Whale PE-12-025 User Manual

Portable ultrasound diagnostic scanner
Table of Contents

Advertisement

Quick Links

1.00
ENG This user manual includes detailed usage instructions for Portable
Ultrasound Diagnostic Scanner. Please read this manual before using
equipment.

Advertisement

Table of Contents
loading

Summary of Contents for Whale PE-12-025

  • Page 1 1.00 ENG This user manual includes detailed usage instructions for Portable Ultrasound Diagnostic Scanner. Please read this manual before using equipment.
  • Page 2 The text of this manual was originally written, approved and published by the manufacturer in En- glish. This manual may not be reproduced, in whole or in part, without the written permission of Beijing East Whale Imaging Technology Co., Ltd..
  • Page 3: Table Of Contents

    Contents Contents Contents ..................3 Chapter 1 - Introduction..............9 Overview .......................11 Intended Use ......................11 Contraindications....................11 Configuration ......................11 Attention .......................12 Copyright ......................13 Chapter 2 - Safety................15 Safety Introduction.....................17 Safety Hazard Alerts...................17 Owner Responsibility ..................17 Regulatory Information ..................18 Safety Classification ...................18 Safety - Power Supply ..................18 2.6.1 Adapter......................18 2.6.2 Printer ......................19...
  • Page 4 Contents 2.10.2 Electrical Hazards..................25 2.10.3 Biological Hazards ..................25 2.10.4 Pacemaker Hazards..................25 2.11 Electrical Safety ....................26 2.12 Allergic Reaction ....................27 2.13 Electromagnetic Compatibility (EMC) ............27 2.13.1 Information on electromagnetic compatibility........27 Chapter 3 - Specifications ............. 33 Power Supply .......................35 Dimensions......................35 3.2.1 Console ......................35 Adapter........................36 Probe ........................36 3.4.1 Supported Probes..................36...
  • Page 5 Contents Chapter 6 - System Description............. 69 Console........................71 Control Panel .......................72 Main User Interface ....................81 Cart.........................83 6.4.1 Specifications ....................84 6.4.2 The Cart’s Ports .....................85 6.4.3 Locking wheels .....................86 6.4.4 Cart Height Adjustment................86 6.4.5 Portable Combination Lock...............88 Chapter 7 - System Operation............93 Introduction ......................95 Patient Management..................95 7.2.1 Patient Information..................96...
  • Page 6 Contents 7.7.2 ECG Controls....................176 7.7.3 Display the ECG curve ................176 7.7.4 Adjust the display of the ECG curve ............176 Measurement....................177 7.8.1 Measurement Overview ................177 7.8.2 General Measurement ................180 7.8.3 Special Measurement................188 7.8.4 Measurement Errors ................. 199 7.8.5 Measurement Result Table ..............
  • Page 7 Contents 8.1.2 System Inspection ..................272 8.1.3 Unit Cleaning ....................273 8.1.4 Static Electricity Interference Prevention..........275 Probe Care And Maintenance................275 8.2.1 Planned Maintenance................275 8.2.2 Probe Inspection..................276 8.2.3 Probe Cleaning and Disinfecting ............276 Chapter 9 - Components and Accessories ........279 Components.......................281 Appendix I - Labels and Nameplates ........... 283 Safety labels .......................285 Appendix II - Acoustic Output Report Table ........
  • Page 8 This page is intentionally left blank.
  • Page 9: Chapter 1 - Introduction

    Chapter 1 - Introduction CHAPTER 1 INTRODUCTION...
  • Page 10 This page is intentionally left blank.
  • Page 11: Overview

    The system is based on the accumulated experience of the research department and clinical depart- ment of Beijing East Whale Imaging Technology Co., Ltd. The Portable Ultrasound Diagnostic Scanner provides image generation in 2D (B) Mode, M-Mode, Anatomical M Mode (AMM), Color Doppler, Power...
  • Page 12: Attention

    Never modify this product, including system components, software, cables, and so on. User modification may cause safety hazards and degradation in system performance. All modification must be accomplished by the Service Engineer from Whale or others authorized by Whale. Portable Ultrasound Diagnostic Scanner User Manual...
  • Page 13: Copyright

    This manual is exclusively for the operation, repair, and maintenance of the - Portable Ultrasound Diagnostic Scanner. Beijing East Whale Imaging Tech- nology Company owns the copyright of this non-publicly published manual, and has the right to use it as confidential data.
  • Page 14 This page is intentionally left blank.
  • Page 15: Chapter 2 - Safety

    Chapter 2 - Safety CHAPTER 2 SAFETY...
  • Page 16 This page is intentionally left blank.
  • Page 17: Safety Introduction

    The owner should make certain to have only authorized and quali- fied Whale service personnel undertake the installation, maintenance, trou- bleshooting, calibration, and repair of the equipment. The owner of the Portable Ultrasound Diagnostic Scanner should ensure that only properly trained, fully qualified personnel are authorized to oper- ate the system.
  • Page 18: Regulatory Information

    It is advisable to maintain a list of autho- rized operators. Should the system fail to operate correctly, or if the unit does not respond to the commands described in this manual, the operator should contact the nearest Whale Service Office. Regulatory Information AAMI/ANSI ES60601-1:2005/A2:2010/CAN/CSA-C22.2 No. 60601-1: 2008 EN60601-1:2006/AC: 2010...
  • Page 19: Printer

    Chapter 2 - Safety Acoustic Output 2.6.2 Printer Digital Graphic Printer (Printer A) Model: UPD897 Input: 100-240V~ 50/60Hz 1.5-0.8A Video Copy Processor (Printer B) Model: P93C Input: 100-240V~ 50/60Hz 0.8A Connect the power supply of the printer to the respective power recep- tacle on the Cart.
  • Page 20: Acoustic Output Display

    It is recommended that all users receive proper training in applica- tions before performing them in a clinical setting. Contact the Whale Ultrasound sales representative for training assistance.
  • Page 21: Probe Safety

    Chapter 2 - Safety Probe Safety only when the maximum combination of control settings is selected. Most settings result in a much lower output. It is important to note the following: The duration of an ultrasound examination is just as important as the acoustic output, since patient exposure is directly related to the expo- sure time.
  • Page 22: Electrical Hazards

    Immersing the whole probe in liquid will damage the probe. Do not use the probe if the surface of probe is scratched which may cause electric shock. Please contact WHALE local representative immediately. A damaged probe can also increase the risk of electrical shock if conductive solutions come in contact with internal parts.
  • Page 23: Mechanical Hazards

    Chapter 2 - Safety Patient Safety Do not apply excessive force to the probe cable, to prevent insulation WARNING failure. 2.8.2 Mechanical Hazards Take precaution to avoid mechanical hazards. Inspect probes for sharp edges or rough surfaces that could injure sen- WARNING sitive tissue.
  • Page 24: Diagnostic Information

    These limitations must be considered before making any decision based on quantitative values. If in doubt, the nearest Whale Service Office should be consulted. Equipment malfunction or incorrect settings can result in measurement er- rors or failure to detect details in the image. The user must become thor- oughly familiar with the operation of the unit in order to optimize its performance and to recognize possible malfunctions.
  • Page 25: Electrical Hazards

    Fuses blown within 36 hours of being replaced may indicate a malfunction- ing electrical circuit within the system. In this event, the unit must be checked by Whale Ultrasound service personnel, No attempt should be made to replace the fuse with another fuse of a higher rating.
  • Page 26: Electrical Safety

    Chapter 2 - Safety Electrical Safety 2.11 Electrical Safety Device Classification The Portable Ultrasound Diagnostic Scanner is a Class I device, type BF. The ultrasound system must use a separate, dedicated power outlet. The connected power supply must be reliably grounded. Connecting the optional hard-copy device (printer) incorrectly may re- duce the electrical safety of the system.
  • Page 27: Allergic Reaction

    Suitable electrical isolation of such external AC out- lets may be required in order to meet IEC 60601-1 (2005) standards for elec- trical leakage. External devices can be used only if Whale marked them to be in com- CAUTION pliance with related standards (IEC 60601-1: 2005).
  • Page 28 Note: Beijing East Whale Imaging Technology Co.,Ltd. cannot guarantee that accessories, lines and transformers not delivered by Beijing East Whale Imaging Technology Co.,Ltd. will correspond with EMC requirements of EN 60601-1-2.
  • Page 29 Chapter 2 - Safety Electromagnetic Compatibility (EMC) sonnel should attempt to correct the problem by one or more of the follow- ing measures: Re-orient or re-locate the affected device. Increase the separation between the unit and the affected device. Power the equipment from a source other than that of the affected de- vice.
  • Page 30 Chapter 2 - Safety Electromagnetic Compatibility (EMC) Table 2-3: Electromagnetic Environment specified Emissions test Compliance Electromagnetic environ- ment - guidance Voltage fluctuations/ Complies flicker emissions IEC 61000-3-3 2.13.1.2 Electromagnetic immunity Table 2-4: Electromagnetic Immunity Environment Emissions test Compliance Electromagnetic environ- ment - guidance Electrostatic discharge ±2kV,±4kV, ±6kV...
  • Page 31 Chapter 2 - Safety Electromagnetic Compatibility (EMC) Table 2-4: Electromagnetic Immunity Environment Emissions test Compliance Electromagnetic environ- ment - guidance Conducted RF 3Vrms Portable and mobile RF communications equip- IEC 61000-4-6 150kHz to 80MHz ment should be used no 3Vrms closer to any part of the 150kHz to 80MHz Portable Ultrasound Diag-...
  • Page 32 Chapter 2 - Safety Electromagnetic Compatibility (EMC) Table 2-4: Electromagnetic Immunity Environment Emissions test Compliance Electromagnetic environ- ment - guidance Field strengths from fixed transmitters, such as base stations for radio (cellu- lar/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
  • Page 33: Chapter 3 - Specifications

    Chapter 3 - Specifications CHAPTER 3 SPECIFICATIONS...
  • Page 34 This page is intentionally left blank.
  • Page 35: Power Supply

    Chapter 3 - Specifications Power Supply Power Supply The Portable Ultrasound Diagnostic Scanner is powered either by its internal battery or by a separate power supply adapter unit connected to a separate power outlet for any range of 100 – 240 VAC, 50–60 Hz. The cart power supply requirement is 100~240V ~ 220VA 50/60 Hz Operating the unit with the wrong voltage causes damages, voiding WARNING...
  • Page 36: Adapter

    Chapter 3 - Specifications Adapter Adapter Input: 100~240 V ~ 220 VA 50/60 Hz Output: DC 19 V, 7.89 A Probe 3.4.1 Supported Probes The Portable Ultrasound Diagnostic Scanner supports four types of probes: Linear array probe Convex array probe Phased array probe Intracavitary convex probe The detailed description of the supported probes is illustrated as bellow.
  • Page 37 Chapter 3 - Specifications Probe Table 3-2: Probe Details Type Name and Icon Technical Data Supported Mode Application Convex 2C5V Center frequency:3.5 MHz 2D mode array probe Pitch: 0.48 mm M-Mode Renal Number of elements: 128 AMM-Mode Elevation Aperture: 12.5 Color Doppler ABD_Large Power Doppler...
  • Page 38: Probe Orientation

    Chapter 3 - Specifications Probe Table 3-2: Probe Details Type Name and Icon Technical Data Supported Mode Application Phased array 1P4V Center frequency: 2.5MHz 2D mode probe Pitch: 0.254 mm M-Mode Cardiac Number of elements: 64 AMM-Mode Elevation Aperture: 12 Color Doppler Ped-Cardiac Power Doppler...
  • Page 39: Probe Labeling

    Chapter 3 - Specifications Probe 3.4.3 Probe labeling Each probe is labeled with the following information: Type of the probe Name of the probe Probe serial number Center frequency The degree of preventing into the liquid The probe labeling displays as the following figure: Figure 3-3: The Probe Labeling 3.4.4 Probe Connection...
  • Page 40 Chapter 3 - Specifications Probe Align the probe connector with the probe port and carefully put it STEP 4 into the port (Refer to Figure 3-5). Figure 3-4: Probe Connection to the Unit Portable Ultrasound Diagnostic Scanner User Manual...
  • Page 41 Chapter 3 - Specifications Probe Lift the connector locking lever upwards to the absolute horizontal STEP 5 position to lock it in place (See Figure 3-6). Figure 3-5: Probe Connection Locking lever Carefully position the probe cord so that it can be free to move. STEP 6 Fault conditions can result in electric shock hazard.
  • Page 42: Printers

    Chapter 3 - Specifications Printers Pull the probe and connector straight out of the probe port. STEP 3 Carefully slide the probe and connector away from the probe port. STEP 4 Ensure the probe cables are free. STEP 5 Ensure the probe head is clean before putting it in its storage case. STEP 6 Printers Sony UP-D897...
  • Page 43: Environment Requirements

    Chapter 3 - Specifications Transport and Storage Environment 3.7.2 Environment Requirements The Portable Ultrasound Diagnostic Scanner requires consistent operational environment. Temperature and humidity requirements are displayed as be- low. Table 3-4: Operation Environment Requirements Temperature Humidity Air Pressure Operational 15%~70% 700—1060hPa 10℃~40℃...
  • Page 44 This page is intentionally left blank.
  • Page 45: Chapter 4 - Start-Up

    Chapter 4 - Start-up CHAPTER 4 START-UP...
  • Page 46 This page is intentionally left blank.
  • Page 47: Positioning The Unit

    Move the unit slowly and carefully. STEP 6 Unit Connection Eligible personnels from Whale should perform the initial unit installation. Before connecting the Portable Ultrasound Diagnostic Scanner, check the power adapter unit and cord, and ensure the voltage level conforms to elec- trical safety requirements.
  • Page 48: Check The Voltage Level

    Connect the output of the AC power adapter to the proper socket STEP 1 on the rear side of the Portable Ultrasound Diagnostic Scanner. Use the AC power adapter specially designed and approved by Whale CAUTION for the Portable Ultrasound Diagnostic Scanner.
  • Page 49: Battery

    Chapter 4 - Start-up Unit Connection Ensure there is nothing is placed over AC adapter’s power cable, and CAUTION the cable is not located where it can be tripped over or stepped on. Placed the AC adapter in a ventilated area before it is used to power CAUTION the unit.
  • Page 50 Chapter 4 - Start-up Unit Connection Don’t cause the battery to short circuit. WARNING Don’t disassemble or change the battery safety system. WARNING Keep the battery away from the fire or other heat sources. WARNING Avoid piercing the battery with sharp objects. WARNING Don’t use a damaged battery.
  • Page 51 ℃ ℉ ℉ Note: Only use batteries authorized by Whale. The status of the power sources are indicated both by a graphic icon on the bottom of the display while the system is operating, and by a pair of LED lights which lights up even while system is shut down.
  • Page 52: Battery Replacement

    Chapter 4 - Start-up Unit Connection Table 4-2: Battery Status Icons Status Icon Status Description No AC power and the battery is in use. Battery is almost empty (0%~25%). Users can click the status icon to check the power level of the battery. 4.2.4 Battery Replacement To replace the battery:...
  • Page 53: Connect The Peripheral/Accessory

    Chapter 4 - Start-up Unit Connection 4.2.6 Connect the peripheral/accessory The peripherals/accessory connectors are located on the rear and side of the unit, and the peripherals/accessory can be properly connected using the connector panel. The details of the connector panel are illustrated as below: Figure 4-2: Peripheral/Accessory Connector Panel S-Video Remote Control...
  • Page 54 IEC standards. Don’t touch the conductive parts of USB or Ethernet cables when con- WARNING necting equipment to unit. Use peripherals and accessories that are approved by Whale. CAUTION Portable Ultrasound Diagnostic Scanner User Manual...
  • Page 55: Connect The Digital Printer

    Chapter 4 - Start-up Unit Connection Alternate connections will require verification of compatibility to avoid CAUTION electric shock. 4.2.7 Connect the digital printer You can properly connect the digital printer to the unit using USB port 1 or 2. 4.2.8 Connect the video printer You can properly connect the video printer to the unit using the Remote Control port.
  • Page 56: Connect The External Monitor

    Chapter 4 - Start-up Unit Connection 4.2.10 Connect the external monitor You can properly connect the external monitor to the unit using the VGA port. 4.2.11 Connect to the Wireless Local Area Networks The unit supports the connection of the Wireless Local Area Networks (WLAN) to access files from other computers, connect to a printer, etc.
  • Page 57: Cart Docking Connection

    Chapter 4 - Start-up Unit Connection 4.2.12 Cart docking connection The cart is an optional device that facilitates clinical needs by offering more flexibilities such as connection to printers, additional probe ports, and easy mobility. Figure 4-3: The Cart Place the unit on the cart: when the display screen turns on or off, you can put the Portable Ultrasound Diagnostic Scanner on the cart.
  • Page 58 Chapter 4 - Start-up Unit Connection Push up the locking lever of the cart. STEP 1 Figure 4-4: The Locking Lever Portable Ultrasound Diagnostic Scanner User Manual...
  • Page 59 Chapter 4 - Start-up Unit Connection Grasp and hold the unit, let it slightly tilted towards your direction, STEP 2 and then put the unit flat on the cart guide, and ensure the front of the unit is closely to the baffle of the cart guide. Figure 4-5: Position the unit Push down the locking lever to lock the unit.
  • Page 60: Switch On/Off The Unit

    Chapter 4 - Start-up Switch on/off the unit Lift the unit from the cart guide. STEP 2 Figure 4-6: The Cart Power Interface AC Adapter: Plug in the AC Adapter power cable Printer A: Plug in the power cable of Digital Graphic Printer Printer B: Plug in the power cable of Video Printer Do not place the cart into a corner where it is difficult to reach the CAUTION...
  • Page 61 Chapter 4 - Start-up Switch on/off the unit Press On/Off button on the top right corner of the control pan- STEP 1 el, the Exit dialogue pops up. Select Yes option, the system enters the full shutdown se- STEP 2 quence.
  • Page 62 This page is intentionally left blank.
  • Page 63: Chapter 5 - Storage And Transportation

    Chapter 5 - Storage and Transportation CHAPTER 5 STORAGE AND TRANSPORTATION...
  • Page 64 This page is intentionally left blank.
  • Page 65: Unit Transportation

    Carefully put the probe head into the carrying case. STEP 3 Do not use excessive force or bump the probe face. CAUTION Remove the unit from the cart and put it in the travel bag supplied by Whale. Rev 1.00...
  • Page 66: Related Symbols

    Chapter 5 - Storage and Transportation Related Symbols Related Symbols Table 5-1: The Related Symbols Symbols Description Indicates that the transport package should be verti- cal during the transportation. Indicates that the package content is fragile, please handle with care. Indicates that the package content is sensitive to wet- ness.
  • Page 67 Chapter 5 - Storage and Transportation Related Symbols Table 5-1: The Related Symbols Symbols Description Indicates that the transport package can’t stack. Indicates that the transport package can’t turn over when transporting. Indicates that the transport package can’t use the hook when handling or transporting.
  • Page 68 Chapter 5 - Storage and Transportation Related Symbols Table 5-1: The Related Symbols Symbols Description Indicates the temperature limits to which the medical device can be safely exposed. Indicates the range of humidity to which the medical device can be safely exposed. Miscellaneous dangerous substances and articles Portable Ultrasound Diagnostic Scanner User Manual...
  • Page 69: Chapter 6 - System Description

    Chapter 6 - System Description CHAPTER 6 SYSTEM DESCRIPTION...
  • Page 70 This page is intentionally left blank.
  • Page 71: Console

    Chapter 6 - System Description Console Console fifi Figure 6-1: Console Overview LCD display Power indicator On/Off button Battery Probe locking handle Probe connector Speakers Monitor hinge Main menu and Soft menu buttons Alpha-numeric keyboard Functional keyboard Trackball Rear connector Panel Rev 1.00...
  • Page 72: Control Panel

    Chapter 6 - System Description Control Panel Control Panel The layout of the Portable Ultrasound Diagnostic Scanner control panel is il- lustrated by the following picture. The control panel is divided into seven sections. The buttons and controls are grouped together for ease of use. Figure 6-2: Control Panel Key Illumination The keys on the control panel are illuminated according to their availability:...
  • Page 73 Chapter 6 - System Description Control Panel Menu Area Figure 6-3: Menu Area Keys Main Menu keys Soft Menu key Main Menu Keys Table 6-1: Main Menu keys Description Four-way menu key There are 4 four-way menu keys in main menu, whose mode-specific functions vary according to the imaging mode and position that is currently active.
  • Page 74 Chapter 6 - System Description Control Panel Power Area Table 6-3: Power Area Key Description On/Off key Turn the unit On/Off, or set the unit to Standby Mode. Gain Area Table 6-4: Gain Area keys Description TGC Sliders There are six sliding keys, which compensate for depth- related attenuation in an image.
  • Page 75 Chapter 6 - System Description Control Panel Table 6-5: Freeze and Print keys Description Print This key is used to print the current imaging screen con- tent to a selected printer. Freeze Stops or restarts the data acquisition. For future use. PC Keyboard Area Keyboard Keys The functions of these keys are the same as those on the computer key-...
  • Page 76 Chapter 6 - System Description Control Panel Table 6-6: Alphabetical Function keys Description Protocol For future use Biopsy Gets into the biopsy function. Body Mark Displays the available body mark for the current examina- tion. Page Erase Erases all texts, body marks and measured results on the current screen.
  • Page 77 Chapter 6 - System Description Control Panel Table 6-7: Mode Selection keys Description Displays the 2D live acquisition mode that is the default mode for the unit. Displays the M-Mode examination screen and enables the M-Mode functions. Used for viewing motion patterns. Display the examination screen in Color Doppler mode.
  • Page 78 Chapter 6 - System Description Control Panel Table 6-7: Mode Selection keys Description Displays more Main menu controls. Trackball and its surrounding keys Table 6-8: Trackball and its surrounding keys Description Trackball Used for navigation and together with the surrounding keys, to move, select or activate objects on the screen.
  • Page 79 Chapter 6 - System Description Control Panel Table 6-8: Trackball and its surrounding keys Description Store Stores the current imaging window to disk. Stored images are displayed in the thumbnail area. Table 6-9: Basic Operations List of Trackball Operation item Description Click Move the cursor by scrolling the trackball to a...
  • Page 80 Chapter 6 - System Description Control Panel Patient Management Area Table 6-12: Patient Management Area keys Description Patient Displays the patient management window, user can cre- ate a new patient, search for the patient, view the exam- ination list of the patient, edit the patient information, browse the image list and so on.
  • Page 81: Main User Interface

    The Main User Interface is divided into seven areas as follow: Top Area From the left: Logo: displays the Whale company logo. Patient information: displays the patient ID and name. Hospital name: displays the institution or hospital name which config- ured in Preset window.
  • Page 82 Chapter 6 - System Description Main User Interface Examination date & time: displays the current examination date and time. Probe and application: displays the currently selected probe and appli- cation or the probe and application that were used for retrieved imag- MI and TIS/TIC/TIB indexes: displays the values of MI and TIS/TIC/TIB in- dexes for the active image.
  • Page 83: Cart

    Chapter 6 - System Description Cart Cart The cart provides several features that facilitates clinical use, including con- sole docking, better mobility, printing ability, more ports, more probes, etc. Rev 1.00...
  • Page 84: Specifications

    Chapter 6 - System Description Cart 6.4.1 Specifications Figure 6-5: The Size of the Standard Cart Figure 6-6: The Size of the Premium Cart Portable Ultrasound Diagnostic Scanner User Manual...
  • Page 85: The Cart's Ports

    Chapter 6 - System Description Cart Table 6-13: Specifications of the Carts Name Model Material Weight Standard Cart PC10 Plastic 45Kg Premium Cart PC20 Plastic 55Kg Note: Premium Cart is equipped with MUX (Probe Connector) and removable CD-ROM, while Standard Cart not. 6.4.2 The Cart’s Ports Figure 6-7: The Premium Cart’s Ports...
  • Page 86: Locking Wheels

    Chapter 6 - System Description Cart 6.4.3 Locking wheels The cart has four swivel wheels. Two frontal locking wheels have Wheel LOCKs, which can be pressed down to lock the wheels. Figure 6-8: Locking the Wheels Wheel LOCK Swivel wheel 6.4.4 Cart Height Adjustment The cart can be vertically adjusted by 200 mm(7.87 inch).
  • Page 87 Chapter 6 - System Description Cart To lower the cart, apply proper amount of pressure to the cart while pulling and holding the height adjustment handle. Figure 6-9: Adjust the Cart Height Cart height adjustment handle Rev 1.00...
  • Page 88: Portable Combination Lock

    Chapter 6 - System Description Cart 6.4.5 Portable Combination Lock Figure 6-10: The Portable Combination Lock Coiling cable Combination number Combination number Screwhead location Push button T-bar Portable Ultrasound Diagnostic Scanner User Manual...
  • Page 89 Chapter 6 - System Description Cart To install the combination lock: Take out of the combination lock connected with the cart via the STEP 1 coiling cable (1 in Figure 6-10). Figure 6-11: The combination cable lock Rotate the combination number (2 in Figure 6-10) to the number STEP 2 previously set for the lock, for example, the initial combination 0000.
  • Page 90 Chapter 6 - System Description Cart Keep the push button pressed, and then align and insert the T-bar STEP 4 to the security slot (1 in Figure 6-12) of the Portable Ultrasound Di- agnostic Scanner. Figure 6-12: Back of the equipment Security slot Release the push button.
  • Page 91 Rotate counter-clockwise the screwhead 90° to the initial position. STEP 4 Note: Please safeguard your personal combination code. If you lose your com- bination code, Beijing East Whale Imaging Technology Co., Ltd. will not be able to retrieve it for you. To unlock the combination lock Set the combination number to previous number.
  • Page 92 This page is intentionally left blank.
  • Page 93: Chapter 7 - System Operation

    Chapter 7 - System Operation CHAPTER 7 SYSTEM OPERATION...
  • Page 94 This page is intentionally left blank.
  • Page 95: Introduction

    Chapter 7 - System Operation Introduction Introduction This chapter describes all operations related to the Portable Ultrasound Di- agnostic Scanner. To begin an examination with a Portable Ultrasound Diagnostic Scanner consists of three steps: Create a new patient record or start a new examination from an ex- STEP 1 isting patient record.
  • Page 96: Patient Information

    Chapter 7 - System Operation Patient Management 7.2.1 Patient Information Create a new patient record Press Patient key on the control panel, the Patient Main window is STEP 1 displayed: Figure 7-1: Patient Handling Screen Portable Ultrasound Diagnostic Scanner User Manual...
  • Page 97 Chapter 7 - System Operation Patient Management Click Search/Create Patient button, the Search/Create Patient win- STEP 2 dow is displayed: Figure 7-2: The Search/Create Patient window Enter the Patient ID and other information. Patients in the database STEP 3 with matching information will be displayed in the Patient List field. If the inputted information is not in the patient record list, select STEP 4 Create Patient on the main menu, the new patient and a new exam...
  • Page 98 Chapter 7 - System Operation Patient Management create a new examination, and then select Begin Exam to start the new examination. Figure 7-3: Examination List Window In the scanning screen, the patient information is displayed on the left side of the Top Area. Delete a patient/examination To delete a patient record: Enter the Search/Create Patient window.
  • Page 99 Chapter 7 - System Operation Patient Management A warning message prompts user to confirm the action. STEP 4 Edit Patient information There are two methods get into the Patient Information window: Alternative 1: Press Patient key on the control panel, the Patient Main window is STEP 1 different compared to Figure 7-1 when an examination has already started.
  • Page 100 Chapter 7 - System Operation Patient Management Click Current Patient Info button, the Patient Information window STEP 2 is displayed (See "Patient Information Window" (s 7-5, p 100)). User can edit patient information in this window. Figure 7-5: Patient Information Window Alternative 2: Enter the Examination List window.
  • Page 101 Chapter 7 - System Operation Patient Management In the main menu, select Image List to enter the Image Browser STEP 2 screen. Figure 7-6: Image Browser Screen The images and cine loops of the selected patient are grouped ac- STEP 3 cording to the examinations.
  • Page 102 Chapter 7 - System Operation Patient Management If the Yes button were selected, the Search/Create Patient window is displayed. Users can create a new patient or select an exist patient to save these information. Enter the Patient Information window, and select End Exam in the main menu to end the exam.
  • Page 103 Chapter 7 - System Operation Patient Management Select a method and click Next button. A message “Are you sure you STEP 3 want to delete?” will be displayed. Click Yes button. “Deleting” will be displayed. STEP 4 After a brief delay, “Delete complete” will be displayed, click Yes but- STEP 5 ton to close.
  • Page 104 Chapter 7 - System Operation Patient Management Select the Backup radio button, Destination Device text box will STEP 4 display the file path to the removable disk. Figure 7-9: Backup Click Next button, “Are you sure you want to back up” will be dis- STEP 5 played.
  • Page 105: Review

    Chapter 7 - System Operation Patient Management Select More on the Main Menu, select the Manage Local Data to dis- STEP 3 play the Manage Local Data window. Select Restore radio button, the Source Device text box will display STEP 4 the file path to the removable disk.
  • Page 106 Chapter 7 - System Operation Patient Management Press Review key on the control panel, the stored images of the selected ex- amination or default examination are displayed in the Review Screen. Figure 7-10: Review Screen Page number. Selected image (blue border). Export the imagas and cines to a removable disk.
  • Page 107 Chapter 7 - System Operation Patient Management Import the images/cines from a removable disk Insert a removable disk into the USB port of the unit. STEP 1 Click Import button, the Import window is displayed. STEP 2 Figure 7-11: Import window Rev 1.00...
  • Page 108 Chapter 7 - System Operation Patient Management Choose folder which you want to load. The folder must contain two STEP 3 directories: AVI and Images. Figure 7-12: Two Sub-level Folders Select the AVI or Images folder and click OK button, then a Loading STEP 4 message is displayed.
  • Page 109 Chapter 7 - System Operation Patient Management Click Export button, the Export window is displayed. STEP 2 Figure 7-13: Export window Choose the Removable Disk path which you want to export and in- STEP 3 put a file name. Click OK button, then a message box pops up to dis- play the exporting status.
  • Page 110: Report

    Chapter 7 - System Operation Patient Management Click Save As button, the Save As window is displayed. STEP 4 Figure 7-14: Save As Window Modify the Save as type, there are three types: DICOM, AVI, JPG, STEP 5 MP4, BMP. If you select DICOM, you can also modify Store settings. Click OK button, a message box pops up to display the export sta- STEP 6 tus.
  • Page 111 Chapter 7 - System Operation Patient Management reviewed before the report is saved. Use the Worksheet window to adjust the measured data, use the Examination List window to edit comments and diagnosis, use Patient Information window to edit patient information. The final report can be printed on a regular laser printer.
  • Page 112 Chapter 7 - System Operation Patient Management In Report screen, select Template in the Main menu. The Tem- STEP 1 plate selection menu is displayed showing the available report templates: Figure 7-16: Template Selection Menu Select a template from the template list (Select Cancel to re- STEP 2 turn to the current template).
  • Page 113 Chapter 7 - System Operation Patient Management To remove an image from the report, press and hold down the STEP 4 Set key on the control panel, and using the trackball, drag the selected image out of the Images area. Add texts to the Findings and Diagnoses text boxes To add texts, perform either of the following operations: Type texts...
  • Page 114 Chapter 7 - System Operation Patient Management To retrieve an archived report Select Retrieve in the main menu, a list of the available reports STEP 1 for the current examination is displayed. The default name for the report is formatted: <template type>_<store date>-<store time>.
  • Page 115: Probe/Application

    Chapter 7 - System Operation Probe/Application Click OK to save the changes and close the Export dialog. STEP 5 Note: The file is saved in the PDF format. Probe/Application Probes and their related applications are selected from the Probes and ap- plications pop-up menus as described below.
  • Page 116 Chapter 7 - System Operation Probe/Application Move the trackball to the desired application. STEP 4 Figure 7-19: Applications Window Press Set key on the control panel to launch the application. STEP 5 Note: Make sure that the probe and application names displayed on the screen correspond to the actual probe and application selection.
  • Page 117 Chapter 7 - System Operation Probe/Application If the current application is a custom application, the Presets window will be displayed directly. Figure 7-20: Presets window If the current application is an user-defined application, a mes- sage box will be displayed. Click Yes button, the message box is closed and the imaging parameters of the current application will be updated.
  • Page 118 Chapter 7 - System Operation Probe/Application To select an application from the More window In the Applications window, trackball to the More option and press STEP 1 Set key on the control panel. The More window will be displayed. Figure 7-21: More window In the More window, trackball to the desired application and press STEP 2 Set key on the control panel to launch it.
  • Page 119: Imaging Mode

    Chapter 7 - System Operation Imaging Mode Imaging Mode 7.4.1 2D Mode 2D Mode Overview The 2D mode displays a two-dimensional gray scale image of the tissue within the probe's field of view. - + - + Figure 7-22: 2D Mode Screen Probe orientation marker GrayScale Bar Focus marker...
  • Page 120 Chapter 7 - System Operation Imaging Mode Table 7-1: 2D Mode Main Menu Controls Parameter Description Freq. Adjusts the probe's operating frequency. The selected fre- quency is displayed in the status window. For some probes/applications the lowest frequency settings will be Octave imaging settings.
  • Page 121 Chapter 7 - System Operation Imaging Mode Table 7-2: 2D Mode Soft Menu Controls Parameter Description Tint Control and adjust the image tint. Tint is available in both Live and Freeze. Power Controls the amount of acoustic power applied in all modes.
  • Page 122: Panoramic Imaging Mode

    Chapter 7 - System Operation Imaging Mode Use the Freq. (decrement frequency) to increase penetration. Use the Width to provide a greater field of view. 7.4.2 Panoramic Imaging Mode Panoramic Imaging Mode Overview Panoramic Imaging mode generates a two-dimensional image by rendering and merging continuous data acquired by the linear movement of the probe.
  • Page 123 Chapter 7 - System Operation Imaging Mode Table 7-4: Panoramic Imaging Mode Main Menu Controls Parameter Description Freq. Enables the adjustment of the probe's operating fre- quency. The selected frequency is displayed in the status window. For some probes/applications the lowest fre- quency settings will be Octave imaging settings.
  • Page 124 Chapter 7 - System Operation Imaging Mode Table 7-6: Other Panoramic Imaging Mode Controls Parameter Description Compensates for depth-related attenuation in an image. The sliders nearest the operator affect the far field. TGC amplifies returning signals to compensate for the attenua- tion caused by tissue at increasing depths.
  • Page 125: M Mode

    Chapter 7 - System Operation Imaging Mode 7.4.3 M Mode M Mode Overview Display a distance/time plot of a cursor line in the axial plane of the 2D im- age. Figure 7-24: M Mode Screen Time motion cursor conventional M-Mode Gray bar M-Mode parameters windows Depth scale...
  • Page 126 Chapter 7 - System Operation Imaging Mode Table 7-7: M Mode Main Menu Controls Parameter Description Up/Down Enable a mirror image of the 2D image to be created. The reference marker W moves correspondingly to the other side of the image. Up/Down is available in both Live and Freeze.
  • Page 127 Chapter 7 - System Operation Imaging Mode Table 7-9: Other M Mode Controls Parameter Description Depth Adjusting the depth will influence diagram frame rate and resolution, Decreasing the depth may allow for higher frame rate and lower resolution. Increasing the depth will increase image resolution and decrease frame rate.
  • Page 128: Anatomical M-Mode

    Chapter 7 - System Operation Imaging Mode 7.4.4 Anatomical M-Mode Anatomical M-Mode Overview Displays a distance/time plot from a cursor line, which is independent from the axial plane. Anatomical M-Mode (AMM) is available in gray scale mode. Figure 7-25: AMM Mode Screen Time motion cursor conventional M-Mode Gray bar M-Mode parameters windows...
  • Page 129 Chapter 7 - System Operation Imaging Mode Table 7-10: AMM Mode Main Menu Controls Parameter Description Up/Down Enable a mirror image of the 2D image to be created. The reference marker W moves correspondingly to the other side of the image. Up/Down is available in both Live and Freeze.
  • Page 130 Chapter 7 - System Operation Imaging Mode Table 7-12: Other AMM Mode Controls Parameter Description Display more Main menu controls: AMM, TDI (phased array probe). Using AMM Mode From 2D Live, press Alt key on the control panel. STEP 1 Select AMM to access AMM mode.
  • Page 131: Cm Mode

    Chapter 7 - System Operation Imaging Mode 7.4.5 CM Mode In CM Mode, the monitor displays a live color image about blood flow, in- cluding the direction of blood flow and spectrum information. This function is available only when the phased array probe is being installed. Figure 7-26: CM Mode Activate CM Mode To activate CM Mode,...
  • Page 132 Chapter 7 - System Operation Imaging Mode Table 7-13: The Main Menu Controls of CM Mode Parameter Description Freq. Adjusts the transmission frequency to control the sensitiv- ity and the level of penetration. The selected frequency is displayed in the status window. Adjusting Frequency may affect Sample Volume and LVRej settings.
  • Page 133: Color Mode

    Chapter 7 - System Operation Imaging Mode Table 7-15: Other Controls of CM Mode Parameter Description Compensates for depth-related attenuation in an image. The sliders nearest the operator affect the far field. TGC amplifies returning signals to compensate for the attenua- tion caused by tissue at increasing depths.
  • Page 134 Chapter 7 - System Operation Imaging Mode Color ROI Color Mode parameters window Color Mode Soft menu Color Mode Main menu Color Mode Main Menu Controls Table 7-16: Color Mode Main Menu Controls Parameter Description Scale Adjusts the repetition rate of the Doppler pluses transmit- ted when acquiring the data for color flow mapping.
  • Page 135 Chapter 7 - System Operation Imaging Mode Table 7-17: Color Mode Soft Menu Controls Parameter Description Invert Inverts the color scheme for positive and negative veloci- ties. Invert is available in Live and Freeze. Power Controls the amount of acoustic power applied in all modes.
  • Page 136 Chapter 7 - System Operation Imaging Mode Optimizing Color Mode Preset generally gives optimal performance with minimal adjustment. If necessary, the following controls can be used to further optimize the Color Mode display: Select Gain to adjust gain in the color flow area. Set Scale to the highest setting that provides adequate flow detection.
  • Page 137: Power Mode

    Chapter 7 - System Operation Imaging Mode 7.4.7 Power Mode Power Mode Overview Power Mode allows the observation of blood flow using backscatter intensi- ty of ultrasound signal from blood cells. - + Figure 7-28: Power Mode Screen Probe orientation marker Color bar Color ROI Power Mode parameters window...
  • Page 138 Chapter 7 - System Operation Imaging Mode Table 7-19: Power Mode Main Menu Controls Parameter Description Steer Controls the angle of the color sample volume (Only for Liner Array Probe). Freq. Adjusts the transmission frequency to control the sensitiv- ity and the level of penetration. The selected frequency is displayed in the status window.
  • Page 139 Chapter 7 - System Operation Imaging Mode Table 7-21: Other Power Mode Controls Parameter Description Depth Adjusting the depth will influence diagram frame rate and resolution. Decreasing the depth may allow for higher frame rate and lower resolution. Increasing the depth will increase image resolution and decrease frame rate.
  • Page 140: Dpdi Mode

    Chapter 7 - System Operation Imaging Mode Use Priority to emphasize either the color flow overlay, or the detail of the underlying gray scale tissue. Use Persist to enhance the sensitivity of the signal’s correlation in the time domain. Use Radial Avg. and Lateral Avg. to reduce noise in the color flow area. 7.4.8 DPDI Mode In DPDI Mode, the monitor displays a live color image about blood flow, in-...
  • Page 141 Chapter 7 - System Operation Imaging Mode The Main Menu Controls of DPDI Mode Table 7-22: The Main Menu Controls of DPDI Mode Parameter Description Freq. Adjusts the transmission frequency to control the sensitiv- ity and the level of penetration. The selected frequency is displayed in the status window.
  • Page 142: Elastography Image Mode

    Chapter 7 - System Operation Imaging Mode Other Controls of DPDI Mode Table 7-24: Other controls of DPDI Mode Parameter Description Gain Adjusted by the knob below the TGC on the control panel. When rotated clockwise, increases the overall gain applied to the received echo signals equally for all depth.
  • Page 143 Chapter 7 - System Operation Imaging Mode Probe orientation marker 2 Color bar Color ROI EI Mode parameters window EI Mode Soft menu EI Mode Main menu EI Mode Main Menu Controls Table 7-25: EI Mode Main Menu Controls Parameter Description Scale Adjusts the repetition rate of the Doppler pluses transmit-...
  • Page 144 Chapter 7 - System Operation Imaging Mode Other EI Mode Controls Table 7-27: Other EI Mode Controls Parameter Description Gain Adjusted by the knob below the TGC on the control panel. When rotated clockwise, increases the overall gain applied to the received echo signals equally for all depth. Compensates for depth-related attenuation in an image.
  • Page 145: Tvi Mode

    Chapter 7 - System Operation Imaging Mode Select LVRej to remove low velocity blood flow and tissue movement to improve image quality. Select Freq. to optimize the color flow display. Higher settings improve resolution. Lower settings improve depth penetration and sensitivity. Select Frame Rate to adjust the Frame Rate.
  • Page 146 Chapter 7 - System Operation Imaging Mode TVI Mode Main Menu Controls Table 7-28: TVI Mode Main Menu Controls Parameter Description Scale Adjusts the repetition rate of the Doppler pluses transmit- ted when acquiring the data for color flow mapping. The Scale (Nyquist limit) should be adjusted so that no aliasing occurs, while still having good resolution of velocities.
  • Page 147 Chapter 7 - System Operation Imaging Mode Other TVI Mode Controls Table 7-30: Other TVI Mode Controls Parameter Description Gain Adjusted by the knob below the TGC on the control panel. When rotated clockwise, increases the overall gain applied to the received echo signals equally for all depth. Compensates for depth-related attenuation in an image.
  • Page 148: Strain Rate Imaging Mode

    Chapter 7 - System Operation Imaging Mode Select LVRej to remove low velocity blood flow and tissue movement to improve image quality. Select Freq. to optimize the color flow display. Higher settings improve resolution. Lower settings improve depth penetration and sensitivity. Select Power to obtain an acceptable image using the lowest setting possible.
  • Page 149 Chapter 7 - System Operation Imaging Mode SRI Mode Main Menu Controls Table 7-31: SRI Mode Main Menu Controls Parameter Description Scale Adjusts the repetition rate of the Doppler pluses transmit- ted when acquiring data for color flow mapping. The Scale (Nyquist limit) should be adjusted so that no aliasing occurs, while still having good resolution of velocities.
  • Page 150 Chapter 7 - System Operation Imaging Mode Other SRI Mode Controls Table 7-33: Other SRI Mode Controls Parameter Description Gain Adjusted by the knob below the TGC on the control panel. When rotated clockwise, increases the overall gain applied to the received echo signals equally for all depth. Compensates for depth-related attenuation in an image.
  • Page 151: Dual Live Mode

    Chapter 7 - System Operation Imaging Mode Select LVRej to remove low velocity blood flow and tissue movement to improve image quality. Select Freq. to optimize the color flow display. Higher settings improve resolution. Lower settings improve depth penetration and sensitivity. Select Power to obtain an acceptable image while using the lowest set- ting possible.
  • Page 152 Chapter 7 - System Operation Imaging Mode The Soft Menu Controls of The Dual Live Mode Determine the current imaging mode being one of Color, Power, DPDI, EI, TVI or SRI mode. The Soft menu controls refer to the corresponding mode (Color Mode/Power Mode/DPDI Mode/EI Mode/TVI Mode/SRI Mode).
  • Page 153: Pw Mode

    Chapter 7 - System Operation Imaging Mode 7.4.13 PW Mode PW Mode Overview PW Mode allows the observation of the speed and characteristics of blood flow in the way of pulse wave Doppler single samples. Figure 7-34: PW Mode Screen Sample Volume Angle correction marker Angle Value...
  • Page 154 Chapter 7 - System Operation Imaging Mode Table 7-35: PW Mode Main Menu Controls Parameter Description Scale Enables the low velocity portions of the spectrum to be fil- tered, since the Doppler spectrum and audio may contain strong wall-motion signals. The scale value may affect FPS and LVRej.
  • Page 155 Chapter 7 - System Operation Imaging Mode Table 7-36: PW Mode Soft Controls Parameter Description Tint Adjusts Doppler spectrum tint. Tint is available in live and Freeze. Invert Enables the Doppler spectrum to be flipped 180 degrees, so that negative velocities are displayed above the base- line and positive velocities below the baseline.
  • Page 156 Chapter 7 - System Operation Imaging Mode Optimize the image by adjusting the image controls describe in the previous section. In PW mode, user can start CW/TVD mode by pressing CW key or se- lecting TDI parameter of the Main menu. Optimizing PW Doppler Mode Preset generally gives optimal performance with minimal adjustment.
  • Page 157: Cw Mode

    Chapter 7 - System Operation Imaging Mode 7.4.14 CW Mode CW Mode Overview CW Mode allows observation of the highest speed and characteristics of blood flow using continuous wave Doppler signal. Figure 7-35: CW Mode Screen Angle correction marker Sample Volume CW cursor line Invert marker Angle Value...
  • Page 158 Chapter 7 - System Operation Imaging Mode Table 7-38: CW Mode Main Menu Controls Parameter Description Scale Enables the low velocity portions of the spectrum to be fil- tered, since the Doppler spectrum and audio may contain strong wall-motion signals. The scale value may affect FPS and LVRej.
  • Page 159 Chapter 7 - System Operation Imaging Mode Table 7-39: CW Mode Soft Menu Controls Parameter Description Invert Enables the Doppler spectrum to be flipped 180 degrees, so that negative velocities are displayed above the base- line and positive velocities below the baseline. Power Controls the amount of acoustic power applied in all modes.
  • Page 160 Chapter 7 - System Operation Imaging Mode In CW mode, user can start PW/TVD mode by pressing PW key or select- ing TDI parameter of the Main menu. Optimizing CW Doppler Mode Preset generally gives optimal performance with minimal adjustment. If necessary, the following controls can be used to further optimize the CW mode display: Select Gain to adjust gain in the spectral Doppler area.
  • Page 161: Tvd Mode

    Chapter 7 - System Operation Imaging Mode 7.4.15 TVD Mode TVD Mode Overview TVD Mode allows for observations on the speed and characteristics of tissue using pulse waved based tissue Doppler signal. Figure 7-36: TVD Mode Screen Angle correction marker Sample Volume Color ROI TVD cursor line...
  • Page 162 Chapter 7 - System Operation Imaging Mode Table 7-41: TVD Mode Main Menu Controls Parameter Description Scale Enables the low velocity portions of the spectrum to be fil- tered, since the Doppler spectrum and audio may contain strong wall-motion signals. The scale value may affect FPS and LVRej.
  • Page 163 Chapter 7 - System Operation Imaging Mode Table 7-42: TVD Mode Soft Menu Controls Parameter Description Invert Enables the Doppler spectrum to be flipped 180 degrees, so that negative velocities are displayed above the base- line and positive velocities below the baseline. Power Controls the amount of acoustic power applied in all modes.
  • Page 164 Chapter 7 - System Operation Imaging Mode In TVD mode, user can start PW/CW mode by pressing PW/CW key. Optimizing TVD Doppler Mode Preset generally gives optimal performance with minimum adjustment. If necessary, the following controls can be adjusted to further optimize the TVD mode display: Select Gain to set the gain in the spectral Doppler area.
  • Page 165: The Split Screen

    Chapter 7 - System Operation Imaging Mode 7.4.16 The Split Screen The Split Screen Overview Figure 7-37: The Split Screen Active window marker Gray bar of the active window Parameter window of the active window 4 Soft menu of the active window Inactive window Main menu of the active menu Active window...
  • Page 166 Chapter 7 - System Operation Imaging Mode Other Controls on The Split Screen Determine the current imaging mode being one of 2D or Color mode. The following controls are responsible for the regulating the corresponding mode. Table 7-44: Other Controls on The Split Screen Parameter Description Gain...
  • Page 167: The Quad Screen

    Chapter 7 - System Operation Imaging Mode 7.4.17 The Quad Screen The Quad Screen Overview Figure 7-38: The Quad Screen Inactive window(1) Active window marker Active window Color bar of the active window Parameter window of the active window Soft menu of the active window Inactive window(2) Main menu of the active window inactive window(3)
  • Page 168: Cine Loop

    Chapter 7 - System Operation Cine Loop The Quad Screen Other Controls Determine the current imaging mode being one of 2D or Color mode. The following controls are responsible for the regulating the corresponding mode. Table 7-45: The Quad Screen Other Controls Parameter Description Gain...
  • Page 169: Cine Storage Buffer

    Chapter 7 - System Operation Cine Loop 7.5.1 Cine Storage Buffer You can view all the images stored in the cine buffer. When cine storage buf- fer is full, the earliest images will be replaced. The data coverage process is ongoing.
  • Page 170: Cine Loop Overview

    Chapter 7 - System Operation Cine Loop 7.5.2 Cine Loop Overview The cine loop icon and main menu controls are displayed as the following figure: Figure 7-39: Cine Loop Screen Left marker Frame indicator Right marker Frame counter Cineloop speed Cineloop Main menu 7.5.3 Cine Loop Main Menu Controls...
  • Page 171: Cine Loop Other Controls

    Chapter 7 - System Operation Cine Loop 7.5.4 Cine Loop Other Controls The cine loop other controls on the control panel are displayed as the follow- ing figure: Figure 7-40: Cine Loop Other Controls On the Control Panel Table 7-47: Cine Loop Main Menu Other Controls Controls Description Trackball...
  • Page 172: Store And View A Cine File

    Chapter 7 - System Operation Cine Loop To view a cine loop automatically Press Freeze key on the control panel to start the Cine mode. STEP 1 Cine icon shows the current total number of frames stored, Frame STEP 2 indicator stays in the last frame.
  • Page 173 Chapter 7 - System Operation Cine Loop To view a single-frame cine file Press Trackball key on the control panel to activate the mouse, and STEP 1 navigate the cursor to the single-frame cine file. Use the Set key to double click the single -frame cine file. The image STEP 2 of this single -frame cine file will be displayed.
  • Page 174: Zoom

    Chapter 7 - System Operation Zoom You can stop or start automatic playback using the controls in the STEP 3 cine loop menu. Automatic playback status can be changed to manual single-frame STEP 4 playback by scrolling the trackball. Exit cine file playback Press Freeze key on the control panel or scan mode key on the control panel to exit the cine file, reload the preset parameters, and return to the 2D real- time mode.
  • Page 175: Ecg Overview

    Chapter 7 - System Operation The ultrasound image displayed is synchronized with the ECG curve. In M- Mode or Doppler, the ECG curve is synchronized to that particular mode's sweep. The operator can control the gain, the position and the sweep rate of the ECG curve using the Main menu keys on the control panel.
  • Page 176: Ecg Controls

    Chapter 7 - System Operation 7.7.2 ECG Controls Press Physic key on the control panel to activate physiological ECG controls. The physiological ECG controls are displayed on the Main menu of the scan- ning screen. Table 7-50: ECG Controls Parameter Description Horizontal Sweep Adjusts the refresh rate of the physiological curve.
  • Page 177: Measurement

    Chapter 7 - System Operation Measurement Select the assignable button Gain to adjust the amplitude of the STEP 2 ECG curve. Note: The ECG signal's amplitude may vary between patients due to different skin moisture and other physiological parameters. Adjust the ECG curve position Press Physic key on the control panel.
  • Page 178 Chapter 7 - System Operation Measurement Measurement Interface As shown in the following figure, we need to have a detailed understanding of the following five areas/menus including the measurement process: Figure 7-43: Measurement Interface Measurement result table Image measurement area Special Measurement main menu General measurement menu Image measurement area...
  • Page 179 Chapter 7 - System Operation Measurement Use the soft menu to select the measurements of the corresponding general items. Caliper Type The caliper is a measuring mark used to indicate the starting point and the end point. Some caliper groups need multiple measurements to calculate a value.
  • Page 180: General Measurement

    Chapter 7 - System Operation Measurement on the sub-menu above the screen. The default settings for general measurement methods can be adjusted here. Basic Measurement Procedures In general, the basic measurement procedures are as follows: Enter the desired mode, and press Measure key on the control pan- STEP 1 Select a general measurement or use the trackball to select a Special STEP 2...
  • Page 181 Chapter 7 - System Operation Measurement Distance Measurement Function: measuring the distance between two points of the ultrasonic im- age. Enter B mode and press Measure key on the control panel. STEP 1 Select Distance in the main menu. STEP 2 Roll the trackball to put the caliper on the image, and position the STEP 3 first measurement mark and press Set key.
  • Page 182 Chapter 7 - System Operation Measurement Access the measurement menu. STEP 1 Press the function key to select the toggle Area, then select El- STEP 2 lipse. Roll the trackball to position the first measurement mark, and STEP 3 press Set key. Roll the trackball again, and the system will dis- play an ellipse.
  • Page 183 Chapter 7 - System Operation Measurement Diameter Stenosis Rate and Area Stenosis Rate Measurements Diameter stenosis rate and area stenosis rate can be set in the Preset or through the soft menu; refer to distance measurement and ellipse area mea- surement for its measurement methods, which will not be repeated here.
  • Page 184 Chapter 7 - System Operation Measurement Formula: Segment length D1, D2, D3, Formula unit: cm. 1-Ellipse Method Function: Use the ellipse method to calculate the volume, and please refer to the Ellipse Area Method for specific measurement methods. Af- ter the measurement is completed, the volume will be displayed in the Measurement result table.
  • Page 185 Chapter 7 - System Operation Measurement Roll the trackball to set the vertical dotted lines, and align the end STEP 4 points on the dotted line, press Set key on the control panel to lo- cate the vertical line segment. The measured results will be dis- played in the Measurement result table.
  • Page 186 Chapter 7 - System Operation Measurement el to position the focus, and the velocity measurement will be displayed in the Measurement result table. PI (Auto) Measurement After entering PW measurement mode, click PI Auto to select auto- STEP 1 matic tracing measurement. A tracing caliper occurs on the screen, press Set key on the control STEP 2 panel to position the caliper and set the starting point.
  • Page 187 Chapter 7 - System Operation Measurement Refer to Acceleration Measurement for measuring methods. HR Measurement Function: To depict a cardiac cycle through the measurement mark dis- played in the form of vertical line, based on which the HR can be deter- mined.
  • Page 188: Special Measurement

    Chapter 7 - System Operation Measurement Roll the trackball to put the caliper on the end point, and press STEP 5 Set key on the control panel. The measurement results will be displayed in the Measure- STEP 6 ment result table. Making an Area Flow Volume Method Function: Area flow is an estimated value of the blood flow that is cal- culated based on area.
  • Page 189 Chapter 7 - System Operation Measurement Press Measure key on the control panel. The Measurement soft STEP 1 menu is displayed. Figure 7-49: Special Measurement Soft Menu Exam subjects Selected Exam Opened Exam Measurements related to the Opened Exam Enter a Exam and perform measurement Press Measure key on the control panel to enter Measurement STEP 1 mode.
  • Page 190 Chapter 7 - System Operation Measurement Press Set key on the control panel to activate the measurement tool. STEP 5 The cursor will move back to the scanning window. Perform the measurement. Completed measurements are marked STEP 6 with a check mark after completion, and the next measurement on the list is automatically selected.
  • Page 191 Chapter 7 - System Operation Measurement Check whether the system date is the current date. Press Patient key on the control panel to input the information of the patient. Press Application key on the control panel to select cardiac Exam. Cardiac Measurement in 2D Mode In 2D mode, users can carry out the measurements of cardiac area listed in the main menu.
  • Page 192 Chapter 7 - System Operation Measurement Table 7-52: Cardiac Measurement Item In 2D Mode Item Description Measurement methods LVAs sax PM Left ventricular short axis Trace/Ellipse area at papillary muscle level at end systole LVAs sax MV Left ventricular short axis Trace/Ellipse area at mitral valve level at end systole...
  • Page 193 Chapter 7 - System Operation Measurement Table 7-52: Cardiac Measurement Item In 2D Mode Item Description Measurement methods LVIDs Left Ventricular Internal Trace/Ellipse Diameter in Systole Teichholz Diastole Diastole RVAWd Right ventricular anterior Distance wall thickness at end systole RVDd Right ventricular dimension Distance at end systole...
  • Page 194 Chapter 7 - System Operation Measurement Operations of elliptical area measurements are: Enter 2D mode and press Measure key on the control panel to enter STEP 1 the measurement interface. Select the MV measurement item from the cardiac measurement STEP 2 soft menu (taking mitral valve for example);...
  • Page 195 Chapter 7 - System Operation Measurement Table 7-53: Cardiac Measurement Item In M Mode Item Description Measurement methods DE Excursion Mitral valve leaflet excursion Slope in anterior diastole DE amp DE-wave amplitude Distance EPSS Distance from E point to sep- Distance tal separation, that is, dis- tance from mitral valve...
  • Page 196 Chapter 7 - System Operation Measurement On the soft menu, the cardiac menu options are displayed. Trackball STEP 2 to the measurement item. The appropriate measurement methods will be displayed in the soft menu. A vertical line will be displayed on the system screen. Move the STEP 3 trackball to position the starting point of the vertical line and press Set key on the control panel, then move the trackball to position the...
  • Page 197 Chapter 7 - System Operation Measurement Table 7-54: Cardiac Measurement Item In Doppler Mode Item Description Measurement methods LVET Left Ventricular Ejection Time Time MV C.O dur Mitral valve close/open dura- Time tion Tel index Tel Index a Time interval between Time beginning and end of mitral valve filling...
  • Page 198 Chapter 7 - System Operation Measurement Table 7-54: Cardiac Measurement Item In Doppler Mode Item Description Measurement methods TV A pt Tricuspid valve A point Velocity RVET Right ventricle ejection time Time TV C.O dur Tricuspid valve close/open Time duration Pulmonic Valve PV Vmax Pulmonary valve maximum...
  • Page 199: Measurement Errors

    Chapter 7 - System Operation Measurement On the soft menu, click IMT, and then select RT or LT from the IMT STEP 5 list. A sampling frame will be displayed on the monitor. Note: RT represents the carotid which is at the right side of one’s neck, while LT represents the left one.
  • Page 200 Chapter 7 - System Operation Measurement turn, the image quality depends on system design and skills of operators as well as their understandings of system features. The ultrasound imaging algorithms are based on the sound traveling 1540m/s in tissue. However, in different tissue types, velocity of sound may be different.
  • Page 201: Measurement Result Table

    Chapter 7 - System Operation Measurement 7.8.5 Measurement Result Table The Measurement Result Table can be moved to prevent the table obscuring parts of the ultrasound image (See "Measurement Result Table" (s 7-51, p 201)). Figure 7-51: Measurement Result Table Moving the Measurement Result Table Trackball to the symbol on the heading of the Measurement Result STEP 1...
  • Page 202: Worksheet

    Chapter 7 - System Operation Worksheet Table 7-55: Measurement Units Calculation item Unit Acceleration cm/s Volume flow Ratio Worksheet Press Worksheet key on the control Panel, the Worksheet window will be displayed: Figure 7-52: Worksheet Window Measurement type Measurement parameter Value: Averaging, Max, Min or Last 4 Value type Measured/ calculated values...
  • Page 203 Chapter 7 - System Operation Worksheet To select the type of value Figure 7-53: Type Of Value Trackball to the relevant cell in the Method column. STEP 1 Press Set key. A pop-up menu is displayed showing the different op- STEP 2 tions available.
  • Page 204: Text

    Chapter 7 - System Operation Text Export a worksheet To export a report, follow these steps: In the USB port, insert a USB disk. STEP 1 On the control panel, press Worksheet. STEP 2 On the Worksheet dialog, click Export. STEP 3 The system will pop up the Export dialog.
  • Page 205: Insert A Text

    Chapter 7 - System Operation Text Anchored text Free text Text Soft menu: displays a lirbary of application-specified Text Main menu text 7.10.2 Insert a text Free text Press Text key on the control panel to activate the Text function. STEP 1 Move the cursor by scrolling the trackball to the insertion position.
  • Page 206: Modify A Text

    Chapter 7 - System Operation Text 7.10.4 Modify a text To modify a text, On the control panel, press Text to activate the Text function. STEP 1 Click in the text that you want to modify. STEP 2 Note: The cursor’s position is an insertion point. Edit the text with the keyboard.
  • Page 207: Arrow

    Chapter 7 - System Operation Arrow 7.11 Arrow 7.11.1 Arrow Overview Press Arrow key on the control panel to activate the Arrow function. Figure 7-55: Arrow Function Anchored Arrow Free Arrow Arrow Main menu Rotate Arrow 7.11.2 Insert an arrow Press Arrow key on the control panel to activate the Arrow function.
  • Page 208: Delete An Arrow

    Chapter 7 - System Operation Body Mark Press Set key on the control panel to anchor the arrow. STEP 4 7.11.4 Delete an arrow To use Arrow controls Press Arrow key on the control panel to activate the Arrow function. STEP 1 Select Next in the Arrow main menu to select the arrow to be delet- STEP 2...
  • Page 209: Body Mark Overview

    Chapter 7 - System Operation Body Mark 7.12.1 Body Mark Overview Press Body Mark key on the control panel to activate the Body Mark func- tion. Figure 7-56: Body Mark Function Body mark Soft menu: displays the pre-defined body mark Probe marker library Body mark...
  • Page 210: Modify A Body Mark

    Chapter 7 - System Operation Body Mark Move the cursor by scrolling the trackball to the first item of the STEP 2 Body Mark soft menu, and then press Set. A list of the available body mark libraries will pop up: Figure 7-57: The list of available body mark libraries Move the cursor to the desired body mark library and then press Set STEP 3...
  • Page 211: Puncture Guide Line

    Chapter 7 - System Operation Puncture Guide Line 7.13 Puncture Guide Line To accurately pierce body cavity with the puncture needle, mark the point of puncture and identify the puncture path based on the puncture guide line in a real-time image. Press Biopsy, the puncture guide line will be displayed on the monitor.
  • Page 212 Chapter 7 - System Operation Preset ers to customize the global configuration for the unit and change application-specified settings. The configuration management package consists of a Preset dialogue win- dow divided in different setup categories with sub-levels. The table below summarizes the contents and access rights of the different categories and sub-levels of the Portable Ultrasound Diagnostic Scanner configuration package: Table 7-57: The contents and access rights of the preset...
  • Page 213 Chapter 7 - System Operation Preset Table 7-57: The contents and access rights of the preset Category and sub-levels Description Access Create pre-defined body mark libraries. Mark Library Sets the body Mark Property. Patient Sets the default dataflow. DataFlow Configures the DICOM server. Sets the options relate to the patient Patient management function.
  • Page 214 Chapter 7 - System Operation Preset Press Config-Diag key on the control panel, the Preset dialogue window is displayed: Figure 7-59: Preset Dialogue Window Sublevel tabs for the selected Preset category Load Factory button Selected Preset category Preset categories Save and exit button Cancel (Exit without saving) Apply button Portable Ultrasound Diagnostic Scanner User Manual...
  • Page 215: System

    Chapter 7 - System Operation Preset 7.14.1 System Region Figure 7-60: The Region Preset Sheet Using the Region preset sheet, users can configure the hospital information, system language and time, and H&W format. The detail description of this sheet is illustrated below. Table 7-58: The Region Configuration Items Item Description...
  • Page 216 Chapter 7 - System Operation Preset Table 7-58: The Region Configuration Items Item Description Language: select the desired language for the system from the pop-up menu. Time Format: select the desired format. Date Format: select the desired format. System Time: set the time. Press the arrow head buttons to set the time.
  • Page 217 Chapter 7 - System Operation Preset General Figure 7-61: The General Preset Sheet Using the General preset sheet, users can configure the trackball behavior, volume, storage, screen saver/automatic freeze, and ECG lead select items. The detail description of this sheet is illustrated as below. Table 7-59: The General Configuration Items Item Description...
  • Page 218 Live Capture: Set Cine Storage time (1- 30s) Enable Screen Saver: set screen saver. Save Type: Set the save type (Black screen, Whale). Wait: Set the time of inactivity before entering screen saver. ECG LeadSelect: Select the lead. OK: press this button to save the set- tings and then exit the Preset dialogue window.
  • Page 219: Exam

    Chapter 7 - System Operation Preset 7.14.2 Exam Applications Before doing a medical examination, first of all, you should select an exam from the Applications or the More menus. Figure 7-62: The Application and the More menu Rev 1.00...
  • Page 220 Chapter 7 - System Operation Preset The exam on those menus corresponds to the display on the Applications di- alog. Figure 7-63: The ApplicationMenu Preset Sheet On the Applications dialog, all the exams of a body part are displayed on the All Applications list.
  • Page 221 Chapter 7 - System Operation Preset To add an exam to the More list, follow these steps: Open the Applications dialog (“Config/Diag” > “Exam” > STEP 1 “Applications”). On the Category drop-down list, select a body part. STEP 2 On the All Applications list, select the exam that you want STEP 3 to add to the Applications list, and then click the right arrow On the Applications list, select the body part which has...
  • Page 222 Chapter 7 - System Operation Preset Global Figure 7-64: The Global Preset Sheet Using the Global preset sheet, users can configure General, Doppler/M Mode and End Exam Setup items. The detail description of this sheet is illus- trated below. Table 7-60: The General Configuration Items Item Description Freeze Response: select the desired...
  • Page 223: Meas/Text

    Chapter 7 - System Operation Preset Table 7-60: The General Configuration Items Item Description Biopsy: set the Biopsy parameters. OK: press this button to save the set- tings and then exit the Preset dialogue window. Cancel: press this button to exit the Preset dialogue window without sav- ing the settings.
  • Page 224 Chapter 7 - System Operation Preset Table 7-61: The Meas Property Configuration Items Item Description Caliper Size: select the desired caliper size from the pop-up menu. Caliper Pattern: select the desired cali- per pattern. Result Font Color (Active): select the desired color of the active result font from the pop-up menu.
  • Page 225 Chapter 7 - System Operation Preset General Meas Figure 7-66: The General Meas Preset Sheet Using this General Meas preset sheet, users can configure the default mea- surement method by mode, and default measurement method. The detail description of this sheet is illustrated below. Table 7-62: The Meas Property Configuration Items Item Description...
  • Page 226 Chapter 7 - System Operation Preset Table 7-62: The Meas Property Configuration Items Item Description Default Measurement Method: select the default Measurement method. OK: press this button to save the set- tings and then exit the Preset dialogue window. Cancel: press this button to exit the Preset dialogue window without sav- ing the settings.
  • Page 227 Chapter 7 - System Operation Preset Meas Parameters Figure 7-67: The Meas Parameters Preset Sheet Using this Meas Parameters preset sheet, users can configure the Measure- ment and report parameters. The detail description of this sheet is illustrated as below. Table 7-63: The Meas Parameters Configuration Items Item Description...
  • Page 228 Chapter 7 - System Operation Preset Text Library Figure 7-68: The Text Library Preset Sheet Using this Text Library preset sheet, users can create a pre-defined library, add a user-defined text into a library, modify the selected text phrase, and delete the selected text phrase or all the text phrases.
  • Page 229 Chapter 7 - System Operation Preset Repeat step 6 and 7 to populate the Text Library. STEP 8 Click the right double arrow, all text phrases of the selected Avail- STEP 9 able Texts Library will be added into the selected Text Library. Reorder a text phrase inside the text items list of the selected Text Library.
  • Page 230 Chapter 7 - System Operation Preset Table 7-64: The Meas Parameters Configuration Items Item Description OK: press this button to save the set- tings and then exit the Preset dialogue window. Cancel: press this button to exit the Preset dialogue window without sav- ing the settings.
  • Page 231 Chapter 7 - System Operation Preset Table 7-65: The Meas Parameters Configuration Items Item Description Text Font Size: configures the text font size. Text Color: sets the text color in the sta- ble state. Word Wrapping: when checked, Text inputted can wrap. Arrow Size: sets the Arrow size.
  • Page 232 Chapter 7 - System Operation Preset Mark Library Figure 7-70: The Mark Library Preset Sheet Using this Mark Library preset sheet, users can create a pre-defined library, add a pre-defined mark into the library, and delete the selected mark. Users can easily select a pre-defined mark from the pre-defined library using Body Mark key.
  • Page 233: Patient

    Chapter 7 - System Operation Preset Remove a mark from the selected Mark Library Select the desired Mark Library from the combo box. STEP 1 Move the cursor to the marks list of the selected Mark Library. Click STEP 2 the mark and click Delete button.
  • Page 234 Chapter 7 - System Operation Preset Dataflows are configured in the DataFlow preset sheet in the Patient setup category as described below. Figure 7-71: The DataFlow Preset Sheet A set of pre-defined dataflows is available on the unit as listed in the table below.
  • Page 235 Chapter 7 - System Operation Preset Select the default dataflow Select the dataflow in the Name drop-down menu. STEP 1 Check the Default box. STEP 2 The dataflow will be selected by default after unit restart. STEP 3 Adjust the assigned devices Select the device in the Selected devices field.
  • Page 236 Chapter 7 - System Operation Preset Table 7-68: The Device Specific Parameters Device Settings Definition Retry Sets the maximum number of attempts to connect, the time interval between the attempts and the timeout. Patient Figure 7-72: The Patient Preset Sheet Using the Patient preset sheet, users can set the patient information, exam screen/report headings, some settings related to the patient windows, oper- ator and Ref.M.D.
  • Page 237 Chapter 7 - System Operation Preset Table 7-69: The Patient Configuration Items Item Description Title bar: select patient information from the title bar line, two-line display the pop-up menu. Anonymous Patient: show or hide patient information. Comments/Diagnoses/Referral Rea- sons: with this text box, user can enter text to change the text keywords? Birth date: whether the date of birth is need to create a patient.
  • Page 238 Chapter 7 - System Operation Preset Add operator and Ref.M.D Click Add button to enter the operator name in the text box (up to STEP 1 10 characters). Click OK button to complete the add operation and update patient STEP 2 information (select Cancel button to cancel the add operation).
  • Page 239 Chapter 7 - System Operation Preset Click Insert. STEP 2 Figure 7-74: Column Sub-menu On the Column sub-menu, select the title that you want to add STEP 3 to the heading row. The title will be displayed on the heading row. Modify a table title To modify a table title, follow these steps: On the heading row, click the title that you want to modify.
  • Page 240 Chapter 7 - System Operation Preset Report Enables the configuration of the Template selection menu and the export/ import of user-defined templates. Figure 7-75: The Report Preset Sheet Show Examination Categories report template Press the Config-Diag key on the control panel. STEP 1 In the Preset dialogue window, select Patient.
  • Page 241 Chapter 7 - System Operation Preset Click the up arrow. The report template is moved one step up. STEP 2 Click the down arrow. The report template is moved one step down. STEP 3 Delete a report template Select a report template. STEP 1 Click Delete button.
  • Page 242 Chapter 7 - System Operation Preset Related Topic: For instructions on how to add texts to the Findings and Diag- noses text boxes, see “7.2.3 Report”. Figure 7-76: Insert Text Dialog Add texts related to a root node To add texts related to a root node, follow these steps: Open the Insert Text dialog (“Config/Diag”...
  • Page 243: Network

    Chapter 7 - System Operation Preset On the Node list, discontinuously click the root node or the STEP 3 child node twice, and then name it. In the Findings or Diagnoses text boxes, type texts. STEP 4 Note: Only can the text of a indivisible node be displayed in the text boxes.
  • Page 244 Chapter 7 - System Operation Preset a Color video print and storage to a DICOM media). The Additional outputs sheet can also be used to setup the report printer. Figure 7-77: The AdditionalOutput Preset Sheet Configure the Print or Alt. Print key In Button field select Print or Alt.
  • Page 245 Chapter 7 - System Operation Preset Local TCP/IP Figure 7-78: The TCP/IP Preset Sheet Using this Local TCP/IP preset sheet, users can configure the TCP/IP settings and DICOM Local information. Table 7-71: The TCP/IP Configuration Items Item Description Host name: set the host name. Do not change the computer name.
  • Page 246: Peripheral

    Chapter 7 - System Operation Preset Table 7-71: The TCP/IP Configuration Items Item Description AE Title: set the AE title. Port: set the port number. PDU: set the protocol data unit. 7.14.6 Peripheral Printer Figure 7-79: The Printer Preset Sheet Using this Printer preset sheet, users can view all the printers and their work- ing state in the Device list, add a printer, delete a printer and configure a printer setting.
  • Page 247 Chapter 7 - System Operation Preset Click Add Printer Button, according to the operation instructions to STEP 2 complete adding the printer. Delete a printer Click the printer name to be deleted in the Device list. STEP 1 Click the Delete Printer button. STEP 2 Configure the printer settings Click the printer name to be configured in the Device list.
  • Page 248 Chapter 7 - System Operation Preset Output Figure 7-80: The Output Preset Sheet Using the Output preset sheet, users can setup the video output size and for- mat, and configure the screen settings. Table 7-72: The Output Configuration Items Item Description Output Size: select the type of the out- put image.
  • Page 249 Chapter 7 - System Operation Preset Table 7-72: The Output Configuration Items Item Description OK: press this button to save the set- tings and then exit the Preset dialogue window. Cancel: press this button to exit the Preset dialogue window without sav- ing the settings.
  • Page 250: Service

    Chapter 7 - System Operation Preset Table 7-73: The Output Configuration Items Item Description OK: press this button to save the set- tings and then exit the Preset dialogue window. Cancel: press this button to exit the Preset dialogue window without sav- ing the settings.
  • Page 251: Dicom

    Chapter 7 - System Operation DICOM 7.15 DICOM This section describes the communication and connection options for the Portable Ultrasound Diagnostic Scanner with other devices in the hospital information system. This sector covers the procedures to configure the opti- mal data management for the Portable Ultrasound Diagnostic Scanner in the following scenarios: A stand-alone Portable Ultrasound Diagnostic Scanner A Portable Ultrasound Diagnostic Scanner and a DICOM Server in a net-...
  • Page 252 Chapter 7 - System Operation DICOM Worklist - Local/ Dicom Storage: search the patient information in a DICOM Modality Worklist. Images are stored to the internal hard drive and to a DICOM server. 7.15.1.2 Dataflow Selection Select a dataflow from the Search/Create Patient window Press the Patient key on the control panel.
  • Page 253 Chapter 7 - System Operation DICOM If you select the Local-Local or Local-Local/Dicom Storage dataflow, the Search/Create Patient window will be displayed as below: Figure 7-84: Search/Create Patient window If you select the Worklist-Local or Worklist- Local/Dicom Storage data- flow. The Search/Create Patient window will be displayed as below: Figure 7-85: Search/Create Patient window To query the patient records from the worklist server If the worklist server is configured and connected correctly, click the...
  • Page 254 Chapter 7 - System Operation DICOM If the worklist server is not configured, click the Query button, a mes- sage box will be displayed. Click Yes button to close the message box. If the worklist server is configured and disconnected, click the Query button, a message box will be displayed.
  • Page 255 Chapter 7 - System Operation DICOM If this patient ID exists in the local database, the Warning win- dow will be displayed. Figure 7-86: Warning window Click Yes button, the Source information will be used in the Destination, and the Examination List window is displayed. Click Cancel button, the Source information will not be used in the Destination, and the Examination List window is dis- played.
  • Page 256 Chapter 7 - System Operation DICOM Select the Dicom prop. in the main menu, the PatientDicomProper- STEP 2 ty window is displayed. Figure 7-87: PatientDicomProperty window Click OK button to close this window. STEP 3 Configure the system with a default dataflow in the Preset dia- log window Press Config-Diag key on the control panel.
  • Page 257: Stand-Alone Ultrasound Unit Scenario

    Chapter 7 - System Operation DICOM Select the Patient category and DataFlow sub-level tab. The Data- STEP 2 Flow preset sheet is displayed as below: Figure 7-88: Default Dataflow Settings Select the desired dataflow in the Name pull-down menu and check STEP 3 the option Default.
  • Page 258: A Ultrasound Unit And A Dicom Server In A Network

    Chapter 7 - System Operation DICOM Image review Review images uses the same data stream in the system. 7.15.3 A ultrasound unit and a DICOM server in a net- work In this scenario Portable Ultrasound Diagnostic Scanner is configured to work with a DICOM server in a network environment.
  • Page 259 Chapter 7 - System Operation DICOM Select the Network category and Local TCP/IP subgroup. The Local STEP 2 TCPIP preset sheet is displayed as below. Figure 7-89: The Local TCP/IP Preset Sheet In the TCP/IP Config area, select DHCP or Static option to configure STEP 3 the IP address, Subway Mask, Gateway and DNS.
  • Page 260 Chapter 7 - System Operation DICOM 7.15.3.2 Setup of the DICOM server in the scanner's Preset window DICOM dataflow selection Press Config-Diag key on the control panel, select the Patient cate- STEP 1 gory, then select the DataFlow sub-group tab. The DataFlow preset sheet will be displayed.
  • Page 261 Chapter 7 - System Operation DICOM If the selected device is Worklist, the Worklist Property window will be displayed as below: Figure 7-91: The Worklist Property Window Rev 1.00...
  • Page 262 Chapter 7 - System Operation DICOM If the selected device is Dicom Storage, the Dicom Storage Proper- ty window will be displayed as below: Figure 7-92: The Dicom Storage Property Window Select the DICOM server from the IP-address pull-down menu. Fol- STEP 2 low the steps below if the IP address settings for the DICOM server need to be modified or created:...
  • Page 263 Chapter 7 - System Operation DICOM Enter the name and/or IP address of the server and click OK button to return to the Properties window. Figure 7-93: Modifying/Creating the IP address For each properties window, the user needs to set to the following STEP 3 requirements.
  • Page 264 Chapter 7 - System Operation DICOM Table 7-74: The different properties window configuration Window Item Description Port Input the DICOM server port. Image settings Whether to multi-frame or not, setup the Max Framer- ate.Compression and Qual- ity. Retry Setup the maximum num- ber and interval for connec- tion attempts.
  • Page 265 Chapter 7 - System Operation DICOM Click Add to list button. STEP 7 Click OK button to close the Search criteria window. STEP 8 Check the connection to the DICOM server In the Dataflow sheet, select the DICOM device to verify the connec- STEP 1 tion.
  • Page 266: Dicom Job Spooler

    Chapter 7 - System Operation DICOM Click Resend button. The images of the selected studies are STEP 3 transferred to the DICOM server. 7.15.4 DICOM JOB SPOOLER DICOM JOB SPOOLER displays the current DICOM output jobs. The jobs may be Storage, Print, Modality Performed Procedure Step or Storage Commit- ment.
  • Page 267 Chapter 7 - System Operation DICOM Start the DICOM JOB SPOOLER On the alphanumeric keyboard, press and hold down the ALT key and press S key. The DICOM JOB SPOOLER window is displayed as below: Figure 7-95: The DICOM JOB SPOOLER Window The DICOM JOB SPOOLER window is automatically updated every minute.
  • Page 268 Chapter 7 - System Operation DICOM Send a job to a new destination Trackball to the job to send in the DICOM JOB SPOOLER window. STEP 1 Press Set key on the control panel. STEP 2 Trackball to Send to button. STEP 3 Figure 7-96: The Send To...Window Press Set key on the control panel.
  • Page 269: Chapter 8 - User Maintenance

    Chapter 8 - User Maintenance CHAPTER 8 USER MAINTENANCE...
  • Page 270 This page is intentionally left blank.
  • Page 271: System Care And Maintenance

    Chapter 8 - User Maintenance System Care and Maintenance System Care and Maintenance This section covers: Planned maintenance System inspection Unit cleaning Prevention of the static electricity interference 8.1.1 Planned Maintenance The user is responsible for ensuring that ultrasound system is safe for daily use.
  • Page 272: System Inspection

    This method significantly reduces the risk of a virus attack on the system. Whale Company is continuously addressing the need for additional actions to reduce vulner- ability of equipment, including scanning vulnerability of the products and evaluating new security patches for the 3rd party technology used.
  • Page 273: Unit Cleaning

    Chapter 8 - User Maintenance System Care and Maintenance 8.1.3 Unit Cleaning Frequent and diligent cleaning of Portable Ultrasound Diagnostic Scanner can reduce the risk of infection spreading from person to person, and also help to maintain a clean working environment. To avoid the risk of electric shock, please turn the system off before CAUTION cleaning.
  • Page 274 Chapter 8 - User Maintenance System Care and Maintenance Cleaning The LCD Display Gently wipe the LCD display with a dry, soft, lint-free non-abrasive STEP 1 cloth. Note: In the event of a scratch-like mark on the screen, this may be a stain transferred from the keyboard or trackball when the LCD display cover was pressed from the outside.
  • Page 275: Static Electricity Interference Prevention

    Chapter 8 - User Maintenance Probe Care And Maintenance bing alcohol. Wipe the surface of keycaps with cloth, and ensure that no liquid drips on or between the keys. Cleaning The Peripherals Clean the peripherals in accordance with the respective manufacturers’ di- rections.
  • Page 276: Probe Inspection

    Chapter 8 - User Maintenance Probe Care And Maintenance After Each Use Inspect the probe Clean the probe Disinfect the probe Before Each Use Inspect the probe. 8.2.2 Probe Inspection After each use Inspect the lens, the probe housing and the cable. Look for damage that might allow liquid into the probe.
  • Page 277 Chapter 8 - User Maintenance Probe Care And Maintenance Table 8-1: Different probes cleaning and disinfecting requirements Probe Model Cleaning/Disinfection 4EC9V Cleaning+ High-level disinfection Probe Cleaning Disconnect the probe from the unit. STEP 1 Wipe the probe lens with a soft cloth to remove the coupling gel. STEP 2 Wipe the probe and cable with a soft cloth dampened with the rec- STEP 3...
  • Page 278 Chapter 8 - User Maintenance Probe Care And Maintenance Do not steam autoclave or subject the probe to Ethylene Oxide. WARNING Do not immerse the probe in liquid beyond the level specified for that WARNING probe. Never immerse the probe connector in liquid. WARNING The probe should not be exposed to the germicide longer than speci- WARNING...
  • Page 279: Chapter 9 - Components And Accessories

    Chapter 9 - Components and Accessories CHAPTER 9 COMPONENTS AND ACCESSORIES...
  • Page 280 This page is intentionally left blank.
  • Page 281: Components

    Chapter 9 - Components and Accessories Components Components Table 9-1: P5 Components Configuration Optional/Standard Console Standard Cart Optional Adapter Standard Probe (1P4V) Standard Probe (2P4V) Standard Probe (3P8V) Standard Probe (2C5V) Standard Probe (4EC9V) Standard Probe (12L5V) Standard Optional 3-ECG Lead Optional 4-ECG Lead Optional...
  • Page 282 Chapter 9 - Components and Accessories Components Table 9-2: P7 Components Configuration Optional/Standard Video Copy Processor (Printer B) Optional Table 9-3: P9 Components Configuration Optional/Standard Console Standard Cart Optional Adapter Standard Probe (1P4V) Standard Probe (2P4V) Standard Probe (3P8V) Standard Probe (2C5V) Standard Probe (4EC9V)
  • Page 283: Appendix I - Labels And Nameplates

    Appendix I - Labels and Nameplates APPENDIX I LABELS AND NAMEPLATES...
  • Page 284 This page is intentionally left blank.
  • Page 285: I-1 Safety Labels

    Appendix I - Labels and Nameplates ComponentsSafety labels Safety labels The following table describes the purpose of safety labels: Table I-1: Symbols used on equipment Symbols Description Location Warning: crush- On the surface of the cart’s ing hazard-foot rear wheel Warning: step- On the upper pallet of ping prohibited...
  • Page 286 Appendix I - Labels and Nameplates ComponentsSafety labels Table I-1: Symbols used on equipment Symbols Description Location Warning: dan- Power output port of cart gerous voltage Probe label On the surface of Trans- ducer connector(1P4V) Probe label On the surface of Trans- ducer connector(2P4V) Probe label On the surface of Trans-...
  • Page 287 Appendix I - Labels and Nameplates ComponentsSafety labels Table I-2: Nameplates Nameplates Location Located inside the groove of back cover of the scanner (The nameplate is identified by differ- ent electric current and applies to P5, P7 and P9). Rev 1.00...
  • Page 288 Appendix I - Labels and Nameplates ComponentsSafety labels Table I-2: Nameplates Nameplates Location Located on the back cover of the lifting column Located on the back cover of the lifting column Portable Ultrasound Diagnostic Scanner User Manual...
  • Page 289 Appendix I - Labels and Nameplates ComponentsSafety labels Table I-2: Nameplates Nameplates Location Located on the enclosure of digital graphic printer Located on the enclosure of the video copy proces- Rev 1.00...
  • Page 290 Appendix I - Labels and Nameplates ComponentsSafety labels Table I-2: Nameplates Nameplates Location Located On the packaging of the ECG electrode lead wire Portable Ultrasound Diagnostic Scanner User Manual...
  • Page 291 Appendix I - Labels and Nameplates ComponentsSafety labels Table I-3: Symbols used on nameplates Symbols Description Location Caution On the nameplates of the Portable Ultrasound Diag- nostic Scanner Manufacturers On the nameplates of the Portable Ultrasound Diag- nostic Scanner Serial number On the nameplates of the Portable Ultrasound Diag- nostic Scanner...
  • Page 292 This page is intentionally left blank.
  • Page 293: Appendix Ii - Acoustic Output Report Table

    Appendix II - Acoustic Output Report Table APPENDIX II ACOUSTIC OUTPUT REPORT TABLE...
  • Page 294 This page is intentionally left blank.
  • Page 295: Ii-1 Acoustic Output Report Table

    Appendix II - Acoustic Output Report Table ComponentsAcoustic Output Report Table ii-1 Acoustic Output Report Table Besides this user’s manual, the related document is shipped together with the product, as below: Table II-1: Acoustic Output Report Table Doc Number P-12-025(1) Version Rev 1.00...
  • Page 296 This page is intentionally left blank.
  • Page 298 Fax: +1 781 609 2164 Fax: +32 2732 6003 Fax: +86 10 6789 2105 © 2017 All rights reserved. No part of this user manual can be reproduced or transmitted without the permission of Beijing East Whale Imaging Technology Co., Ltd.

Table of Contents