Delete Limit Line; Pass Fail Testing - Promax AE-166 User Manual

Spectrum analyser
Table of Contents

Advertisement

4.2.4

Delete Limit Line

Description
Any one of the 5 limit lines can be deleted.
Activate Correction
Press
line (limit line 1~5) to delete.
Press Delete Limit Line[F6]. The data from the chosen limit line will be
deleted.
4.2.5

Pass Fail Testing

Description
Before pass/fail testing can begin, limit lines for the upper and lower limits
must first be saved.
Operation
Press
To set a high limit, press High Limit[F1] and choose one of the limit lines
as the upper (high) limit.
To set the low limit, press Low Limit[F2] and select one of the limit lines
as the lower limit.
Press Pass Criterion[F3] and select the pass criteria.
Criteria: All-In, Max-In, Min-In
Press Pass/Fail Mode[F5] to select what the AE-166/167 will do on a
fail judgment. Single will stop testing after a single fail. Continue will
continue testing after each fail judgment.
Pass/Fail Mode: Single, Continue
Press Pass/Fail Test[F4] and turn the testing on.
The test result appears in the bottom of the display, and the high and
low limit lines (if enabled) appear on the display.
Pass:
Fail:
Display Icon
The alarm icon is shown at the bottom of the display whenever
testing is turned on.
February 2017
> Edit Select Limit[F1] > Limit Line[F1] and choose a limit
> Pass/Fail Test.
, with green grid border.
, with red grid border.
138