Operation
Data Server Software
3-14
Omni FTIR Multi Gas CEMS Instruction Manual
the user to enter basic configuration data. In order to make changes, it is
necessary that the program be in the Stopped mode.
Configure
This program uses the ASCII concentration data files (con) to drive the
current loops in directory C:\CONS.
Use the Add and Remove buttons to browse for the disk location of each
file type to be used.
The Refresh Interval specifies the time in milliseconds between refreshes of
the board's output. The software examines each of the paths to locate the
most recent file, and updates the data in its output buffers if they have
changed since the last query.
The DAQ Board specifies the National Instruments device number
assigned to the current loop board. Value 1 must be specified.
The AutoStart feature automatically begins data transmission when the
current loop program is opened.
Thermo Fisher Scientific