10.2 Maintenance and maintenance intervals at
a glance
Using the list below and the other sections, check the possible ranges of con-
sumables and the presence of any unusual damage that may have occurred since
the time of installation and commissioning. Equipment cleaning and documen-
tation of the work performed complete the maintenance.
DANGER
Incorrect or non-performed maintenance work
Health and property damage as well as incorrect measurement results
are possible
The operator must regularly check all safety devices for operability, if ne-
cessary, also outside the maintenance interval. The provisions for ex-
plosion protection (e.g. per IEC EN 60079-14) must be observed.
DANGER
Travaux de maintenance incorrects ou non exécutés
Des dommages à la santé et à la propriété ainsi que des résultats de
mesure incorrects sont possibles.
L'opérateur doit vérifier régulièrement le fonctionnement de tous les dis-
positifs de sécurité, si nécessaire également en dehors de l'intervalle de
maintenance. Les dispositions relatives à la protection contre les ex-
plosions (par exemple selon la norme CEI EN 60079-14) doivent être re-
spectées.
Not all maintenance work, for example replacing the battery, can be carried out by
the user. The following overview lists all possible maintenance work and intervals.
This monitoring can also be performed more frequently in order to detect de-
viations at an early stage. If necessary, please arrange a service appointment.
Visual inspection. Fill in the "First inspection"
l
form one month after commissioning.
Close-up inspection(s) every twelve months. Document results (e.g. use "Ser-
l
vice report template"
formed
(p. 285))
Complete a service calibration see
l
(p. 273)every twelve month.
GasLab Q2
10.6.2 Template for documentation of the work per-
➪  
10.3.11 Completing a service calibration
➪  
User Manual (N )
10.6.1 First inspection
➪ 
(p. 284)