Whale PE-12-025 User Manual page 135

Portable ultrasound diagnostic scanner
Table of Contents

Advertisement

Imaging Mode
Table 7-17: Color Mode Soft Menu Controls
Parameter
Invert
Power
Other Color Mode Controls
Table 7-18: Other Color Mode Controls
Parameter
Gain
TGC
Depth
1/2/4
Alt
Using Color Mode
STEP 1
STEP 2
STEP 3
STEP 4
STEP 5
STEP 6
STEP 7
Rev 1.00
Description
Inverts the color scheme for positive and negative veloci-
ties. Invert is available in Live and Freeze.
Controls the amount of acoustic power applied in all
modes. Acoustic power cannot be set above the amount
permitted by the FDA. The Thermal index (TI) and the
Mechanical index (MI) are displayed on the screen.
Description
Adjusted by the knob below the TGC on the control panel.
When rotated clockwise, increases the overall gain applied
to the received echo signals equally for all depth.
Compensates for depth-related attenuation in an image.
The sliders nearest the operator affect the far field. TGC
amplifies returning signals to compensate for the attenua-
tion caused by tissue at increasing depths.
Adjusting the depth will influence diagram frame rate and
resolution. Decreasing the depth may allow for higher
frame rate and lower resolution. Increasing the depth will
increase image resolution and decrease frame rate.
Enables multiple image display windows in which two or
four images can be viewed simultaneously.
Display more Main menu controls: Simultaneous, TVI
(phased array probe), TDI (phased array probe), EI (linear
array probe and convex array probe), SRI (phased array
probe).
While in 2D mode, press CFM key on the control panel.
Use the trackball to position the Color ROI over the area to be exam-
ined.
Rolling the trackball to adjust the position of the Color ROI.
To adjust the Color ROI size, press Set key and use the trackball.
To enlarge or narrow the Color ROI, roll the trackball to the left or
right.
To lengthen or shorten the Color ROI, roll the trackball up or down.
Press Set key on the control panel when the desired size is obtained.
Reposition the Color ROI as needed.
Press Freeze key on the control panel to stop imaging.
To quit Color Doppler imaging, please press CFM or 2D key.
Chapter 7 - System Operation
135