Heat; Imacc; Autostart - Thermo Scientific Omni FTIR Multi Gas CEMS Instruction Manual

Table of Contents

Advertisement

System Installation, Start Up and Shut Down
System Startup
2-12
Omni FTIR Multi Gas CEMS Instruction Manual

Heat

With the IMACC IFSS running, open the IMACC script editor. Refer to
Appendix E,
in the script to read system temperatures. Once the power has been applied,
wait until temperatures are at their set points. The cell heater control is set
on a 90 degree C/hour ramp rate.
Ejector Pump Temp
Heated Cell Temp
Heated line Inlet Temp
Heated line Outlet Temp

IMACC

The IMACC FSS software is preconfigured and starts automatically on
computer power up. Refer to Chapter 3, "Operation" for details on
software operation.

Autostart

For automatic system start on power-up and after power interruptions,
verify that the following shortcuts are pre-loaded into the Windows Startup
menu:
C:\layouts\nnnn.layout (Where nnnn is the desired layout)
C:\Program files\Imacc\IMACC FTIR Software Suite\Imacc System Process
Monitor.exe
C:\Program files\Imacc\IMACC FTIR Software Suite\Imacc Modbus
Server.exe
C:\Program files\Imacc\IMACC FTIR Software Suite\Imacc Modbus TCP
Server.exe
C:\Program files\Imacc\IMACC FTIR Software Suite\Imacc CurrentLoop
Server.exe
"IMACC Script
Setting" to open the Modbus control block
: 180 °C
: 185 °C
: 180 °C
: 180 °C