Optional accessories

Calibration adapter

The calibration adapter for the instrument is a simple 6-inch Tygon tubing. During calibration,
simply insert thetubing into the regular gas inlet probe of the instrument and to the gas
regulator on the gas bottle.

Calibration regulator

The Calibration Regulator is used in the calibration process. It regulates the gas flow rate from
the Span gas cylinder into the gas inlet of the instrument during calibration process. The
maximum flow rate allowed by the flow controller is about 0.5L/min (500 cc per min.).
Alternatively, a demand-flow regulator or a Tedlar gas bag may be used to match the pump flow
precisely.

Organic vapor zeroing kit

The Organic vapor zeroing kit is used for filtering organic air contaminants that may affect the
zero calibration reading. To use the Organic vapor zeroing kit, simply connect the filter to the
inlet port of the instrument.
AutoRAE 2 automatic test & calibration system
The AutoRAE 2 Automatic Test and Calibration System/portable gas monitors makes
compliance with monitor test and calibration requirements as easy as pressing a button. Simply
cradle the monitor and the system will take care of all calibration, testing, and recharging.
The AutoRAE 2 is a flexible, modular system that can be configured to meet your calibration
requirements effectively and efficiently. An AutoRAE 2 system can be as simple as a single
cradle deployed in standalone mode to calibrate one instrument at a time, or as powerful as a
networked, controller-based system supporting ten monitors and five distinct calibration gas
cylinders.