Whale PE-12-025 User Manual page 193

Portable ultrasound diagnostic scanner
Table of Contents

Advertisement

Measurement
Table 7-52: Cardiac Measurement Item In 2D Mode
Item
LVIDs
Teichholz
Diastole
RVAWd
RVDd
IVSd
LVIDd
LVPWd
Systole
IVSs
LVIDs
LVPWs
As shown in the table above, the measurement Item mainly includes Dis-
tance methods and Trace/Ellipse methods.
Operations of distance method measurement are:
STEP 1
STEP 2
STEP 3
STEP 4
STEP 5
Rev 1.00
Description
Left Ventricular Internal
Diameter in Systole
Diastole
Right ventricular anterior
wall thickness at end systole
Right ventricular dimension
at end systole
Interventricular septum
dimension at end diastole
Left ventricular internal
dimension at end diastole
Left ventricular posterior wall
dimension at end diastole
Cardiac systole
Interventricular septum
dimension at end systole
Left ventricular internal
dimension at end systole
Left ventricular posterior wall
dimension at end systole
Press Application key on the control panel, then select Cardiac, and
enter 2D mode.
Press Measure key on the control panel and select MV measure-
ment Item in the soft menu (using mitral valve as example), and
then click MV to display the measurement items for mitral valve.
Put the first caliper on the starting point of line segment, Press Set
key on the control panel.
Roll the trackball to put the second caliper on the end point of line
segment, Press Set key on the control panel.
The measurements will be displayed in the Measurement result ta-
ble, and saved to the worksheet and report.
Chapter 7 - System Operation
Measurement methods
Trace/Ellipse
Distance
Distance
Distance
Distance
Distance
Distance
Distance
Distance
193