NOTE:
4.1.14
Composite Second Order (CSO)
Description
►
Composite Second Order measurement calculates the difference in amplitude
between the carrier signal and the composite second order beat.
Parameters
►
CSO CH SP: The channel space.
Measurement items
►
Visual Carrier: frequency, amplitude
Channel Space: frequency
CSO: amplitude difference
Example
►
Operation
►
Press
and choose CSO.
Press CSO[F2] and toggle CSO on.
Any other measurement mode will automatically be disabled.
The display splits into two screens. The top shows the trace with the
visual carrier marker and the CSO beat marker. The bottom screen
shows the CSO measurements.
123
Ensure the aural and color subcarriers are turned off when
making CNR measurements.
Figure 86.
> More[F7] > CNR/CSO/CTB[F1] > Setup[F1] > CSO[F2]
February 2017