Probe
3.4.3
Probe labeling
Each probe is labeled with the following information:
1
2
3
4
5
The probe labeling displays as the following figure:
3.4.4
Probe Connection
Connecting the probe
Probes can be connected or exchanged at any time, regardless of whether
the unit is powered On/Off. The operation is executed with the following
steps:
STEP 1
STEP 2
STEP 3
Rev 1.00
Type of the probe
Name of the probe
Probe serial number
Center frequency
The degree of preventing into the liquid
The patient is not allowed to touch any of connectors on the ultra-
WARNING
sound unit, including ultrasound probe connectors.
Handle the probe gently while connecting or disconnecting.
CAUTION
Put the probe carrying case on a stable surface and open the case.
Be careful to take out the probe and untied the probe cord. Do not
hang the probe head freely. Any impact to the probe head could re-
sult in irreparable damage.
Press the probe connector locking lever (Refer to Figure 3-6) down-
wards.
Figure 3-3: The Probe Labeling
Chapter 3 - Specifications
39