7.3.10.7 Firmware Upgrade of HCI-TRL/2 or FCM-TRL/2 through CAN-SD Card

Part No.: 4417340 Revision 1
Follow the steps to upgrade firmware for HCI-TRL/2 or FCM-TRL/2
through CAN-SD card.
1. Insert the card in a card reader and connect it to PC.
2. Create a folder in the card and name it as "Firmware".
3. Add/Copy the bin format file to the card inside the Firmware
folder.
For Example: 0_A1001.025
4. Remove the card from the PC and insert in CAN-SD (TII-SD)
card.
5. Connect the card to the gauge.
6. During Firmare upgrade the following process occurs:
The health LED blinks indicating the health status. After few seconds
Amber light (LED 2) blinks indicating the firmware upgrade process.
Once the upgrade starts the card will go to bootloader state.
After upgrade is completed, the card gets back to normal state.
7. The Firmware is upgraded for HCI-TRL/2 or FCM-TRL/2
through CAN- SD card