut Down

System Shut
2-14
Omni FTIR Multi Gas CEMS Instruction Manual
To shut down the system, switch the Measurement/Maintenance switch to
the Maintenance position.
Down
Turn off the power to "Heater"; this will switch the system to fail safe
mode.
Wait 5 minutes to flush the complete system.
Close all Programs running on the PC.
Use the 'Startup' menu of Windows to switch off the power on the PC.
Turn all other circuit breakers OFF one by one, and then turn off the main
breaker on the breaker panel.
Note Once the initial startup procedure has been done, remember to turn
the Maintenance switch to the "Measurement" position each time the
system is powered up.
Thermo Fisher Scientific