Ae-166/167 Specifications; Frequency - Promax AE-166 User Manual

Spectrum analyser
Table of Contents

Advertisement

8.5

AE-166/167 Specifications

The specifications apply when the AE is powered on for 45 minutes* to warm-
up to a temperature of 20°C to 30°C, unless specified otherwise:
8.5.1

Frequency

Frequency
Range
Resolution
Frequency Reference
Accuracy
Aging Rate
Frequency Stability
over Temperature
Supply Voltage
Stability
Frequency Readout Accuracy
Start, Stop, Center,
Marker
Trace points
Marker Frequency Counter
Resolution
Accuracy
Frequency Span
Range
Resolution
Accuracy
* 45 minutes typical, 90 minutes maximum
215
9 kHz to 3 GHz
1 Hz
±(period since last adjustment X aging rate)
+ stability over temperature + supply
voltage stability
±1 ppm max.
±0.025 ppm
±0.02 ppm
±(marker frequency indication X frequency
reference accuracy + 10% x RBW +
frequency resolution
Max 601 points, min 6 points
1 Hz, 10 Hz, 100 Hz, 1 kHz
±(marker frequency
indication X frequency
reference accuracy +
counter resolution)
0 Hz (zero span),
100 Hz to 3 GHz
1 Hz
± frequency resolution
1 year after last
adjustment
0 to 50 °C
)
1
RBW/Span > =
0.02 ;
Mkr level to DNL >
30 dB
1
RBW: Auto;
February 2017

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ae-167

Table of Contents