MAINTENANCE

Balance test

Entity name
[Balance test number of
test points]
Entity name
[Lock Test]
[Balance test]
[Unlock]
258
The balance test measures the unbalance and quality of the bearings
over one drum rotation over a configurable number of points. The
resulting unbalance data is used for the unbalance compensation for
both level and density when enabled (refer to sections
).
Entity name
Value
(SmartView)
Range
[BT nr test
2-32
point]
(Product
level)
The measuring drum unbalance can be measured by the following
procedure:
Entity name
(SmartView)
[Lock Test]
(Product level)
[Balance test]
(Product level)
[Unlock]
(Product level)
Service Manual
954 SmartServo FlexLine
Default
Explanation
Value
If necessary, change the number
8
of test points.
Note: When the number has been
changed to a different value the
TII-SRV board needs a reset
before the new setting becomes
active.
Description
The Lock test raises the displacer to the
programmed [lock test motor limit switch level]
position. Then the gauge goes into block mode
(BL). If [Lock test motor limit switch level] is set
higher than [Motor limit switch high] the displacer
stops at the programmed [Motor limit switch
high] position and stays in lock test mode (LT).
With this command the balance of the measuring
drum is checked. This measurement takes
approximately 5 minutes.
When the balance test is ready, the status on the
display changes from BT to FR and the displacer
level primary value status code changes from
'Balance Test command active' into 'Freeze
command active'.
Cancel the balance test / freeze command.
Part No.: 4417340 Revision 1
7.7.1
and
10.2.3

Hide quick links: