Press Offset/Limit[F3] to set the offset parameters:
SelectOffset[F1]: Select which offset to edit.
[F2]: Toggles the selected offset on/off.
StartFreq[F3]: Sets the start frequency of the selected offset.
StopFreq[F4]: Sets the Stop Frequency of the selected offset.
RBW[F5]: Sets the RBW of the selected offset.
Press More 1/2[F6] to set absolute and relative level limits and
conditions:
Abs Start[F2]: Sets the absolute start level limit for the selected
offset.
Abs Stop[F3]: Sets the absolute stop level limit for the selected offset
Man: Allows a user-defined Abs Stop level
Couple: Sets the Abs Stop level to the Abs Start level.
Rel Start[F4]: Sets the relative start level limit for the selected offset.
Rel Stop[F5]: Sets the relative stop level for the selected offset.
Man: Allows a user-defined Abs Stop level.
Couple: Sets the Rel Stop level to the Rel Start level.
Press Fail Mask[F6] to set the Fail Mask conditions:
Absolute[F1]: Sets the fail condition to the Absolute level limit.
Relative[F2]: Sets the fail condition to the relative level limit.
Abs AND Rel[F3]: Sets the fail condition as both the absolute and
relative level limits.
Abs OR Rel[F4]: Sets the fail condition to either the absolute or
relative level limits.
Press Select Offset[F1] and repeat the above steps for any other offsets.
Offset: 1 ~ 5
117
February 2017