3.3.7
Vertical Scale Units
Description
►
Change the vertical units for both linear or logarithmic scales.
Operation
►
Press
The units are changed accordingly.
Units: dBm, dBmV, dBuV, Watts, Volts
3.3.8
Reference Level Offset
Description
►
The Reference Level Offset function sets an offset value to the reference level
to compensate for any loss or gain from an external network or device.
The offset value does not affect the input attenuation or the on-screen trace.
This setting will change the reference level readout, the scale readout and the
marker readout.
Operation
►
Press
unit.
To remove the offset level, set the reference offset to 0 dB.
Range: 0 dB ~ 50 dB
Display Icon
►
The AMP icon is displayed at the bottom of the screen.
31
> More[F7] > Y Axis[F1] and then choose the desired units.
> More[F7] > RefLvlOffset[F4] and set the offset level and
February 2017