Troubleshooting
Service Locations
Locations
6-8
Omni FTIR Multi Gas CEMS Instruction Manual
●
Service
For additional assistance, Thermo Fisher Scientific has service available
from exclusive distributors worldwide. Contact one of the phone numbers
below for product support and technical information or visit us on the web
at www. thermoscientific.com/aqi.
Check that the SYSTEM.STATUS value is showing the
expected mode; see
pins should have a continuous DC voltage supplied (Max 24
VDC).
Check wiring from terminal block to breakout board on
●
opposite side of electrical panel.
Check cable from breaker board to NI PCI-6709 board inside
●
PC.
Maintenance digital output is not reporting the expected mode
Check the key switch position. The vertical position is
●
maintenance mode and voltage should be conducted from the
odd terminal 25 to the even terminal 26.
1-866-282-0430 Toll Free
1-508-520-0430 International
Table B–3
in Appendix B. Odd numbered
Thermo Fisher Scientific