COMMISSIONING

7.3.4.15 Maintenance

Name
[Wire tension
protection]
[Drum slip
detection]
[Drum slip
occurrence]
[Motor position
errors]
[Set to
Maintenance mode]
[Operational mode]
[Find reference
encoder position]
[Frequency 0]
[Frequency 1]
[Frequency 2]
[Frequency 3]
[Measure
frequency]
Part No.: 4417340 Revision 1
Value Range
Default Value
<enable>
<enable>
<disable>
<enable>
<0.0>
<0.0>
<0.0>
<0.0>
Service Manual
954 SmartServo FlexLine
Explanation
With this entity the possibility
exists to enable or disable the
wire tension high and low check.
If enabled a high or low fail code
will be generated if the tension
exceeds the low or high
thresholds.
With this entity the possibility
exists to enable or disable the
drum slip detection. If enabled a
drum slip fail code will be
generated if the measured
frequency is below a certain
limit.
Shows the number of drum slip
occurrences
Shows the number of observed
slip detections of the reference
encoder
This command resets the Servo
board and after that enters the
maintenance mode. In this mode
there are no level
measurements. Now the 4
frequencies can be measured or
the find reference encoder can
be done. A limited number of
commands can be issued.
Shows the mode the gauge is in:
Operational or Maintenance
This command determines the
reference encoder position and
synchronizes this with position.
Only available in Maintenance
mode.
Frequency measured at resp.
25, 100, 175 and 250 g. load on
the wire.
This command will measure the
frequency with the actual
displacer/weight. The result is
placed in entity [Measured
frequency].
89

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents