Commissioning Of Tii-Lcd - Honeywell FlexLine SmartServo 954 Service Manual

Servo-based level gauge
Hide thumbs Also See for FlexLine SmartServo 954:
Table of Contents

Advertisement

COMMISSIONING

7.3.11.2 Commissioning of TII-LCD

Name
[Level type]
[Display contrast]
[Identification]
[Temperature Units]
[Zero format]
[Level units]
[Decimal Separator]
[Tenth millimeter
selection]
190
For correct functioning of the TII-LCD module in the instrument, the
following entities can be set by using either Engauge or HART
SmartView.
Value Range
<Innage>
<Ullage>
<0...63>
2 digits
8 characters
E.g. Tank_123
<Celsius>
<Fahrenheit>
<format Numerical 0>
<format letter O>
<Metres>
<Feet>
<Inches>
<Fraction>
<Point>
<Comma>
<enable>
<disable>
Service Manual
954 SmartServo FlexLine
Default Value
Explanation
<Innage>
It determines level
value to be displayed
on the LCD is Innage
Type or Ullage Type
<35>
Used to set the
contrast of the LCD
<-------->
Name of the tank or
instrument. This tank
name will be displayed
during start up and
showing of initial
screen is enabled. Only
special character
allowed is "_"
<Celsius>
Temperature value is
converted based on the
Temperature unit
selected and displayed
on the LCD
<format
Zero of Process value
Numerical 0>
Level and Temperature
is displayed as per this
configuration
<Metres>
Based on the level unit
selected process value
read from servo is
converted and
displayed on the LCD
<Point>
Based on this
configuration decimal
separator is displayed
on the LCD screen
<enable>
Determines whether
the tenth millimeter is
shown on the display in
the [PV screen] in case
a level entity is shown.
Part No.: 4417340 Revision 1