Example Of A Smooth Servo Unbalance Diagram (16 Points) - Honeywell FlexLine SmartServo 954 Service Manual

Servo-based level gauge
Hide thumbs Also See for FlexLine SmartServo 954:
Table of Contents

Advertisement

MAINTENANCE
Entity name
[Balance test result
unbalance]
[Balance test result
average weight]
[Balance test result
maximum weight]
[Balance test result
minimum weight]
[Balance test result
average frequency]
[Measured bearing
status]
Figure 9-1
Part No.: 4417340 Revision 1
The balance test results in several quality indicators that can be viewed
in Engauge and partly on the SmartView.
Entity name (SmartView)
[Drum unbalance [kg]]
(Product level)
[Avg meas weight [kg]]
(Product level)
[Max unbal weight [kg]]
(Product level)
[Min unbal weight [kg]]
(Product level)
[Avg meas freq [kg]]
(Product level)
-
The result of the balance test can also be viewed graphically by the
diagram viewer in Engauge (Servo Unbalance Diagram).

Example of a smooth servo unbalance diagram (16 points)

Figure 9-1 shows an example of a balance test diagram of 16 points.
The blue line shows the actual unbalance table in [Hz], while the red
lines show the measured hysteresis.
Service Manual
954 SmartServo FlexLine
Description
The drum unbalance should be within 3
grams. When the drum unbalance is more,
check for contamination of the drum. If the
unbalance is still more than 3 grams replace
the drum bearings (refer to section 9.1.2).
Refer to section
9.1.2
when the displacer's
weight differs more than 3 g from the value in
entity [Displacer weight].
Refer to section
9.1.1
259