Creating Ramp Programs

Thermo Fisher Scientific
Use the following procedure to create ramp programs.
1. Right-click anywhere on the CALogix window. Select 'Edit Programs'
from the pop-up menu.
2. Select the 'Module 1' tab. Enter Segment 1: Segment type RAMP,
Ramp rate 90, Target Setpoint 185, Holdback band 0, EOP Status
None. Enter Segment 2: Segment type SOAK, Soak time Continuous,
EOP Status None.
3. Select the 'Module 2' tab. Enter Segment 1: Segment type RAMP,
Ramp rate 90, Target Setpoint 185, Holdback band 0, EOP Status
Programming the Controllers and Setting Parameters