8.5.2

Amplitude

Amplitude Range
Attenuator
Maximum Safe Input Level
1 dB Gain Compression
Displayed Average Noise Level (DANL)
217
Measurement
Range
Input Attenuator
Range
Average Total
Power
DC Voltage
Total Power at 1st
Mixer
Total Power at the
Preamp
Preamp off
9 kHz to 100 kHz
100 kHz to 1 MHz
1 MHz to 2.7 GHz
2.7 GHz to 3 GHz
Preamp on
100 kHz to 1 MHz
1 MHz to 10 MHz
10 MHz to 3 GHz
0 to 50 dB, in 1 dB
step
= +33 dBm
± 50 V
> 0 dBm
> -22 dBm
mixer power level (dBm)= input power
(dBm)-attenuation (dB)
4
0 dB attenuation; RF Input is terminated
with a 50 Ω load. RBW 10 Hz; VBW 10 Hz;
span 500 Hz; reference level = -60dBm;
trace average = 40
< -93 dBm
< -90 dBm - 3 x
(f/100 kHz) dB
< -122 dBm
< -116 dBm
0 dB attenuation; RF Input is terminated
with a 50 Ω load ; RBW 10 Hz; VBW 10 Hz;
span 500 Hz; reference level = -60dBm;
trace average = 40
Displayed Average
Noise Level (DANL)
to 18 dBm
DANL to 21 dBm
DANL to 30 dBm
Auto or manual
setup
Input attenuator
=10 dB
Typical;Fc = 50
MHz; preamp. off
Typical;Fc = 50
MHz; preamp. on
Nominal
February 2017