Honeywell FlexLine SmartServo 954 Service Manual page 137

Servo-based level gauge
Hide thumbs Also See for FlexLine SmartServo 954:
Table of Contents

Advertisement

COMMISSIONING
Name
[Element type]
[Number of elements]
[Lowest element offset]
[RTD element positions]
Name
[Element type]
[Number of elements]
[Lowest element offset]
[RTD element positions]
Part No.: 4417340 Revision 1
Product Temperature
The following entities must be set by Engauge or HART
SmartView for a correct functioning of the FII-VT module in an
instrument.
NOTE: Using Engauge, following entities are set within the Engauge
Product temperature tab.
Explanation
The supported element
type is: SPL
The number of elements a
RTD probe has
The distance from tank zero
till the lowest position of the
multi-RTD probe.
So not used for the 3-wire
version and it must be 0.
The positions of the RTDs
from tank zero level
Explanation
The supported element
type is : SPL
The number of elements a
RTD probe has
The distance from tank zero
till the lowest position of the
multi-RTD probe
The positions of the RTDs
from the lowest position of
the probe.
So not tank zero level!
Service Manual
954 SmartServo FlexLine
RTD 3 spots (see
Figure 7-19
Value Range
3 characters
<SPL>
<1 .. 3>
floating point number:
<-x.x .. +x.x>
MUST BE
<0.0>
3 floating point numbers:
<-x.x .. +x.x>
Note: Only actually used
elements to be entered.
Multi RTD probe (see
Value Range
3 characters
<SPL>
<1 .. 14>
floating point number:
<-x.x .. +x.x>
14 floating point numbers:
<-x.x .. +x.x>
Note: Only actually used
elements to be entered.
left side)
Default
<--->
<0>
<80.0>
<0,0,0>
Figure 7-19
right side)
Default
<--->
<0>
<80.0>
<0,0,0,0,0,0,0,0,0,0,0,0,0,0>
129

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents