Electrical Hazards; Biological Hazards; Pacemaker Hazards - Whale PE-12-025 User Manual

Portable ultrasound diagnostic scanner
Table of Contents

Advertisement

Personnel And Equipment Safety
2.10.2

Electrical Hazards

To Avoid Injury:
Fuses blown within 36 hours of being replaced may indicate a malfunction-
ing electrical circuit within the system. In this event, the unit must be
checked by Whale Ultrasound service personnel, No attempt should be
made to replace the fuse with another fuse of a higher rating.
2.10.3

Biological Hazards

To avoid the risk of disease transmission:
2.10.4

Pacemaker Hazards

The possibility of the system interfering with pacemakers is minimal. How-
ever, as the system generates high frequency electrical signals, the operator
should be aware of the potential hazard this could cause.
Rev 1.00
The internal circuits of the unit use high voltages, capable of causing
WARNING
series injury or death by electrical shock.
Do not remove the unit's protective covers. No user-serviceable parts
are inside. If servicing is required, contact qualified technical personnel.
Connect the attachment plug to a hospital-grade ground outlet to en-
sure adequate grounding.
Do not put liquids on or above the unit. Conductive fluids seeping into
the active circuit components may cause short circuiting, which could
result in an electrical fire.
An electrical hazard may exist if any light, monitor or visual indicator re-
mains on after the unit is turned off.
Use protective barriers whenever necessary. Follow sterile procedures
as required.
Thoroughly clean probes and reusable accessories after each patient
examination and disinfect to sterilize as needed.
Follow all in-house infection control policies if they apply to personnel
and equipment.
Chapter 2 - Safety
25

Advertisement

Table of Contents
loading

Table of Contents