Operation: Setting up the main channel
Press
Any other measurement mode will automatically be disabled.
The display splits into two screens. The top screen shows the main
channel, adjacent channels and their corresponding limits. The bottom
screen shows the ACPR measurement results in real time.
Press
following:
Main CHBW[F1]: Set the bandwidth of the main channel.
Main CH H Limit[F2]: Set the low limit for the main channel.
Main CH Limit[F3]: Set the high limit for the main channel.
CH SPC[F4]: Specify the channel spacing
Operation: Setting up the adjacent channel(s)
Press ADJCH Setup[F5] to setup the adjacent channels:
Select AdjCh[F1]: Choose an adjacent channel number: 1, 2, 3
[F2]: Toggle the selected channel on/off.
ADJCHBW[F3]: Choose the bandwidth of the selected channel.
ADJCH Offset[F4]: Set the adjacent channel offset.
ADJCH HLimit[F5]: Set the adjacent channel high limit.
ADJCH LLimit[F5]: Set the adjacent channel low limit.
Repeat the above steps for the other adjacent channels, if needed.
February 2017
> Channel Analysis[F1] > ACPR[F2] and turn ACPR on.
Figure 71.
> Channel Analysis[F1] > ACPR Setup[F1] > and set the