Karl Storz AUTCON III 400 Instruction Manual page 90

Table of Contents

Advertisement

84
Fehler erkennen und
beheben
Header
Message text
Confirmation of
You have selected an AUTOSTART mode.
AUTOSTART mode
Setting the AUTOSTART mode can result in unintentional coagulations, e.g. when bipolar forceps are used for gripping while the
AUTOSTART mode is on.
AUTOSTART Fault
The instrument is in contact with tissue.
AUTOSTART cannot be selected when the instrument is in contact with tissue. Open up the instrument.
TSI Warning
The annual Technical Safety Inspection (TSI) is due.
Neutral Electrode
No neutral electrode connected. No neutral electrode is connected. Connect a neutral electrode.
Fault
Neutral Electrode
Wrong neutral electrode connected.
Fault
The selection does not match the neutral electrode connected. Connect the neutral electrode which matches the selected mode, or
change the mode to match the neutral electrode.
Neutral Electrode
Poor contact with the patient.
Fault
The resistance between the neutral electrode and the patient is too high. Establish better contact of the neutral electrode.
Neutral Electrode
Poor contact with the patient.
Warning
The contact resistance between the neutral electrode and the patient is getting worse. Establish better contact of the neutral
electrode.
Neutral Electrode
No cable for neutral electrode connected.
Fault
No neutral electrode cable is connected. Connect a neutral electrode.
Neutral Electrode
No cable for neutral electrode connected.
Notice
The cable for neutral electrode has been removed. Unipolar activation is not possible.
Mode Fault
No mode selected. No mode was selected for this type of activation. Select the desired mode or change the footswitch assignment.
Mode Fault
This mode is not allowed for baby neutral electrodes. Use split neutral electrodes with a large surface and the suitable mode for this.
Mode Fault
This mode is not allowed for this socket. The current mode remains active.
Choose another socket for this mode.
Footswitch Fault
No compatible footswitch. The connected footswitch is not compatible with this device. Connect a compatible footswitch with a
changeover switch.
Footswitch Fault
Footswitch not assigned to a socket.
The footswitch has not been assigned to a socket yet.
Assign a socket to the footswitch using the 'Pedal' button.
Footswitch Fault
Error on footswitch connection.
Check the footswitch. If this message appears again, please contact the Technical Support. Contact: MENU – SERVICE.
Recognizing and remedying
errors
The following table describes the cause of the
error and the appropriate corrective action.