COMMISSIONING
7.3.6.5.3
7.3.6.6

LED Association

7.3.6.7

Terminal Allocation

Part No.: 4417340 Revision 1

Acknowledge

The
!Acknowledge!
command will cause the specified relay to be
physically deactivated, but the relay status will be set to
!Acknowledged!. This command is only available in
see 7.3.6.4.1. This command can only be given when the concerned
relay has already been activated.
For example, this command is useful if the relay is connected to an
alarm system. The alarm can be silenced by the command, but it is still
possible to determine if an alarm has occurred. When the alarm
condition is then removed, the relay state will become deactivated, and
normal operation will resume.
Depending on the value set in the
individual relays on the FII-DO board can be associated with one out of
2 LEDs, LE2 or LE3 (see
The associated LED will be ON when the relay state is <Activated>, and
the LED will be OFF when the relay state is
NOTE: The LEDs do not indicate the physical relay state (coil state or
contacts state), as this depends on the software settings and the
physical settings (jumper), see Relay Configuration
Set the value of the
For an example, see table below.
LED Association
1 + 2
3 + 4
Terminal Number
14
15
16
17
18
19
20
21
*) See jumper settings, section
Service Manual
954 SmartServo FlexLine
[LED Association]
Figure
7-7).
<Deactivated>.
[LED Association]
entity as desired (optional).
LED Number
LE2
LE3
LE2
LE3
Name
Ry1_a
Ry1_b
Ry2_a
Ry2_b
Ry3_a
Ry3_b
Ry4_a
Ry4_b
7.3.6.3.1
[PV Monitor]
mode;
entity, the 4
7.3.6.3.
Relay Number
1
2
3
4
Function
Relay 1 Common
Relay 1 NO or NC*
Relay 2 Common
Relay 2 NO or NC*
Relay 3 Common
Relay 3 NO or NC*
Relay 4 Common
Relay 4 NO or NC*