Chapter 8 - User Maintenance
278
Do not steam autoclave or subject the probe to Ethylene Oxide.
WARNING
Do not immerse the probe in liquid beyond the level specified for that
WARNING
probe.
Never immerse the probe connector in liquid.
WARNING
The probe should not be exposed to the germicide longer than speci-
WARNING
fied to achieve the desired effect.
Do not soak or saturate probes with solutions containing alcohol,
WARNING
bleach, ammonium chloride compounds.
Neurological use on patients must be avoided. If a probe becomes con-
WARNING
taminated, there is no adequate means of disinfection.
List of cleaning agent and disinfections
Clean Agent: 3M Epizyme multi enzyme cleaner (1:100 diluted by wa-
ter)
LLD Agent: 70% isopropanol
HLD Agent: Cidex OPA
Portable Ultrasound Diagnostic Scanner User Manual
Probe Care And Maintenance