Honeywell FlexLine SmartServo 954 Service Manual page 84

Servo-based level gauge
Hide thumbs Also See for FlexLine SmartServo 954:
Table of Contents

Advertisement

COMMISSIONING
Name
[Block mode]
[Block]
[Unlock]
[Lock test]
[Lock test limit level]
[Calibrate]
76
Value Range
<continuous>
<non continuous>
Service Manual
954 SmartServo FlexLine
Default Value
Explanation
<continuous>
With this entity the user can
select whether the Block
command
remains active after the selected
interface has reached the
displacer.
Set to continuous the Block
command will
remain active until another
operational command with
sufficient priority
interrupts the Block command.
The displacer will follow the
interface.
Set non continuous the Block
command will also
be canceled if the product
reaches the displacer.
This command stops the level
measurement,
See also Block mode
This command is used to break
off any active operational
command.
After an Unlock command the
954 returns to the default
operation mode.
This command will raise the
displacer up till it reaches [Motor
limit switch high] or [Lock test
limit level].
If motor limit high reached, the
displacer remains there (until
e.g. Unlock has been given)
If lock test limit reached, the
command mode will be changed
to Block.
<50.0>
Level at which the Lock test
command will stop the displacer
and the command mode will
change to Block.
This command will raise the
displacer up until it reaches the
calibration plate. This position
(innage level) will be compared
with [Tank top]. If the difference
is too big a fail is issued.
Part No.: 4417340 Revision 1