Whale PE-12-025 User Manual page 264

Portable ultrasound diagnostic scanner
Table of Contents

Advertisement

Chapter 7 - System Operation
Table 7-74: The different properties window configuration
Window
Adjust the Search criteria
STEP 1
STEP 2
STEP 3
STEP 4
STEP 5
STEP 6
264
Press Config-Diag key on the control panel, open the Preset dialog
window. Select the Patient category, then select the DataFlow sub-
group tab. The DataFlow preset sheet will be displayed.
Select a DICOM Worklist dataflow.
Select the Worklist device in the Selected devices pane and click
Properties button.
Click the Search criteria button. The Search criteria window will be
displayed.
Figure 7-94: The Search Criteria Window
Select a search criteria from the Select tag pull-down menu.
Enter a value if required or leave blank if not to be used. This entry is
case sensitive and must match exactly.
Item
Port
Image settings
Retry
Timeout
Portable Ultrasound Diagnostic Scanner User Manual
Description
Input the DICOM server
port.
Whether to multi-frame or
not, setup the Max Framer-
ate.Compression and Qual-
ity.
Setup the maximum num-
ber and interval for connec-
tion attempts.
Input the timeout.