Karl Storz AUTCON III 400 Instruction Manual page 232

Table of Contents

Advertisement

226
EMV
Recommended separation distances between portable and mobile HF communications equipment and
The AUTOCON
®
 III 400 is intended for use in an electromagnetic environment in which the HF disturbances are controlled. The customer or user of the
AUTOCON
®
 III 400 can help to prevent electromagnetic interference by maintaining a minimum distance between portable and mobile HF communications
equipment (transmitters) and the AUTOCON
Rated maximum output power of
Separation distance d according to frequency of transmitter (m)
transmitter (W)
150 kHz­to­80 MHz
f
0.01
0.12
0.1
0.38
1
1.2
2
1.7
10
3.8
100
12
For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection by structures, objects
and people.
EMC
the AUTOCON
®
 III 400 (IEC 60601-1-2, Table 6)
®
 III 400 as recommended below, according to the maximum output power of the communications equipment.
80 MHz­to­800 MHz
f
0.12
0.38
1.2
1.7
3.8
12
CEM
800 MHz­to­2.5 GHz
h
0.23
0.73
2.3
3.3
7.3
23