COMMISSIONING
7.3.9.7

Hardware Configuration

7.3.9.7.1
Jumper
Position
Number
JP1
ON
JP2
ON
JP3
ON
JP4
ON
JP5
ON
JP6
ON
JP7 & JP8
A
JP7 & JP8
P
7.3.9.7.2
158

Jumper Allocation

The following are typical jumper settings done on the HCI-HAO board.
Connection
Description
Details
Short 2 & 3
W&M Entity Protection
Short 2 & 3
Password Read Protection
Short 2 & 3
Write Protection All Entities
Short 2 & 3
Free
Short 2 & 3
Free
Short 2 & 3
CAN termination 120E resistor
Short 1 & 2
Active mode for Analog Output
Short 2 & 3
Passive mode for Analog Output

LED Allocation

LED
Function
Number
LE2
HART data Transmit
LE3
HART data Receive
Service Manual
954 SmartServo FlexLine
Default
Default
Position
Connections
OFF
Short 1 & 2
OFF
Short 1 & 2
OFF
Short 1 & 2
OFF
Short 1 & 2
OFF
Short 1 & 2
OFF
Short 1 & 2
A
Short 1 & 2