Display
►
Tip
►
Signals that are masked by the noise floor level should have a ratio of less
than 1 to smooth the noise out.
Signals with strong frequency components should use a ratio equal to or
greater than 1.
3.5.4
Average Trace
Description
►
The Average function averages the trace for a user-defined number of times
before it is displayed. This feature smoothes the noise level, but has the
drawback of slowing down the display update rate.
Operation
►
Press
Set the number of averages.
Range: 4 ~ 200
Default: 20
Display Icon
►
The AVG icon is displayed at the bottom of the screen when the
Average function is on.
February 2017
Figure 38.
> Average[F4] and toggle Average on or off.
44