4.4.2

Normalize the Tracking Generator

Background
The normalize function subtracts the trace after each sweep with a reference
trace. The resultant trace is added to a normalized reference level.
Connection
When normalizing the TG output, connect the TG output directly to the RF
input.
After normalization, connect the DUT to the tracking generator and connect
the output of the DUT to the RF input.
Operation
Press
generator on.
Press Normalize[F6] to enter the Normalization menu.
Press Norm. Ref. Level[F2] to set the vertical level of the normalized
reference.
Range: -100 dB ~ 100 dB
Press Norm. Ref. Position[F3] offsets the normalized trace on the screen.
Range: 10 ~ 0 grid divisions. (top to bottom)
Press Norm.[F4] to toggle the normalized data on/off.
Alternatively, press Exe. Norm.[F1] to perform the normalization again.
145
Figure 116.
> Tracking Generator[F1] > TG[F1] and toggle the tracking