Display
Tip
Signals that are masked by the noise floor level should have a ratio of less
than 1 to smooth the noise out.
Signals with strong frequency components should use a ratio equal to or
greater than 1.
3.5.4

Average Trace

Description
The Average function averages the trace for a user-defined number of times
before it is displayed. This feature smoothes the noise level, but has the
drawback of slowing down the display update rate.
Operation
Press
Set the number of averages.
Range: 4 ~ 200
Default: 20
Display Icon
The AVG icon is displayed at the bottom of the screen when the
Average function is on.
February 2017
Figure 38.
> Average[F4] and toggle Average on or off.