3.9

Marker

A Marker shows the frequency and amplitude of a waveform point. The
AE-166/167 can activate up to 6 markers or marker pairs simultaneously as
well as up to 10 peak markers in the marker table.
The marker table helps editing and viewing multiple markers in a single display.
A delta marker shows the frequency and amplitude difference from a reference
marker.
The AE-166/167 can automatically move a marker to various locations
including the peak signal, center frequency, and start/stop frequency. Other
marker operations regarding signal peaks are available in the Peak Search
function.
Activating a Marker
Move Marker Manually
Move Marker to Preset Locations
Activate Delta Marker
Move Delta Marker(s)Manually
Marker Functions
Move Marker to Trace
Show Markers in Table
Peak Search
Peak Configuration
Peak Table
Activating a Marker
There are two basic marker types, normal markers and delta markers. Normal
markers are used to measure the frequency/time or amplitude of a point on the
trace. Delta markers are used to measure the difference between a reference
point and a selected point on the trace.