Replacing The Pc Console - Thermo Scientific Omni FTIR Multi Gas CEMS Instruction Manual

Table of Contents

Advertisement

Replacing the PC
Console
Thermo Fisher Scientific
6. Pull the ring to release the pin that holds the cable routing arm to the
bracket on the rear of the PC
7. Remove the PC from the rack by depressing the button on each slide
rail and pulling the PC out
8. Using the Phillips screwdriver, remove the slide rails from each side of
the PC, noting the orientation
9. Using the Phillips screwdriver, re-install the slide rails onto the new
PC.
10. Lift the PC and place each slide rail into the 2 extender pieces, then
push in the PC until it stops. Depress the 2 side tabs on the slide rails
and push the PC in until it locks in place, still outside of the rack
(Figure
5–46).
11. Re-attach the cable routing arm to the bracket on the rear of the PC by
lining up the holes and inserting the pin that was removed earlier.
12. Re-connect all of the cables to the rear of the PC in the proper
locations. Follow steps 1–3 in reverse order to complete the procedure.
Use the following procedure to replace the PC Console (monitor and
keyboard).
Equipment Required:
Phillips screwdriver
PC Console
1. Slide the "Lock" lever on the front of the PC console to the "Release"
position. Shut the PC down through Windows, and then shut off the
PC/Console circuit breaker on the breaker box.
2. Using the Phillips screwdriver, remove the blank panel in the front of
the system directly above the PC console
3. At the rear of the system, unplug the PC console power connector, as
well as the connectors for the console, mouse, and keyboard
47).
Preventive Maintenance and Servicing
(Figure
5–46).
(Figure
5–46).
(Figure
5–46).
(Figure
5–46).
Omni FTIR Multi Gas CEMS Instruction Manual

Replacing the PC Console

(Figure 5–
5-57