Color Mode - Whale PE-12-025 User Manual

Portable ultrasound diagnostic scanner
Table of Contents

Advertisement

Imaging Mode
Table 7-15: Other Controls of CM Mode
Parameter
TGC
Depth
1/2/4
Alt
7.4.6

Color Mode

Color Mode Overview
The Color Doppler Mode allows the user to observe blood flow using Dop-
pler multiple samples.
Rev 1.00
Description
Compensates for depth-related attenuation in an image.
The sliders nearest the operator affect the far field. TGC
amplifies returning signals to compensate for the attenua-
tion caused by tissue at increasing depths.
Adjusting the depth will influence diagram frame rate and
resolution. Decreasing the depth may allow for higher
frame rate and lower resolution. Increasing the depth will
increase image resolution and decrease frame rate.
Enables multiple image display windows in which two or
four images can be viewed simultaneously.
Display more Main menu controls: Simultaneous, TVI
(phased array probe), TDI (phased array probe), EI (linear
array probe and convex array probe), SRI (phased array
probe).
Figure 7-27: Color Mode Screen
1
Probe orientation marker
2
Color bar
Chapter 7 - System Operation
133

Advertisement

Table of Contents
loading

Table of Contents