I/O Boards; Console; Datalogger; Electrical Panel Assembly - Thermo Scientific Omni FTIR Multi Gas CEMS Instruction Manual

Table of Contents

Advertisement

System Component Description

Console

I/O Boards

Console

Datalogger

Electrical Panel
Assembly
7-14
Omni FTIR Multi Gas CEMS Instruction Manual
A National Instruments PCI-6704 board is used to provide 4-20mA analog
outputs for all of the gas concentrations. This connects to a 68-screw
terminal connector board on the electrical panel assembly.
A National Instruments PCI-6509 board provides up to 96 digital
inputs/outputs which are used to control system valves, provide status
outputs and control inputs from the user. This board connects to a 30-
screw terminal connector board on the electrical panel assembly.
A 1U keyboard drawer with integrated touchpad is also included in the
system to facilitate user input at the system.
The display is a 9U 19" TFT high brightness industrial LCD monitor.
An optional data logger may be used to:
Capture user-selected measurement data in real-time
Allow the operator to trigger calibration sequences
Manually and automatically control calibration gases
Mark data taken during any calibration purge or activity
The electrical panel
(Figure
the following user connections:
4-20 mA Analog Outputs (one for each gas)
Digital solid state relay outputs (Temperature Alarm, Pressure Alarm,
Maintenance, FID Alarm, FTIR Data Valid, System Gas Mode)
Digital Inputs (System Gas Mode)
Probe Connections (AC Power, Temperature Alarm, and Valve
Control)
7–11) includes screw terminal connections for