Operation: trigger configuration
Press AF Trigger[F5] > Trigger Setup[F2] to set the triggering conditions:
Edge Slope[F1]: Sets the trigger to rising or falling edge.
Trigger Mode[F2]: Sets the triggering mode:
Trigger Level[F3]: Sets the trigger level as a frequency. (The
displayed level will only remain for a few moments).
Trigger Delay[F4]: Sets the trigger delay time: 0 to 1 ks
Press Return[F7] to return to the AF Trigger menu and set the remaining
triggering options:
FreeRun[F1]: Disables the trigger, this is the default setting.
Start Time[F3]: Sets the start time for the x-axis for the FM waveform
in the top-half of the screen.
Stop Time[F4]: Sets the stop time for the x-axis for the FM waveform
in the top-half of the screen.
Action Now[F5]: Turns FreeRun mode off and uses the user-defined
trigger settings.
NOTE:
4.1.7

FM Pass Fail Testing

Description
The Limit Edit function puts a pass limit on the FM deviation, carrier offset and
carrier power.
Measurement Range
Frequency Deviation: 40 Hz ~ 400 kHz, 1 Hz measurable
Carr. Offset: 1 Hz ~ 400 kHz
Carrier Power: -120 dBm ~ 30 dBm
101
Norm.: Normal trigger
Sgl.: Single trigger
Cont.: Continuously trigger
The MAX and MIN measurements are held until higher or lower
values are found. To reset the MAX and MIN measurements,
press
> Demod[F2] > AM Analysis[F1] > Min/Max
Reset[F3].