Honeywell FlexLine SmartServo 954 Service Manual page 297

Servo-based level gauge
Hide thumbs Also See for FlexLine SmartServo 954:
Table of Contents

Advertisement

TROUBLE SHOOTING
Status code
(SmartView)
6191
6208
6209
6210
6211
6212
Part No.: 4417340 Revision 1
Description
(Engauge)
Encoder LED pin malfunction
Wire tension too high
Wire tension too low
Density displace declined
Boot self test failed
Density license not available
Service Manual
954 SmartServo FlexLine
Remarks
Hardware error. Malfunction of the LED
of the reference encoder.
The wire tension is too high. The gauge
will resume functioning when the wire
tension issue is resolved (by e.g. moving
the displacer down). Refer to sections
7.7.3
and 11.1.
The wire tension is too low. The gauge
will resume functioning when the wire
tension issue is resolved (by e.g. moving
the displacer up). Refer to sections
and
11.1
7.7.3
289