Selecting The Trigger Mode - Promax AE-166 User Manual

Spectrum analyser
Table of Contents

Advertisement

Press Action Now[F5] to activate the external trigger.
The system will now wait for the trigger conditions to be matched before
starting a sweep.
Display Icon
The EXT Trigger icon is displayed when the external trigger is
activated.
NOTE:
3.8.4

Selecting the Trigger Mode

Description
In free run mode all signals are captured and the trigger conditions are not
used.
Modes
Normal: The spectrum analyser captures every signal that meets the
trigger conditions.
Single: The spectrum analyser captures the first signal that meets the
trigger conditions.
Continuous: The spectrum analyser captures the first signal that meets
the trigger conditions then switches to free run mode thereafter.
Operation
Press
Nor.: Normal
Sgl.: Single
Cont.: Continuous
Press Action Now[F5] to manually start triggering.
Display Icons
Normal:
February 2017
The trigger will revert back to the Free Run mode if any
parameter settings are changed, such as the span or
amplitude settings.
> Trigger Mode[F3] to toggle the trigger mode:
Single:
Continuous:
62